{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Bowdoin",
      "address_1": "1201 Main St",
      "reason_for_recall": "The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.",
      "address_2": "",
      "product_quantity": "2952 total",
      "code_info": "UDI: 00873263002162",
      "center_classification_date": "20180705",
      "distribution_pattern": "The devices were distributed to the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY.    The devices were distributed to the following foreign countries:  Argentina, Australia, Canada, Columbia, Denmark, Ecuador, France, Germany, Great Britain, Italy, Japan, Norway, Romania, South Africa, Switzerland,  Saudi Arabia, Taiwan.",
      "state": "ME",
      "product_description": "microTargeting\" STar\" Drive  System (Manual), Product Number 70-ZD-MA",
      "report_date": "20180711",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "FHC, Inc.",
      "recall_number": "Z-2351-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80148",
      "termination_date": "20190430",
      "more_code_info": "",
      "recall_initiation_date": "20180510",
      "postal_code": "04287-7302",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}