  "results": [
    {
      "country": "United States",
      "city": "Mahwah",
      "address_1": "325 Corporate Dr",
      "reason_for_recall": "The recalling firm has received a higher than expected number of complaints documenting femoral head/hip stem disassociation for certain sizes of femoral heads manufactured prior to March 4, 2011.",
      "address_2": "",
      "product_quantity": "",
      "code_info": "Lots  201MJD  20RMJD  20TMED  20XMLE  211MJD  217MKD  21AMKD  21DMKD  21KMJD  21LMKD  21LMKDX1  21VMKD  21WMKD  21WMKDX1  21XMKD  21YMKD  223MAE  234MKE  241MJD  26XMHE  26YMHE  270MHE  280MAE  28TMTD  297MTD  2DDMKE  2DJMJD  2HKMKE  2HNMKE  2JPMKE  2JVMKE  2KKMJD  2MMMAE  2MPMAE  2MRMAE  2MTMAE  2NAMKE  2VDMAE  2VMMAE  2W6MAE  2W6MJD  2W7MJD  2W8MAE  2WAMJD  2WPMHE  2X5MHE  2XTMAE  2XVMAE  301MHE  30PMAE  30XMJD  30YMJD  315MJD  31NMLE  31PMLE  31RMLE  31TMLE  338MAE  33PMED  33RMED  33TMED  33VMED  33WMED  33XMED  343MED  344MED  345MED  361MRA  361MRAX1  362MRA  8E3MMA  8E3MMAX1  8E4MMA  8E6MMA  8E8MMA  8E8MMAX1  8E9MMA  8ERMMA  8H8MPD  8J8MMA  8L6MPD  8NJMMA  8NJMMAX1  8PLMRA  8PMMRA  8PNMRA  8PNMRAX1  8PTMPD  8PXMPD  8T3MMA  8VPMMA  91AMPE  925MLD  926MLD  934MMA  934MMAX1  93DMMA  953MPA  954MPA  954MPAX1  976MMA  977MMA  978MMA  979MMA  97AMMA  97DMMA  97PMTA  97PMTAX1  97RMTA  97TMTA  97VMTA  97WMTA  99EMPE  99HMPE  9ADMME  9AEMME  9AHMME  9AJMME  9AVMPE  9D5MPE  9DMMPE  9DNMMA  9DPMMA  9DPMMAX1  9DRMPE  9E8MPE  9E9MMA  9E9MMAX1  9E9MMAX1X1  9EEMMA  9EHMMA  9EHMMAX1  9H7MPE  9H9MPE  9L7MPE  9LPMPE  9TJMJD  9TJMJDX1  9TJMJDX2  9V8MJD  JH9MJE  JHAMJE  JHDMJE  JHEMJE  JHNMPD  JHRMED  JHRMEDX1  JHVMAE  JHWMAE  JHXMAE  JHYMAE  JJ9MLE  JKNMDD  JKNMDDX1  JKPMAE  JKRMAE  JLKMDD  JLLMDD  JLMMDD  JLPMDD  JLRMDD  JLRMDDX1  JLTMDD  JLTMDDX1  JLVMDD  JLVMDDX1  JMXMHE  JN8MPD  JN9MPD  JNXMED  JPRMPD  JT2MED  JTTMED  JVYMED  JWXMME  JWYMED  JWYMEDX1  JX0MME  JX1MME  JX2MME  