  "results": [
    {
      "status": "Terminated",
      "city": "Bridgend",
      "state": "",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74366",
      "recalling_firm": "ORTHO-CLINICAL DIAGNOSTICS",
      "address_1": "FELINDRE MEADOWS",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution - US (Nationwide including Puerto Rico) and foreign distribution to Canada and Bermuda",
      "recall_number": "Z-2302-2016",
      "product_description": "VITROS Immunodiagnostic Products Total T4 Reagent Pack, REF/Product Code  874 4468, Unique Identifier No. 10758750005147, IVD --- Ortho-Clinical Diagnostics,    For in vitro diagnostic use only. For the quantitative measurement of total thyroxine (T4) in human serum and plasma (EDTA or heparin) using the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System to aid in the differential diagnosis of thyroid disease.",
      "product_quantity": "US: 12,015; Foreign: 55 Sales Units to Canada, 223 Sales Units shipped to Bermuda",
      "reason_for_recall": "Ortho-Clinical Diagnostics confirmed that sample results generated from VITROS Immunodiagnostic Products Total T4 Reagent Packs may show a negative bias prior to the open pack storage limit of <= 8 weeks that is located in the Instructions for Use (IFU) for VITROS Immunodiagnostic Products Total T4 Reagent Pack.",
      "recall_initiation_date": "20160602",
      "center_classification_date": "20160729",
      "termination_date": "20180608",
      "report_date": "20160810",
      "code_info": "Lot Numbers:1836, 1840, 1845, 1850, 1860, 1870, 1880, 1890, 1900, 1910, 1920, 1930, 1940, 1950, 1968, 1978, 1988, 1998",
      "more_code_info": ""
    }
  ]
}