  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Lake Forest",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98872",
      "recalling_firm": "ICU Medical, Inc.",
      "address_1": "600 N Field Dr",
      "address_2": "N/A",
      "postal_code": "60045-4835",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of ID, IL, IN, KS, MA, MD, NM, NY, PA, TN, TX and the country of Canada.",
      "recall_number": "Z-2313-2026",
      "product_description": "Plum Duo Precision IV Pump, 40002-0403",
      "product_quantity": "13,613",
      "reason_for_recall": "Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.",
      "recall_initiation_date": "20260429",
      "center_classification_date": "20260605",
      "report_date": "20260617",
      "code_info": "UDI-DI M335400031 Plum Duo Precision IV Pump v1.2.3"
    }
  ]
}