  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Rochester",
      "address_1": "100 Indigo Creek Dr",
      "reason_for_recall": "There is a potential for imprecise results when using the slides.  Increase in imprecision when using PHYT coating 0165. This leads to higher and lower than expected results for both quality control samples and patient samples.",
      "address_2": "",
      "product_quantity": "5712",
      "code_info": "Lot Numbers: 2616-0165-4866, 2616-0165-4904, 2616-0165-4870, 2616-0165-4869, 2616-0165-5266, 2616-0165-5307, 2616-0165-5233, 2616-0165-5232, 2616-0165-5611, 2616-0165-5610, 2616-0165-5648, 2616-0165-5649, 2616-0165-6218, 2616-0165-6275, 2616-0165-6220, 2616-0165-6221, 2616-0165-6219, 2616-0165-6276, 2616-0165-6791, 2616-0165-6832, 2616-0165-6792, 2616-0165-6787, 2616-0165-7225, 2616-0165-7226, 2616-0165-7223, 2616-0165-7214, 2616-0165-7222",
      "center_classification_date": "20180703",
      "distribution_pattern": "Worldwide distribution.  US nationwide, Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Poland, Portugal, Russia, Singapore, Spain, and Sweden",
      "state": "NY",
      "product_description": "VITROS Chemistry Products PHYT Slides, Catalog Number 8298671.  in vitro diagnostic",
      "report_date": "20180711",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Ortho-Clinical Diagnostics",
      "recall_number": "Z-2344-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80163",
      "termination_date": "20200818",
      "more_code_info": "",
      "recall_initiation_date": "20180509",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}