  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Newark",
      "state": "DE",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62709",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "500 GBC Drive, Mailstop 514",
      "address_2": "PO BOX 6101",
      "postal_code": "19714-6101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "USA ( nationwide ) including the states of  AZ, CA, FL, IL, KY, MD, MA, MN, MS, MO, NE, NH, NJ, NY, NC, OH, PA, TX, VA and WV.",
      "recall_number": "Z-2305-2012",
      "product_description": "syngo(R) Lab Data Manager and syngo(R) Lab Process Manager, all versions.    This device is classified as a data calculator and processing module for use with In-vitro diagnostic devices. It also provides work flow management as well as remote instrument control to multiple interfaced instruments.",
      "product_quantity": "44",
      "reason_for_recall": "When using the syngo Quality Control package the system may not perform as intended for Multi-Rule QC violation [e.g., 2(2S), 4(1s) and 10x], unless configured using the instructions on page 2 of the recall notification.  QC status may not be flagged appropriately.  If auto-verification rules are defined to hold patient results when QC status is out, the patient results will not be held as expecte",
      "recall_initiation_date": "20120529",
      "center_classification_date": "20120831",
      "termination_date": "20140410",
      "report_date": "20120912",
      "code_info": "Version VA11B and all previously released versions."
    }
  ]
}