JX3MED  JX4MED  JY5MKD  JY9MPD  K0JMED  K2AMHD  K5MMJE  K66MHD  K68MHD  K68MHDX1  K78MED  K78MEDX1  K7AMED  K7DMED  K7EMED  K7EMJE  K7HMED  K7HMEDX1  K7HMJE  K7JMJE  K7KMJE  K8WMHD  K8XMHD  K8XMHDX1  K8YMHD  K8YMHDX1  K90MHD  K91MHD  K91MHDX1  KA1MLE  MER718  MERDEP  MERDER  MERDH8  MERDH9  MERDHA  MERDHD  MERDHE  MERDHEX1  MERDHH  MERE2V  MERED2  MEREP5  MEREP6  MERHR9  MERHVX  MERK77  MERK79  MERK7A  MET56M  MET56MX1  MET5JX  MET5JXX1  MET5JY  MET5K0  MET635  MET636  MET637  MET638  MET7DK  MET7DKX1  MET7K0  MET7K8  MET7KH  MET7KM  MET7L9  MET7LJ  MET7LM  MET7LW  MET9Y2  MET9Y3  MET9Y4  METAE4  METAML  METARE  METATX  METATXX1  METATY  METATYX1  METAV0  METAVY  METAW5  METAWH  METAYE  METD13  METD13X1  METD20  METD33  METD37  METE6J  METE6T  METE81  METE82  METE83  METE84  METJMA  METJMD  METJME  METJMEX1  METJMH  MHEX2A  MHEX30  MHEX3K  MHEX3L  MHEXD7  MHEXJJ  MHEXK3  MHEXNA  MHEXVV  MHEY78  MHEY8R  MHEY8W  MHEY8WX1  MHEYDJ  MHH0D3  MHH18L  MHH1AK  MHH1L4  MHH1VR  MHH1VRX1  MHH1VRX2  MHH1VT  MHH565  MHH5HL  MHH5HLX1  MHH5HM  MHH5VW  MHH629  MHH65D  MHH65T  MHHHKW  MHHJ1N  MHHJ56  MHHJYA  MHHK3L  MHHK5N  MHHKH8  MHHKM3  MHHKM8  MHHKML  MHHKN8  MHHKT2  MHHKVW  MHHKYP  MHHL1R  MHHL61  MHHL7K  MHHL8T  MHHMPD  MHHPPX  MHHPXW  MHHPYA  MHHPYL  MHHR14  MHHR45  MHHR46  MHHR47  MHHRK6  MHHRK8  MHHRKA  MHHRR5  MHHRT2  MHHRTX  MHHRV6  MHHRWV  MHHT44  MHHT45  MHHT46  MHHTH5  MHHTH6  MHME90  MHME96  MHMEM2  MHMEM3  MHMEM5  MHMEM6  MHMH4K  MHMH4L  MHMH5P  MHMH5R  MHMH5RX1  MHMH72  MHMH8A  MHMH8E  MHMH8H  MHMH8J  MHMHHJ  MHMHHK  MHMHHL  MHMHHM  MHMHJY  MHMJ79  MHMJ7E  MHMJ8K  MHMJ8M  MHMJ93  MHMJ94  MHMJ9N  MHMJA8  MHMJA9  MHMJAA  MHMJAD  MHMJEJ  MHMJEJX1  MHMJEK  MHMJEM  MHMJEN  MHMM6M  MHMN4J  MHMNEX  MHMNM2  MHN5A4  MHN5DD  MHN5DR  MHN5E1  MHN5EM  MHN5HH  MHN5HJ  MHN5HK  MHN5HL  MHN5NR  MHN5NT  MHN5NV  MHN5NW  MHN61R  MHN6PA  MHN6PT  MHN6R3  MHN6R5  MHN73A  MHN73D  MHN73E  MHN73J  