  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Best",
      "state": "",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80271",
      "recalling_firm": "Philips Medical Systems Nederlands",
      "address_1": "Veenpluis 4-6",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US nationwide distribution.",
      "recall_number": "Z-2301-2018",
      "product_description": "OmniDiagnost Classic X-ray system Angiographic and Diagnostic, Product Nos.  70859, 708023, 708024, 708025",
      "product_quantity": "477",
      "reason_for_recall": "The fixation of the upper and lower tilt actuator, of the Omnidiagnost system might break off and the table will start to rotate from 0 to + 90 /- 90 degree with high speed. This rotating movement can not be stopped by the user. This could potentially lead to the patient falling off the table. There is a risk of injury for the patient, user, or bystander.",
      "recall_initiation_date": "20180501",
      "center_classification_date": "20180627",
      "termination_date": "20200608",
      "report_date": "20180704",
      "code_info": "As-Maintained System Equipment Number: 76596  38351  76948  50428  82655  13124  85163  250340  86549  87164  86977  86965  86340  1520  101154  1873  102164  102977  83871  83958  86113  104857  103052  103219  84850  86303  103055  103058  103218  104012  104934  104577  105324  504972  504297  104858  529688  527085  527086  527104  527090  527082  529687  527080  527087  529686  527091  527092  527089  527093  529705  527109  527088  527102  527107  527083  527084      529703  529699  529713  529706  529712    529690  529710  529709  529708  529707  14629273  15893969  35015270  35014092  35127787  34997249  15849132  15857294  158554  15841822  15858211  16231503  15831844  16409964  16802355  18055578  18560502  15857902  15858098  303148  15854028  17492237  17541021  31456486  17758213  17941166  19392346  19579758  19782801  18809817  19257910  31103760  19782331  19935548  30396410    32054303  30194039    30873909  37103066  34090526  31596158  10135906  10166050  10006064  16064  10138539  47843126  479951  47911693  47901391  47844078  47944782  48030899  478435  47891928  47912290  48005677  486819  47837945  484516  47884920  47859977  47923515  48059776  48050512  47884354  47850684  48149946  47872003  47938350  47943336  47898047  48060837  478948  50194622  48016151  48062720  47995620  479809  47999720  47844799  47900809  48017211  47882644  47908199  48020563  48026144  48030684  47890048  48055971  47894387  47943186  47937552  48062279  47884829  47872447  47937672  48078137  47913036  479468  47894218  47926963  47889912  47900468  479897  47893535  47893638  47989775  48025019  48066707  48076135  47905960  14430237  48036987  48034012  48018140  47845684  47924690  15314079  47871520  47947079  48023192  47867463  47942597  486893  48016769  47987041  47871768  47934458  47987152  48037814  48055063  48055336  47989269  47945064  47847747  48025542  47902864  47883970  47880733  48022528  48056372  48056424  48056474  47880372  47954220  47843577  47878837  47951539  48048118  47956370  48008155  47915670  47838516  47921591  48057153  48086181  47840577  483428  47937095  47846955  47897860  48019597  48078998  47994839  48019837  47932196  47889837  48055440  47876560  47937175  47994577          47840758  47835296  47918167  48044073  47903895  48053656  485677  47902524  47894944  47922669  55780435  48027852  47896589  48023262  48015942  481403  47963682  48001403  47999799  48011103  48048560  47905125  47845762  479712  47900712  48173118  47999344  480515  47868786  47999900  47886317  48047270  47960473  47888204  48086244  48036906  47995001  53261099  479999  912567  47996686  47838739  488155  47902340  47981406  48027784  47843335  47888783  478723  912554  912570  912559  581026  186984  912572  912573  912564  912558  912569  912563  912561  912551  912560  912552  912574  912575  912562  912556  36920957  912568  912566  912576  912565  912577  912555  912550  912553  193444  91002561  35411240  149457  912557  912571  48155079  42612895  51040917  51029767  51047402  50937663  51028970  50988581  50976660  51103727  50901797      50969468  50978845  51020690    50968278    51072875  50997325    51112717  51069644  50977659  50994437  51029062  51029673  50964975  50951499  51040304  51112897  15868176  15867868  48152770    42679080  42680261  162624  186454        47952304  47952194  16216632  10188638  10123241  36962750  36970721  48114348  48156059  48142307  48114234  48114558  75795774  75795892  57742137  18045135  37203863  37283054  37271725  37194973  37178219  37192552  37197676  37126660  37126769  37218356    37134483          37270177  37121779  37112241  37228570  37192635  37129310  37275288  37245565  37210949  33317303  10119343  10146416  141793  15983788  14430350  15313955  15314202  34767769  34750597  49888651  34764332  34764834  47352569  34764731  34764966  54375348  36274326  36348140  36358901  36425595  36196325  36247890  36404575  36306923  36371177  36347422  362891  36321062  36334792    36287232  36384474  36396773  36209250  362623  36318737  36314659",
      "more_code_info": ""
    }
  ]
}