  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63042",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "51 Valley Stream Pkwy",
      "address_2": "N/A",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution- including the states of GA, ID, IL, IA, KS, NC, OH, SD, TX, and WI.",
      "recall_number": "Z-2322-2012",
      "product_description": "Siemens Mobilett Mira.    Mobile X-Ray System.",
      "product_quantity": "20",
      "reason_for_recall": "During regular product monitoring, Siemens has become aware of a potential issue with the Mobilett Mira system.  During normal clinical operation the switch on the motor drive handle of the unit may become loose and slide horizontally to the right and become trapped under the cover.  This may cause the system to continuously move.  In response, Siemens has issued Update Instruction XP038/12/S to r",
      "recall_initiation_date": "20120821",
      "center_classification_date": "20120906",
      "termination_date": "20140403",
      "report_date": "20120912",
      "code_info": "Model number 10273100 -- serial numbers 1034, 1038, 1041, 1042, 1043, 1053, 1057, 1070, 1079, 1081, 1088, 1098, 1099, 1102, 1103, 1105, 1106, 1107, 1108, and 1112."
    }
  ]
}