MHN7EL  MHN7EP  MHN7ER  MHN7ET  MHN7YJ  MHN7YK  MHN7YW  MHT9EW  MHT9EX  MHT9EY  MHT9H0  MHT9H2  MHTA8N  MHTA8P  MHTA8V  MHTA8W  MHTA9K  MHTA9L  MHTAA2  MHTAA3  MHTAA4  MHTAA5  MHTDP1  MHTDX4  MHTE75  MHTE94  MHTEE3  MHTEE8  MHTEHJ  MHTEKN  MHTEPK  MHTEX0  MHTEX5  MHTEYY  MHTH0L  MHTH0M  MHTH16  MJ5MKE  MJA521  MJA522  MJA523  MJA524  MJA525  MJA526  MJA527  MJA528  MJA529  MJA63N  MJA63P  MJA645  MJA646  MJA69R  MJA69T  MJA69V  MJA69W  MJA6DV  MJA6DW  MJA6DX  MJA6DY  MJA6M5  MJA6M6  MJA6M7  MJA6M8  MJA6M9  MJAT16  MJAT17  MJAT18  MJAT19  MJAT8N  MJATDP  MJATE5  MJAV2V  MJAV2W  MJAV2X  MJAV2Y  MJAVJE  MJAW06  MJK923  MJK9EW  MJK9EY  MJK9H0  MJK9H1  MJK9NP  MJK9NR  MJK9R8  MJK9R9  MJK9RA  MJKA07  MJKA08  MJKA09  MJKAM9  MJKAMA  MJKAVX  MJKAW0  MJKAW1  MJKAW2  MJKDNT  MJKDNV  MJKDNW  MJKDNX  MJKHRW  MJKHRX  MJKHV7  MJKHW1  MJKHX1  MJKJ06  MJKJ31  MJKJ64  MJKJ88  MJKJEE  MJKJPL  MJKJPM  MJKJPN  MJKJPR  MJKJT3  MJKJV6  MJKJV7  MJKJV9  MJKJVA  MJKJX5  MJKK1P  MJKK31  MJKK3X  MJKK80  MJKKAK  MJKKHW  MJKKK9  MJKKKR  MJKKKT  MJKM08  MJKM09  MJKM0A  MJKM0E  MJKMEK  MJKMEL  MJKMEM  MJKMEN  MJKMPP  MJKMPT  MJKMPW  MJKMR2  MJL0X5  MJL15K  MJL1AR  MJL2T9  MJL33N  MJL36J  MJNTH3  MJNTY5  MJNVTK  MJNVWJ  MJNWEV  MJNX70  MJNXPN  MJNXTJ  MJNXYJ  MJNY53  MJNY81  MJP06M  MJP088  MJP0H9  MJP0N0  MJP0PD  MJP0PY  MJP19E  MJP1D0  MJP24V  MJP25P  MJP2P6  MJP2T8  MJP3JL  MJP3NA  MJP44K  MJP59V  MJP5HY  MJP5MD  MJP5RT  MJP5V1  MJP6HL  MJP71D  MJP76M  MJP7LR  MJP7W7  MJP7WA  MJP7XM  MJP80T  MJP819  MJP81E  MJP8L3  MJP8M5  MJP8N2  MJP8TL  MJP9EN  MJP9KY  MJP9P2  MJPA2A  MJPA62  MJPATD  MJPATV  MJPD1L  MJPDM0  MJPDRL  MJPE3N  MJPE67  MJPE98  MJPED5  MJPELA  MJPHN4  MJPHRA  MJPJ5Y  MJPJ6X  MJPJ94  MJPJRJ  MJPK8L  MJPKV2  MJPLMT  MJPLV7  MKAXV6  MKAY86  MKAYD3  MKD00V  MKD066  MKD0WA  MKD1JY  MKD1N6  MKD20E  MKD226  MKD299  MKD2R7  MKD3NV  MKD3PX  MKD3T9  MKD48X  MKD4J8  MKD4PM  MKD56A  MKD58W  MKD5DV  MKD5JE  