      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71711",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "51 Valley Stream Pkwy",
      "address_2": "N/A",
      "postal_code": "19355-1406",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide",
      "recall_number": "Z-2341-2015",
      "product_description": "Artis Q systems;  dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.  Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures.",
      "product_quantity": "N/A",
      "reason_for_recall": "The possibility exists for a large amount of liquid to seep through the top cover of the table and penetrate the components inside the table. If this occurs, table movements may no longer be possible. No injuries reported.",
      "recall_initiation_date": "20150629",
      "center_classification_date": "20150805",
      "termination_date": "20161207",
      "report_date": "20150812",
      "code_info": "Model Number of device: 10848282  10848281  10848355  10848354  serial numbers: 121006  121013  109016  121011  123008  123003  123011  121014  109011  121004  121018  121036  121012  109015  109009  121019  121010  121015  121016  109007  111000  121020  121009  121056  121057  109021  121008  109020  109027  109010  123010  121041"
    }
  ]
}