      "status": "Terminated",
      "city": "Irvine",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85424",
      "recalling_firm": "INARI MEDICAL",
      "address_1": "9 Parker Ste 100",
      "address_2": "N/A",
      "postal_code": "92618-1666",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S Distribution in the states of .: IA, GA, MO, MI, AZ, FL, NJ, MN, LA, PA, AL, OH, NY, WI, CA, TX, TN, NH, IL, NC, CT, CO, OK, OR, IN, MA, VA, KY, WA, MS, SC, UT, SD, ND, AR, ID, DE, including DC.",
      "recall_number": "Z-2299-2020",
      "product_description": "Triever20 Aspiration Guide Catheter a component of FlowTriever Retrieval/ Aspiration System, UDI: 00850291007079",
      "product_quantity": "318",
      "reason_for_recall": "The hemostasis valve of aspiration guide catheter devices may fail to seal resulting in a loss of vacuum during thrombus aspiration, which may result the procedure being prolonged as another device must be used or suboptimal thrombus removal if the problem is not detected.",
      "recall_initiation_date": "20200323",
      "center_classification_date": "20200608",
      "termination_date": "20210813",
      "report_date": "20200617",
      "code_info": "Lot: 20010004, 20010005, 20010008"
    }
  ]
}