MKD5LK  MKD6E2  MKD6E5  MKD709  MKD71T  MKD8A2  MKD8M5  MKD8MW  MKD8P9  MKD8X8  MKD9D4  MKD9KN  MKDA48  MKDA5P  MKDA5Y  MKDALD  MKDALY  MKDANV  MKDAP4  MKDAYJ  MKDJP8  MKDKMM  MKDL3D  MKDL3E  MKDL64  MKDMLE  MKDN51  MKDN5V  MKDPH1  MKDPH2  MKDPXY  MKDRAJ  MKDT49  MKDT6T  MKDTRP  MKDWX1  MKDX3P  MKDYAV  MKE15P  MKE18V  MKE2A0  MKE2L2  MKE3E4  MKE3R4  MKE43X  MKE456  MKE54L  MKE7LY  W0DMNE  W0KMNE  W44MKD  W44MKDX1  W49MJE  W4WMKA  W58MNE  W7DMJE  W82MJE  W85MJE  W8LMJE  WAHMAD  WAJMAD  WARMAD  WARMADX1  WATMAD  WAVMAD  WAWMAD  WAXMAD  WD0MAD  WD0MHE  WD1MAD  WD1MHE  WD2MAD  WD2MHE  WD3MAD  WD3MADX1  WD3MHE  WD5MAD  WD9MAD  WDAMAD  WDDMAD  WDWMAD  WHJMKD  WJNMME  WJTMME  WJWMME  WJYMME  WL9MNE  WM4MKA  WM4MKAX1  WN8MNE  WPEMKD  WVAMKA  WVDMKA  WVDMKAX1  WVEMKA  WVHMKA  WVHMKAX1  WW6MME  WWPMME  WWXMKD  WWXMKDX1  WWXMKDX1X1  WWYMKD  WX4MEE  WX6MEE  WX7MEE  WX8MEE  X2KMKD  X2MMKD  X3HMNE  X47MKD  X4EMKD  X4MMKD  X4MMKDX1  X4NMJE  X52MKD  X5LMKD  X5NMHE  362MRAX1  363MRA  363MRAX1  364MRA  36AMAE  39PMAE  3AVMPA  3AVMPAX1  3AWMPA  3DYMAE  3L4MHE  3LYMEE  3M0MEE  3M1MEE  3M2MEE  3PNMLE  3PPMLE  3TLMHE  3V8MLE  3V8MLEX1  3VDMLE  3WLMHE  415MLE  44MMHE  484MPD  48DMRD  48YMPD  490MPD  4DNMED  4DPMED  4DRMED  4DTMED  4DTMEDX1  4DTMRD  4DVMED  4DYMPD  4E0MPD  4E1MPD  4H8MEE  4KAMRD  4NRMRD  4RKMTD  4RMMTD  4RPMTD  4RTMTD  4RVMTD  4VMMRD  4W7MAE  4W8MAE  4W9MAE  4WAMAE  4WEMEE  4WHMEE  4WJMEE  4WKMEE  4WPMRD  4YVMME  542MAE  552MTA  557MTA  55AMTA  55AMTAX1  55DMTA  55MMAD  55PMAD  55RMAD  55TMAD  55VMAD  55WMAD  56MMRD  9V8MJDX1  9V8MJDX1X1  9X8MPA  9X9MPA  9XAMPA  A1HMME  A1KMME  A1LMEE  A1LMME  A1MMEE  A1MMME  A1NMEE  A1PMEE  A2AMMD  A3EMPA  A3LMPA  A43MMD  A49MEE  A58MEE  A5MMME  A5NMLD  A5NMLDX1  A5RMLD  A6VMEE  A96MEE  AD2MJD  AH6MJD  AH6MJDX1  AJVMJE  AMPMHE  AN0MHE  ARHMJD  ARHMJDX1  ARHMJDX1X1  AT0MJD  AT5MJD  AT5MJDX1  AT6MKD  AT6MKDX1  AT7MKD  AT9MJD  AT9MJDX1  ATKMJD  AW2MKD  