  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Snoqualmie",
      "address_1": "35301 SE Center St",
      "reason_for_recall": "The firm has received multiple reports of the Non-Invasive Blood Pressure (NIBP) parameter becoming non-functional with an associated message (No Reading). In addition, there is one of the following three error messages: Inflate Error, HW Error, or No Data.",
      "address_2": "",
      "product_quantity": "3801 modules and 116 PCBAs & valves distributed in the US; 1297 modules and 366 PCBAs & valves distributed outside the US",
      "code_info": "Manifold Lot Codes: 128280, 129420, 129940, or 155640.",
      "center_classification_date": "20160804",
      "distribution_pattern": "Worldwide Distribution - USA Distribution and to the countries of :  AFGHANISTAN, ARGENTINA, AUSTRALIA, BAHRAIN, BOLIVIA, BRAZIL, BRUNEI DARUSSALAM; CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DOMINICAN REPUBLIC;   ECUADOR, EGYPT, ESTONIA, FRANCE, GERMANY, GUATEMALA, HONG KONG, INDIA, IRAN (ISLAMIC REPUBLIC OF), ISRAEL, ITALY, JORDAN, KUWAIT, MEXICO, MOROCCO, NETHERLANDS, NICARAGUA, PAKISTAN, PALESTINE, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, SAUDI ARABIA, Slovakia, SPAIN, SRI LANKA, SWITZERLAND, TAIWAN, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, and VIET NAM.",
      "state": "WA",
      "product_description": "91496 Ultraview SL Command Module, Options A, B, C, and I.    Option A: ECG/Respiration, SpO2, Temperature x2 & adult/neonate NIBP.    Option B: ECG/Respiration, Invasive Pressure x2, SpO2, Temperature x2 & adult/neonate NIBP.    Option C: ECG/Respiration, Invasive Pressure x4, SpO2, Cardiac Output, Temperature x2 & adult/neonate NIBP.    Option I: SpO2, Temperature x2, & adult/neonate NIBP.",
      "report_date": "20160810",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Spacelabs Healthcare Inc",
      "recall_number": "Z-2343-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74549",
      "termination_date": "20170522",
      "more_code_info": "",
      "recall_initiation_date": "20160701",
      "postal_code": "98065-9216",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}