  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Pewaukee",
      "address_1": "N27W23910 Paul Rd Ste A",
      "reason_for_recall": "There is a potential for higher than specified surface temperatures involving the 3T 16ch Flex SPEEDER Large Coil.",
      "address_2": "",
      "product_quantity": "39 (11 US, 28 OUS)",
      "code_info": "\"Neocoil   serial Number\"    046-4-0012  046-4-0013  046-4-0014  046-4-0015  046-4-0016  046-4-0017  046-4-0018  046-4-0019  046-4-0020  046-4-0021  046-4-0022  046-4-0023  046-4-0024  046-4-0025  046-4-0026  046-4-0027  046-4-0028  046-4-0029  046-4-0030  046-4-0031  046-4-0032  046-4-0033  046-4-0034  046-4-0035  046-4-0036  046-4-0037  046-4-0038  046-4-0039  046-4-0040  046-4-0041  046-4-0042  046-4-0043  046-4-0044  046-4-0045  046-4-0046  046-4-0047  046-4-0048  046-4-0049  046-4-0050",
      "center_classification_date": "20160801",
      "distribution_pattern": "Worldwide Distribution - US and Internationally to Georgia, Germany, Korea, Netherlands, Pakistan, Spain, Japan, and the United Kingdom.",
      "state": "WI",
      "product_description": "3T 16ch Flex SPEEDER Coil, Large, NeoCoil Part Number NC046200, Toshiba Medical Systems Model MJAJ-222A.  The NeoCoil 3T 16ch Flex SPEEDER Coils are tuned to receive RF frequency corresponding to the proton precession in a 3 tesla magnetic field, which is governed by the Larmor equation.    To be used in conjunction with Toshiba 3T Magnetic Resonance Scanners with ODU connectors to produce diagnostic images of the upper and lower extremities, chest, abdomen, pelvis, head, neck and spine that can be interpreted by a trained physician.",
      "report_date": "20160810",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "NeoCoil, LLC",
      "recall_number": "Z-2312-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74477",
      "termination_date": "20171121",
      "more_code_info": "",
      "recall_initiation_date": "20160617",
      "postal_code": "53072-6204",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}