      "status": "Terminated",
      "city": "Mentor",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62895",
      "recalling_firm": "Steris Corporation",
      "address_1": "5960 Heisley Rd",
      "address_2": "N/A",
      "postal_code": "44060-1834",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -- USA (nationwide) and the countries of  Bahrain and Hong Kong.",
      "recall_number": "Z-2335-2012",
      "product_description": "The SYSTEM 1E Liquid Sterilant Processing System.    The SYSTEM 1E Liquid Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersible and reusable critical and semi-critical heat sensitive medical devices, including endoscopes and their accessories.",
      "product_quantity": "6,051 units",
      "reason_for_recall": "The firm initiated a recall after complaint investigations revealed that when the units are left on and not in use at night and/or on weekends the hose disconnects/separates resulting in water leakage that can cause damage to property.",
      "recall_initiation_date": "20120801",
      "center_classification_date": "20120907",
      "termination_date": "20140606",
      "report_date": "20120919",
      "code_info": "Model #: P6500, S/N 400000 through 406294"
    }
  ]
}