  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71788",
      "recalling_firm": "GE Healthcare",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Internationally to UNITED ARAB EMIRATES, ARGENTINA, BRAZIL, INDIA, EGYPT, CHINA,  FINLAND, FRANCE, UNITED KINGDOM, HUNGARY,  JAPAN, INDONESIA,   KOREA, MEXICO, NEW ZEALAND, CANADA, AUSTRALIA, ITALY,   RUSSIAN FEDERATION, TAIWAN, and SOUTH AFRICA.",
      "recall_number": "Z-2339-2015",
      "product_description": "GE Healthcare, Signa Infinity MR System; Signa MR/i MR System; SmartSpeed; EchoSpeed Plus; HighSpeed Plus, Signa Infinity MR System with EXCITE Technology, Excite II, Signa 3T Infinity with Excite, Signa Excite 1.5T and Signa Excite 3T, Signa Excite HD, Signa HD 1.5T and Signa HD 3.0T, Signa HDx, Signa HDxt, Signa HDi.   Whole body magnetic resonance diagnostic imaging scanners.",
      "product_quantity": "82 (39 USA; 43 OUS)",
      "reason_for_recall": "A gradient cable in the ACGD cabinet may be missing a spacer that prevents connecting the gradient cable terminals in the wrong polarity on X gradient amplifier. This could cause images to be flipped left-right with incorrect orientation annotation.",
      "recall_initiation_date": "20150710",
      "center_classification_date": "20150805",
      "termination_date": "20151119",
      "report_date": "20150812",
      "code_info": "Mfg Lot or Serial # System ID  00000153282MR9 501362MR1  00000165901MR0 ZA2101MR02  00000213818MR8 37190MRI  00000213341MR1 M27590  00000247386MR6 707459MR2  00000231538MR0 706583MR  00000236418MR0 850862FTS1CH1  00000242668MR2 502868MR  00000267704MR5 314525SAMR2  00000227724MR2 ALLIANCESIG262  00000214667MR8 853270051  00000242173MR3 ALLIANCESIG287  00000241518MR0 502633MR1  00000231910MR1 212956MR2  00000237000MR5 0910272037  00000241867MR1 HU1008MR01  00000243769MR7 920279006  00000047514WH5 YM0774  00000285490MR9 954916MR1  00000284687MR1 309655MRI1A  00000280860MR8 1202470  00000282692MR3 902496MRI3  00000282813MR5 0835270010  00000278669MR7 508755MR1  00000267211MR1 0850270458  000000HDMR3141 SHIMR598  000000HDMR5777 919470MR2  000000HDMR3905 709737MRI  000000HDMR2541 M4018076  000000HDMR5379 M4135138  00000HDMR10710 10543MRM01  00000301019MR6 SMSDML2  00000292334MR0 INSIGHTMR1915  00000290735MR0 704323MR1  00000300812MR5 MR14604YR3  00000297420MR2 082427080028  00000239401MR3 404352PNC  00000209414MR2 732591MR  00000206394MR9 516365MR  00000224271MR7 RADNETMR1515  00000190316MR0 1167491  00000223494MR6 1299290  00000223256MR9 84631  00000227239MR1 2817688  00000230273MR5 2817679  00000240274MR1 083027002100393  00000199094MR4 YM0620  00000237547MR5 600101MR01  00000210080MR8 941262MR  00000214332MR9 208367LX5  00000204255MR4 770438EAMR  00000218279MR8 2214SIGNA15  00000215878MR0 0853270047  00000193672MR3 AE1011MR01  00000001634M25 MRR8049  00000005464M23 082427070113  00000000108M21 2807539  00000267104MR8 864225MR  00000254183MR7 304347MR1  00000264011MR8 EG1522MR01  00000006705YR8 901226MR1  00000007870YR9 2651831  00000009552YR1 FI1012MR04  00000009010YR0 FI1012MR03  00000006662YR1 0850270197  002003-0922-09 ALLIANCESIG293  0002005-0725-4 5360MREXCITE3  00000279502MR9 304766TMR  00000284315MR9 208265BMR  00000285664MR9 904399MR2  00000280274MR2 561798MR1  00000286778MR6 2718385  00000276273MR0 4716MREXCITE  00000299465MR5 EG1279MR01  00000309342MR4 082427010077  00000264527MR3 SMSFFI2  00000263933MR4  A5156029  00000266764MR0 RADNETMR1513  00000291209MR5 415780KMR  00000290984MR4 214820XBMR1   928649MR   08302703148203"
    }
  ]
}