  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Tustin",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85309",
      "recalling_firm": "Canon Medical System, USA, INC.",
      "address_1": "2441 Michelle Dr",
      "address_2": "N/A",
      "postal_code": "92780-7047",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including in the states of PA, AR, OH, NC, CA, IA, GA, IL, AL, WV, LA, MO, FL, MS, TX, AZ, MA, NJ, MD, NY, NV, OK, WI, PR.",
      "recall_number": "Z-2326-2020",
      "product_description": "Infinix-8000C with Catheterization Table CAT-850B or CAT-860B - Product Usage: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.",
      "product_quantity": "24",
      "reason_for_recall": "Bolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.",
      "recall_initiation_date": "20200303",
      "center_classification_date": "20200609",
      "report_date": "20200617",
      "code_info": "Serial Numbers:  B8B1972001,  A9A1592001,  A3A14Z2001,  AEB1812002,  A1B1552001,  A9C18Y2007,  A3E1922002,  B5B1952001,  A9C1822006,  A9B1622002,  A9B1722003,  A5C1622001,  B3B19Y2001,  A1D1772003,  B1B19X2001,  A5C16Y2002,  A9C1822005,  A1E1862008,  A1D1752002,  A1E1832004,  A1E1852005,  A1E1852006,  A5E1862003,  AEA17X2001"
    }
  ]
}