  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "1800 W Center St",
      "reason_for_recall": "Zimmer had received complaints that reported breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). This is a retrospective recall which covers Zimmer's corrective actions taken earlier this year concerning the TASPs.",
      "address_2": "",
      "product_quantity": "70,986 distribution events",
      "code_info": "Item Numbers:  42517600410  42517600510  42517600610  42517600710  42517600810  42517600910      Lot Numbers:   62342526  62423918  62423919  62423920  62427414  62482212  62563745  62612225  62656072  62740698  62342527  62405895  62430401  62504179  62511925  62563784  62638665  62695870  62343066  62406967  62432146  62515486  62564324  62691390  62717335  62351291  62374838  62415307  62430397  62518138  62622527  62696875  62357867  62374796  62427836  62441229  62511924  62565948  62563908  62590523  62695295  62361315  62376662  62420192  62434748  62518137  62540059  62563778  62627131  62698733",
      "center_classification_date": "20140821",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of  AK, AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WY, and the countries of Australia, Canada, Germany, India, Japan, Korea, and Singapore.",
      "state": "IN",
      "product_description": "Persona (TM) The Personalized  Knee System  CPS Tibial Articular Surface Provisional Top  Nonsterile  TASP MLC L 3-5 CD TOP  TASP MLC L 6-9 CD TOP  TASP MLC L 3-5 EF TOP  TASP MLC L 6-9 EF TOP  TASP MLC L 10-11 EF TOP  TASP MLC L 6-9 GH TOP",
      "report_date": "20140827",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer, Inc.",
      "recall_number": "Z-2301-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68817",
      "termination_date": "20140911",
      "more_code_info": "",
      "recall_initiation_date": "20130610",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}