  "results": [
    {
      "status": "Ongoing",
      "city": "Irving",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85642",
      "recalling_firm": "Acuity Surgical Devices, LLC",
      "address_1": "8710 N Royal Ln",
      "address_2": "N/A",
      "postal_code": "75063-2539",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Other",
      "distribution_pattern": "US Nationwide distribution including in the states of TX, FL, NC, LA, KY, WI, SD, IL, CA, and KS.  O.U.S.: None",
      "recall_number": "Z-2354-2020",
      "product_description": "A LINK Z INTERVERTEBRAL BODY FUSION DEVICE    Product Description:  A Link Z Ti Unitary Plate-Cage 32x21x15 7  A Link Z Ti Unitary Plate-Cage 32x21x15 12  A Link Z Ti Unitary Plate-Cage 32x21x15 16  A Link Z Ti Unitary Plate-Cage 32x21x15 16  A Link Z Ti Unitary Plate-Cage 32x21x15 16  A Link Z Ti Unitary Plate-Cage 32x21x15 20  A Link Z Ti Unitary Plate-Cage 32x21x15 20  A Link Z Ti Unitary Plate-Cage 32x21x15 20  A Link Z Ti Unitary Plate-Cage 32x21x17 12  A Link Z Ti Unitary Plate-Cage 32x21x17 16  A Link Z Ti Unitary Plate-Cage 32x21x17 20  A Link Z Ti Unitary Plate-Cage 32x21x19 20  A Link Z Ti Unitary Plate-Cage 32x24x15 7  A Link Z Ti Unitary Plate-Cage 32x24x15 16  A Link Z Ti Unitary Plate-Cage 36x26x15 12  A Link Z Ti Unitary Plate-Cage 36x26x15 12  A Link Z Ti Unitary Plate-Cage 36x26x15 16  A Link Z Ti Unitary Plate-Cage 36x26x15 16  A Link Z Ti Unitary Plate-Cage 36x26x15 16  A Link Z Ti Unitary Plate-Cage 36x26x15 16  A Link Z Ti Unitary Plate-Cage 36x26x15 20  A Link Z Ti Unitary Plate-Cage 36x26x15 20  A Link Z Ti Unitary Plate-Cage 36x26x15 20  A Link Z Ti Unitary Open Interbody 32x24x12 7  A Link Z Ti Unitary Open Interbody 32x24x12 12  A Link Z Ti Unitary Open Interbody 32x24x16 7  A Link Z Ti Unitary Open Interbody 32x24x16 12    Product Usage: is a spinal device that is implanted in the intervertebral body space via an anterior approach to improve stability of the spine while supporting fusion.",
      "product_quantity": "A total of 27 lots producing 178 units",
      "reason_for_recall": "Due to a manufacturing process issue, it allowed vertebral body bone screws to not seat properly within the spinal device/cage.",
      "recall_initiation_date": "20190107",
      "center_classification_date": "20200611",
      "report_date": "20200617",
      "code_info": "Product  Product Description                             Lot  Number                                                 Number  14-32211507 A Link Z Ti Unitary Plate-Cage 32x21x15 7          1000638  14-32211512 A Link Z Ti Unitary Plate-Cage 32x21x15 12     1000259  14-32211516 A Link Z Ti Unitary Plate-Cage 32x21x15 16    1000260  14-32211516 A Link Z Ti Unitary Plate-Cage 32x21x15 16    1000325  14-32211516 A Link Z Ti Unitary Plate-Cage 32x21x15 16    1000529  14-32211520 A Link Z Ti Unitary Plate-Cage 32x21x15 20    1000261  14-32211520 A Link Z Ti Unitary Plate-Cage 32x21x15 20    1000469  14-32211520 A Link Z Ti Unitary Plate-Cage 32x21x15 20    1000556  14-32211712 A Link Z Ti Unitary Plate-Cage 32x21x17 12    1000262  14-32211716 A Link Z Ti Unitary Plate-Cage 32x21x17 16    1000640  14-32211720 A Link Z Ti Unitary Plate-Cage 32x21x17 20    1000641  14-32211920 A Link Z Ti Unitary Plate-Cage 32x21x19 20    1000530  14-32241507 A Link Z Ti Unitary Plate-Cage 32x24x15 7          1000211  14-32241516 A Link Z Ti Unitary Plate-Cage 32x24x15 16    1000536  14-36261512 A Link Z Ti Unitary Plate-Cage 36x26x15 12    1000282  14-36261512 A Link Z Ti Unitary Plate-Cage 36x26x15 12    1000583  14-36261516 A Link Z Ti Unitary Plate-Cage 36x26x15 16    1000232  14-36261516 A Link Z Ti Unitary Plate-Cage 36x26x15 16    1000383  14-36261516 A Link Z Ti Unitary Plate-Cage 36x26x15 16    1000479  14-36261516 A Link Z Ti Unitary Plate-Cage 36x26x15 16    1000652  14-36261520 A Link Z Ti Unitary Plate-Cage 36x26x15 20    1000233  14-36261520 A Link Z Ti Unitary Plate-Cage 36x26x15 20    1000360  14-36261520 A Link Z Ti Unitary Plate-Cage 36x26x15 20    1000584  15-32241207 A Link Z Ti Unitary Open Interbody 32x24x12 7  1000572  15-32241212 A Link Z Ti Unitary Open Interbody 32x24x12 12         1000573  15-32241607 A Link Z Ti Unitary Open Interbody 32x24x16 7  1000576  15-32241612 A Link Z Ti Unitary Open Interbody 32x24x16 12         1000577"
    }
  ]
}