  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Helsinki",
      "state": "N/A",
      "country": "Finland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77198",
      "recalling_firm": "Nexstim PLC",
      "address_1": "Elimaenkatu 9B",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US to GA only, Foreign: Europe",
      "recall_number": "Z-2319-2017",
      "product_description": "Nexstim eXima NBS System Software version 2.2 or higher.  The Nexstim Navigated Brain Stimulation System (NBS System) is indicated for noninvasive mapping of the primary motor cortex of the brain to its cortical gyrus. The NBS System provides information that may be used in the assessment of the primary motor cortex for pre-procedural planning. The NBS System is not intended to be used during a surgical procedure. The NBS System is intended to be used by trained clinical professionals.",
      "product_quantity": "11 units.",
      "reason_for_recall": "Software defect: the NBS software may accidentally generate duplicate copies of one or several files.",
      "recall_initiation_date": "20170314",
      "center_classification_date": "20170608",
      "report_date": "20170614",
      "code_info": "Serial numbers: NBS101, NBS106, NBS140, NBS141, NBS145, NBS148, NBS155, NBS159, NBS163, NBS164, NBS166.  NBS158 has been scrapped."
    }
  ]
}