{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mountain View",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71697",
      "recalling_firm": "Siemens Medical Solutions USA, Inc.",
      "address_1": "685 E Middlefield Rd",
      "address_2": "N/A",
      "postal_code": "94043-4045",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and the countries of Algeria, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Brazil, Brunei, Bulgaria, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Dem. Rep. Congo, Denmark, Dominican Rep., Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Lithuania, Luxembourg, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, P.R. China, Peru, Philippines, Poland, Portugal, Republic Korea, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, U.A.E., Ukraine, United Kingdom,  Venezuela and Vietnam.",
      "recall_number": "Z-2344-2015",
      "product_description": "ACUSON S Family ultrasound systems utilizing the 18L6 HD and/or 6C1 HD biopsy guidelines    6C1HD - model numbers:  10442008 6C1HD Needle Guide, S3000;  10440586 6C1 HD Biopsy Guide, S2000  18L6 - model numbers:  10442007 18L6 Biopsy Guide, S3000  10044200 S2000 Biopsy Starter Kit, 18L6 HD  10855189 18L6 HD Biopsy Starter Kit, S1000    Radiology: The ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.",
      "product_quantity": "4140 total",
      "reason_for_recall": "When using the CIVCO biopsy attachments in combination with the ACUSON S Family on-screen guidelines, the biopsy needle may traverse outside the on-screen  guidelines.",
      "recall_initiation_date": "20150629",
      "center_classification_date": "20150805",
      "termination_date": "20180118",
      "report_date": "20150812",
      "code_info": "6C1HD - model numbers:  10442008 6C1HD Needle Guide, S3000;  10440586 6C1 HD Biopsy Guide, S2000;  18L6 model numbers:  10442007 18L6 Biopsy Guide, S3000  10044200 S2000 Biopsy Starter Kit, 18L6 HD  10855189 18L6 HD Biopsy Starter Kit, S1000    See Consignee list for all serial numbers."
    }
  ]
}