  "results": [
    {
      "status": "Terminated",
      "city": "Palm Beach Gardens",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71212",
      "recalling_firm": "The Anspach Effort, Inc.",
      "address_1": "4500 Riverside Dr",
      "address_2": "N/A",
      "postal_code": "33410-4235",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US including AZ, CA, CO, FL, GA, IL, KY, LA, MN, MO, NC, NJ, NY, OH, OR, PA, TX, and Internationally to Canada and China.",
      "recall_number": "Z-2311-2015",
      "product_description": "ANSPACH  - 5MM Coarse Diamond Ball, 22MM Shaft Exposure;  REF S-5DC-22-G1    Cutting shaping bone including spine and cranium.",
      "product_quantity": "80 tools.",
      "reason_for_recall": "G1 Dissection Tools are not compatible with all attachments listed on the labeling;  G1 cutting tools are labeled with a Specified Shaft Exposure which may not be consistent when used with each attachment on the product label. Shaft exposure may be greater.",
      "recall_initiation_date": "20150413",
      "center_classification_date": "20150729",
      "termination_date": "20160825",
      "report_date": "20150805",
      "code_info": "REF S-5DC-22-G1 ; ALL lot numbers."
    }
  ]
}