  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Westminster",
      "state": "CO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77251",
      "recalling_firm": "Zimmer Biomet Spine Inc.",
      "address_1": "10225 Westmoor Dr",
      "address_2": "N/A",
      "postal_code": "80021-2702",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide, military/VA/govt and foreign consignee (No Canadian consignees)",
      "recall_number": "Z-2318-2017",
      "product_description": "Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle \u0013 90 in-lbs Model Number(s): 07.02066.001, 07.02053.001, 730M0017",
      "product_quantity": "497 units",
      "reason_for_recall": "Zimmer Biomet is conducting a medical device recall for Vitality T27 Final Drivers and Vitality Torque Limiting Handle \u0013 90 in-lbs because when applying torque to tighten the closure top the driver has the potential to create a torsional spring which can cause damage and the potential for the values of torque to go out of specification.",
      "recall_initiation_date": "20170510",
      "center_classification_date": "20170608",
      "termination_date": "20180525",
      "report_date": "20170614",
      "code_info": "Model No: 07.02066.001 (All lots); 07.02053.001 (All lots distributed prior to April 26, 2017), and 730M0017 (Lot 309622)"
    }
  ]
}