  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88341",
      "recalling_firm": "Becton, Dickinson and Company, BD Biosciences",
      "address_1": "2350 Qume Dr",
      "address_2": "N/A",
      "postal_code": "95131-1812",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution in the states of CT, FL, KY, NY, and PA. The country of Taiwan.",
      "recall_number": "Z-2333-2021",
      "product_description": "BD\" Mouse IgG2a Isotype Control APC X39 ASR",
      "product_quantity": "23 reagents",
      "reason_for_recall": "Complaints have been received for performance issues related to high background and staining.  this issues is related to a manufacturing process error that can potentially cause a false positive staining due to contamination with another antibody.",
      "recall_initiation_date": "20210722",
      "center_classification_date": "20210824",
      "termination_date": "20230324",
      "report_date": "20210901",
      "code_info": "Catalog Number:  340757 UDI: 00382903407576 Lot Number: 0253432"
    }
  ]
}