  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "New Freedom",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88365",
      "recalling_firm": "Oakworks Inc",
      "address_1": "923 E Wellspring Rd",
      "address_2": "N/A",
      "postal_code": "17349-8408",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-2349-2021",
      "product_description": "Vascular Table without Fowler (USV1) and Ultrasound Vascular Table with Fowler (USV2). Intended Use: Vascular Ultrasound examination table Catalog Number: 66783 and 64219",
      "product_quantity": "55 units",
      "reason_for_recall": "Weld failure- the bracket attaches the tabletop tilting actuator to the tabletop frame. If the bracket weld fails, the tabletop tilting action is non-functional",
      "recall_initiation_date": "20210629",
      "center_classification_date": "20210825",
      "termination_date": "20231005",
      "report_date": "20210901",
      "code_info": "Serial Number/Part Number VCF766794 USGNV22723723MTTZZCE2S; VCF767303 USGNV23023723MTTZZ2S; VCF771999 USGNV23023723MTTZZ2S; VCF774252 USGNV23023723MTTZZ2S; VCF776073 USGNV23023723MTTZZ2S; VCF776198 USGNV23023723MTTZZ2S; VCF778246 USGNV23023723MTTZZ2S; VCF778247 USGNV23023723MTTZZ2S; VCF779362 USGNV23023723MTTZZ2S; VCF780260 USGNV23023723MTTZZ2S; VCS781126 USGNV13023723MTTZZ2S; VCF783652 USGNV22723723MTTZZF82S; VCF783969 USGNV23023723MTTZZF82S; VCF785961 USGNV22723723MTTZZCE2S; VCS766698 USGNV12723723MTTZZF62S; VCF768278 USGNV23023723MTTZZCEDRMC2S; VCF770016 USGNV23023723MTTZZ2S; VCF770017 USGNV23023723MTTZZ2S; VCF770571 USGNV23023723MTTZZCE2S; VCF771782 USGNV23023723MTTZZF8MC2S; VCF776974 USGNV22723723MTTZZ2S; VCF778602 USGNV23023723MTTZZ2S; VCF779609 USGNV23023723MTTZZF8CE2S; VCF781198 USGNV23023723MTTZZ2S; VCF784255 USGNV23023723MTTZZCE2S; VCF785376 USGNV23023723MTTZZ2S; VCF785650 USGNV23023723MTTZZF8MC2S; VCF786137 USGNV23023723MTTZZ2S; VCF786173 USGNV23023723MTTZZCE2S; VCF771553 USGNV23023723MTTZZCEMC2S; VCF771554 USGNV23023723MTTZZCEMC2S; VCF785065 USGNV22723723MTTZZF8CEDRDL2S; VCF771992 USGNV23023723MTTZZF8DRDL2S; VCF771993 USGNV23023723MTTZZF8DRDL2S; VCF782916 USGNV23023723MTTZZF8DRDL2S; VCF772035 USGNV22723723MTTZZ2S; VCF777288 USGNV22723723MTTZZ2S; VCF779024 USGNV22723723MTTZZ2S; VCF779025 USGNV22723723MTTZZ2S; VCF779026 USGNV22723723MTTZZ2S; VCF780977 USGNV22723723MTTZZ2S; VCF785219 USGNV22723723MTTZZ2S; VCF771311 USGNV23023723MTTZZ2S; VCF785744 USGNV22723723MTTZZCE2S; VCS772698 USGNV12723723MTTZZ2S; VCF771236 USGNV22723723MTTZZF8CE2S; VCF775658 USGNV23023723MTTZZ2S; VCF776370 USGNV23023723MTTZZ2S; VCF792903 USGNV23023723MTTZZMC2S; VCF773748 USGNV23023723MTTZZF8MC2S; VCF773748 USGNV23023723MTTZZCE2S; VCF779429 USGNV23023723MTTZZF8CE2S; VCF779921 USGNV23023723MTTZZF82S; VCF783233 USGNV23023723MTTZZ2S; VCF786329 USGNV22723723MTTZZ2S;"
    }
  ]
}