      "status": "Terminated",
      "city": "Cambridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88461",
      "recalling_firm": "Philips North America Llc",
      "address_1": "222 Jacobs St",
      "address_2": "N/A",
      "postal_code": "02141-2289",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO,  MD, MD, NH, NY, OH, TN, TX, VA and the countries of Canada, Mexico, Australia, Austria,  Chile, China, Czech Republic, Denmark, France, Germany, India, Indonesia, Italy, Japan,  Korea, Republic of Lebanon, Netherlands, Romania, Russian Federation, Saudi Arabia,  Spain, Turkey, Ukraine, United Kingdom.",
      "recall_number": "Z-2342-2021",
      "product_description": "Ingenia 1.5T-Magnetic Resonance Medical Electrical Systems indicated for use as a  diagnostic device. The system can produce cross-sectional images, spectroscopic images  and/or spectra in any orientation of the internal structure of the head, body or extremities  Model Number: 781341",
      "product_quantity": "3 units US; 24 units OUS",
      "reason_for_recall": "Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury",
      "recall_initiation_date": "20210730",
      "center_classification_date": "20210825",
      "termination_date": "20230725",
      "report_date": "20210901",
      "code_info": "S/N and UDI: 84344 (01)00884838055322(21)84344; 87019 (01)00884838055322(21)87019; 87231 (01)00884838055322(21)87231.  OUS: 87171   (01)00884838055322    84943   (01)00884838055322    87142   (01)00884838055322    87249   (01)00884838055322    87252   (01)00884838055322    87195   (01)00884838055322    87129   (01)00884838055322    87136   (01)00884838055322    87175   (01)00884838055322    87196   (01)00884838055322    87197   (01)00884838055322    87235   (01)00884838055322    87187   (01)00884838055322    87212   (01)00884838055322    84572   (01)00884838055322    87116   (01)00884838055322    87152   (01)00884838055322    87161   (01)00884838055322    87237   (01)00884838055322    87244   (01)00884838055322    87090   (01)00884838055322    87093   (01)00884838055322    87259   (01)00884838055322    87209   (01)00884838055322"
    }
  ]
}