AW3MKD  AWYMKD  AX0MKD  AX8MKD  D1MMHE  D1NMHE  D37MLD  D3DMHE  D3DMLD  D3EMLD  D3EMLDX1  D3NMLD  D71MND  D72MND  D73MND  D9KMEE  D9LMEE  D9MMEE  D9NMEE  DHRMHD  DMAMND  DMDMND  DMEMND  DN0MDE  DN1MDE  DN2MDE  KAAMHD  KAXMHD  KD0MHD  KJKMLE  KK6MLE  KKLMLE  KKRMKD  KKXMKD  KL0MKD  KMHMKD  KMJMKD  KMKMKD  KPKMKD  KPLMKD  KPNMKD  KPPMKD  KR5MPA  KR8MPA  KR8MPAX1  KR9MPA  KRAMPA  KV6MHE  KWEMHE  KX5MHE  KY3MNA  KY4MNA  KY4MNAX1  KY5MNA  KY5MNAX1  KY8MNA  KY9MNA  KY9MNAX1  KYPMPA  KYRMPA  L01MJE  L02MPA  L1DMHE  L1RMNA  L1RMNAX1  L1TMNA  L2JMLE  L2PMNA  L2RMNA  L2TMNA  L2VMNA  L2WMNA  L4WMHE  L64MHE  L69MJE  L6AMJE  L6DMJE  L6EMJE  L73MND  L7KMND  L7YMND  L85MND  L8DMND  L96MNA  L96MNAX1  L9VMTA  L9VMTAX1  LA0MTA  LA1MTA  LA8MTA  LAEMTA  LAJMPA  LAJMPAX1  LD2MPA  LD2MPAX1  LEAMRA  METJMJ  METJMK  METJML  METJMM  METKJL  METKK3  METKK4  METKK8  METKK9  METKLM  METKMD  METLL0  METM60  METM6X  METMJH  MHA68V  MHA69X  MHA6KP  MHA6KR  MHA71J  MHA71JX1  MHA71X  MHA72L  MHA74X  MHA751  MHA752  MHA753  MHA9DV  MHA9DX  MHA9DY  MHA9E1  MHA9E3  MHA9EE  MHA9EJ  MHA9EP  MHA9EW  MHA9H0  MHA9H0X1  MHAKK6  MHAKVK  MHAKW2  MHAL00  MHAL01  MHAL3D  MHAL76  MHAL78  MHAL79  MHALLY  MHALMD  MHAN2R  MHAN2T  MHAN43  MHAN4Y  MHAN52  MHAN5M  MHAN5T  MHAN6K  MHAN6KX1  MHAN94  MHAN9N  MHAN9P  MHANA9  MHANAE  MHDJ2J  MHDMM5  MHDMM5X1  MHDMM6  MHDMM7  MHDMM8  MHDMMR  MHHTLY  MHHTNK  MHHTR7  MHHTT3  MHHTT7  MHHTVV  MHHTWD  MHHTWE  MHHTWEX1  MHJ0T3  MHJ0V5  MHJ0V6  MHJ0VY  MHJ0WN  MHJ589  MHJ596  MHJ59V  MHJ5AR  MHJ5M3  MHJ5R8  MHJ5V9  MHJ61H  MHJ655  MHJ6DD  MHJ6E8  MHJ6E9  MHJ6EY  MHJ6H3  MHJ6HP  MHJ6KN  MHJMKE  MHJRDT  MHJRDW  MHJRDX  MHJRW4  MHJT19  MHJT25  MHK1Y1  MHK219  MHK22E  MHK22J  MHK252  MHK253  MHK2EY  MHK2LY  MHK2ML  MHK2N2  MHK2N4  MHK2NA  MHK2ND  MHK2P4  MHK2VW  MHK2VX  MHK2VY  MHK35P  MHK35R  MHK3AD  MHK3EP  MHK3YY  MHK404  MHK4D1  MHK4D6  MHK4X0  MHK545  MHK54L  MHK54N  MHK5AA  MHK5E5  MHK5J4  