{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Winterthur",
      "state": "",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74548",
      "recalling_firm": "Zimmer Gmbh",
      "address_1": "Sulzer Allee 8",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic: CO, IL\t    Foreign: Argentina, Costa Rica, India, Malta, Japan, Philippines, Vietnam, Portugal, Dominican Republic, Argentina, Australia, Russia, Saudi Arabia, Ireland, Ecuador, China, Singapore, Indonesia, Canada, Mexico, South Korea , Spain, Netherlands, Sweden, Malaysia, New Zealand, Switzerland, El Salvador, England, Norway, Finland, Venezuela, Taiwan, Italy, Colombia, Greece, Thailand, France, Israel, Jamaica, Japan,                          Jordan, Germany, Korea, Lebanon, Belgium, Hong Kong,                   Philippines, Iran, Brazil, Turkey, United Arab Emirates  (UAE), Uruguay",
      "recall_number": "Z-2324-2016",
      "product_description": "V-TEK TWIST D 40X90MM 30MM Cann 12MM Rnd; The twist drill is used to drill holes into bones/tissues for foot, ankle and hand implant systems.",
      "product_quantity": "169",
      "reason_for_recall": "Zimmer Biomet initiated a voluntary recall of the Twist Drill & V-Tek Twist Drill due to to a potential tolerance stack up identified between the twist drill and the mating Kirschner Wire (K-Wire) component.",
      "recall_initiation_date": "20160627",
      "center_classification_date": "20160802",
      "termination_date": "20170530",
      "report_date": "20160810",
      "code_info": "Part No.: 502015640; Lot Numbers:27374, 104547/188A9",
      "more_code_info": ""
    }
  ]
}