  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Crawley",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85600",
      "recalling_firm": "Elekta Limited",
      "address_1": "Linac House",
      "address_2": "Fleming Way",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US distribution to IA, NY, PA, TX, WI.      International distribution to Australia, Canada, China, Denmark, Germany, Hong Kong, Italy, Netherlands, Sweden, Switzerland, Turkey, United Kingdom.",
      "recall_number": "Z-2297-2020",
      "product_description": "Elekta Unity, Philips Marlin Software versions: R5.3.30, R5.3.31, R6.0.5331, R5.7.0, R5.7.1, R6.1.571",
      "product_quantity": "26 units",
      "reason_for_recall": "In TSM (Treatment Session Manager) Motion Monitoring workflows, under certain conditions there can be a mismatch between the contour data overlay with respect to the motion monitoring images of the monitored structure.",
      "recall_initiation_date": "20200521",
      "center_classification_date": "20200608",
      "termination_date": "20240819",
      "report_date": "20200617",
      "code_info": "Software versions: R5.3.30, R5.3.31, R6.0.5331, R5.7.0, R5.7.1, R6.1.571.  UDI Number 05060191071321.      Anatomy (Protocol Name): Abdomen (btFFE Cor RealTime, btFFE Cor Sag RealTime, btFFE Cor Sag Tra RealTime, btFFE Fast Sag Cor RealTime, btFFE Sag RealTime); HeadNeck L (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime);    HeadNeck M (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime);  Pelvis L (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime);  Pelvis M (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime); Thorax (btFFE Cor RealTime, btFFE Cor Sag RealTime,   btFFE Cor Sag Tra RealTime, btFFE Sag RealTime, btFFE Sag RealTime)"
    }
  ]
}