  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85647",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-2302-2020",
      "product_description": "Multi Display Manager in the Artis Systems (Image-Intensified Fluoroscopic X-Ray System)  as follows:    Artis Q floor     Model: 10848280;   Artis Q ceiling  Model: 10848281;  Artis Q biplane  Model:10848282;   Artis zee ceiling Model: 10094137;   Artis zee floor     Model: 10094135     Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.",
      "product_quantity": "15 US",
      "reason_for_recall": "Artis systems containing a Multi Display Manager of a specific lot, the power distributor may lead to an increased electrical contact resistance or the interruption of electrical contact, and result in planned procedures to be terminated and performed on an alternative x-ray system",
      "recall_initiation_date": "20200430",
      "center_classification_date": "20200609",
      "termination_date": "20200911",
      "report_date": "20200617",
      "code_info": "Serial Numbers:  Serial  125253  103340  121526  121521  103335  109621  138311  148418  147237  109620  136555  121520  121525  121518  125254"
    }
  ]
}