{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Andover",
      "address_1": "3000 Minuteman Rd",
      "reason_for_recall": "The two bolts that attach the clamping plate to the lateral shaft may break, which could cause the c-arm to move laterally forward, slide out of its bearings, and sag",
      "address_2": "",
      "product_quantity": "993 total",
      "code_info": "All serial numbers.   708034 System serial numbers  6, 264, 112, 106, 233, 136, 31, 83, 224, 144, 71, 123, 5, 249, 134, 103, 266, 97, 132, 61, 19, 82, 196, 3, 12, 39, 244, 197, 119, 204, 75, 151, 69, 70, 37, 115, 55, 200, 180, 175, 78, 93, 234, 130, 142, 53, 113, 65, 27, 242, 202, 14, 208, 247, 261, 225, 28, 11, 110, 25, 128, 171, 81, 237, 211, 54, 76, 163, 207, 216, 122, 135, 205, 63, 238, 254, 262, 72, 1, 206, 187, 52, 143, 154, 155, 156, 73, 150, 98, 248, 80, 22, 133, 209, 167, 265, 105, 217, 79, NTS0427, 129, 13, 15, 16, 17, 18, 24, 84, 146, 41, 190, 10, 194, 125, 268, 177, 109, 42, 43, 44, 172, 213, 181, 212, 189, 26, 95, 9, 4, 117, 251, 2, 48, 183, 157, 245, 116, 62, 56, 160, 152, 127, 94, 174, 176, 259, 50, 58, 51, 277, 68, 159, 77, 40, 199, 253, 114  257, 246, 23, 147, 162, 29, 66, 47, 179, 221, 126, 195, 220, 161, 46, 182, 278, 138, 222, 269, 36, 96, 201, 137, 223, 228, 219, 60, 272, 111, 35, 85, 186, 38, 120, 170, 185, 32, 64, 30, 100, 101, 104, 235, 34, 148, 149, 67, 118, 121, 241, 210, 191, 255, 168, 273, 33, 240, 131, 88, 8, 193.    708037 System Serial numbers  48, 85, 17, 53, 2, 125, 170, 180, 157, 33, 49, 200, 18, 84, 216, 36, 116, 234, 201, 135, 236, 39, 96, 34, 19, 54, 132, 5, 154, 83, 151, 29, 21, 110, 175, 228, 183, 20, 4, 91, 69, 44, 205, 89, 164, 240, 165, 13, 51, 121, 174, 120, 43, 160, 185, 171, 23, 209, 95, 144, 50, 141, 156, 80, 94, 150, 190, 41, 10, 113, 93, 98, 196, 38, 64, 86, 72, 73, 230, 56, 22, 140, 181, 192, 37, 12, 191, 195, 155, 221, 182, 133, 246, 149, 169, 61, 62, 100, 101, 102, 103, 219, 226, 227, 46, 97, 99, 55, 3, 7, 8, 63, 233, 106, 114, 206, 207, 108, 213, 32, 163, 24, 92, 176, 212, 179, 14, 193, 214, 6, 30, 173, 222, 146, 115, 90, 177, 59, 25, 189, 118, 218, 153, 107, 161, 31, 27, 112, 42, 104  76, 26, 217, 35, 124, 139, 15, 66, 241, 245, 40, 239, 123, 58, 88, 172, 129, 81, 136, 159, 244, 238, 82, 65, 249, 70, 142, 188, 178, 158, 128, 117, 148, 45, 203, 79, 229, 199, 119, 126, 152, 71, 52, 60, 197, 198, 87, 105.     708038 System serial numbers 158, 146, 196, 164, 37, 63, 213, 209, 218, 252, 255, 256, 7, 96, 170, 45, 173, 100, 51, 155, 83, 210, 90, 212, 159, 217, 226, 1, 53, 181, 163, 140, 47, 95, 148, 39, 26, 10, 114, 108, 71, 141, 115, 61, 16, 42, 33, 14, 36, 106, 112, 35, 137, 48, 60, 157, 25, 50, 54, 156, 116, 12, 68, 105, 3, 11, 161, 204, 205, 15, 232, 9, 220, 109, 227, 224, 201, 250, 251, 91, 92, 93, 207, 72, 228, 171, 118, 57, 189, 132, 129, 216, 138, 29, 24, 31, 2, 119, 139, 5, 28, 49, 22, 23, 208, 80, 98, 30, 125, 193, 167, 168, 169, 75, 81, 131, 221, 211, 41, 160, 192, 43, 44, 154, 97, 6, 219, 8, 79, 135, 88  107, 65, 194, 126, 69, 247, 248, 249, 13, 215, 78,  243, 245, 246, 175, 38, 177, 253, 191, 32, 259, 235, 264, 258, 52, 67, 111, 257, 94, 142, 143, 260, 121, 236, 268, 265, 74, 101, 182, 174, 124, 239, 4, 172, 263, 254, 149, 151, 64, 183, 184, 76, 152, 188, 261, 242, 244, 34, 162, 147, 241, 134, 178, 237, 176.",
      "center_classification_date": "20160801",
      "distribution_pattern": "Worldwide distribution.  US nationwide, Canada, Algeria, ARGENTINA, Australia, Austria, Belgium, China, Colombia,  Czech Republic,  Denmark, Egypt, Estonia, Faroe Islands, Finland, France, French Guiana, Germany, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Republic of Korea, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, TURKEY, Ukraine, United Arab Emirates, and United Kingdom.",
      "state": "MA",
      "product_description": "MultiDiagnost Eleva with Flat Detector; x-ray system. Product Codes: 708034, 708037, 708038.",
      "report_date": "20160810",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Electronics North  America Corporation",
      "recall_number": "Z-2305-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74731",
      "termination_date": "20201008",
      "more_code_info": "",
      "recall_initiation_date": "20160719",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}