  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77167",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "address_2": "N/A",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and countries of: Australia, Austria, Belgium, Bahrain, Canada, China, Costa Rica, Finland, France, Germany, Hong Kong, India, Ireland, Italy, Lithuania, Latvia, Malaysia, Netherlands, Romania, Russia, Saudi Arabia, Singapore, Spain, Switzerland, United Arab Emirates, and United Kingdom.",
      "recall_number": "Z-2313-2017",
      "product_description": "Endoscopic Curved Needle, Surgical Needle Guide, Supplied sterilized by ethylene oxide gas in peel-open packages.    Used for internal suturing under direct vision during intraabdominal endoscopic procedures. The needle driver design affords precise placement and control of standard curved suture needles, similar to open abdominal suturing.",
      "product_quantity": "327 Lots, 725 units",
      "reason_for_recall": "COOK Medical is initiating a voluntary recall of multiple products because the reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilization of these products.",
      "recall_initiation_date": "20170501",
      "center_classification_date": "20170607",
      "termination_date": "20190326",
      "report_date": "20170614",
      "code_info": "CATALOG NUMBER: and GPN: J-ENH-033100 & G16426,  J-ENH-053100 & G16427, J-ENH-053120 & G16428,   J-ENH-053130 & G16429  Dates of Manufacture: January 2012 to February 2017"
    }
  ]
}