MHK5JT  MHN7YY  MHN85V  MHN85W  MHN85X  MHN85Y  MHN8K8  MHN923  MHN925  MHN926  MHN93A  MHNH8K  MHNHA9  MHNHAA  MHNHAD  MHNHHH  MHNHHK  MHNHHL  MHNHHM  MHNHVM  MHNHVN  MHNJ6X  MHNJ6Y  MHNJ70  MHNJ72  MHNJ8R  MHNJ8T  MHNJAW  MHNJAX  MHNJAY  MHNJD0  MHNJJ3  MHNJJ4  MHNJJ5  MHNJJ6  MHNK9R  MHNKE9  MHNKEA  MHNKHV  MHNMKE  MHNN7E  MHNN8D  MHNN8E  MHNN8H  MHNN8J  MHNPM9  MHNPMA  MHNPPM  MHNPPP  MHNR3E  MHNR3H  MHNRE7  MHNRE8  MHNRE9  MHNREA  MHNRNH  MHNRNJ  MHNRNK  MHNRNL  MHNTE9  MHNTEE  MHNTMD  MHNTMN  MHNTMW  MHNTNT  MHNTPM  MHNTPN  MHNTPP  MHNTPR  MHNTPX  MHNV2D  MJAW35  MJAW42  MJAW4T  MJD1JK  MJD1MH  MJD1NV  MJD1RK  MJD1TT  MJD1X7  MJD22A  MJD2E3  MJD2E4  MJD2EM  MJD2EN  MJD2H6  MJD2H7  MJD2J1  MJD2J2  MJD2JD  MJD2JL  MJD2KP  MJD2NN  MJD2R4  MJD2X5  MJD2Y4  MJD31P  MJD3M3  MJD3M5  MJD3MH  MJD3N0  MJD3PW  MJD3R4  MJD3TV  MJD3XA  MJD410  MJD453  MJD47X  MJD48N  MJD4V9  MJD4VA  MJD4VR  MJD4WP  MJD4Y3  MJD50X  MJD52E  MJD54M  MJD579  MJD59J  MJD5EM  MJD5KP  MJD68Y  MJD6D9  MJD6NL  MJD6TJ  MJD6WT  MJD76A  MJD7M3  MJD7M4  MJD7M5  MJD7P5  MJDHAH  MJDHAJ  MJDHAK  MJDHAL  MJDHJ8  MJDHM4  MJDHRN  MJDHXH  MJDHY6  MJDJ1N  MJL39T  MJL3HA  MJL3JN  MJL3T2  MJL48X  MJL4L2  MJL4N5  MJL5H0  MJL5J0  MJL5VY  MJL73X  MJL748  MJL7EX  MJL7WE  MJL7XJ  MJL9E1  MJL9E2  MJL9E3  MJL9PV  MJL9T2  MJLA8K  MJLA8L  MJLA8M  MJLA9H  MJLAA1  MJLAA2  MJLAA3  MJLAA4  MJLAHN  MJLALN  MJLALP  MJLALR  MJLALT  MJLAY0  MJLD67  MJLEK6  MJLELP  MJLEXT  MJLHRM  MJLHVT  MJLJDD  MJLK78  MJLK7Y  MJLKJ9  MJLKLM  MJLKTD  MJLL02  MJLM2R  MJLM5W  MJLM6P  MJLM8E  MJLN6J  MJLNMA  MJLNPE  MJLNXW  MJLP14  MJLPK4  MJLR8A  MJLRDA  MJLRY7  MJLTK1  MJLV53  MJLV6T  MJLV83  MJLVRA  MJLVVD  MJLVX7  MJLW8D  MJLWD6  MJLWJD  MJPMLD  MJPN5J  MJPNHJ  MJPNKJ  MJPNNK  MJPP1N  MJPP37  MJPP7N  MJPPDW  MJPR57  MJPR8H  MJPR9Y  MJPRD8  MJPT5R  MJPT8W  MJPV99  MJPVE4  MJPVK5  MJPVV3  MJPVYD  MJPWAW  MJPWD5  MJPWEE  MJPWK6  MJPWKX  MJPWL3  MJPX56  MJPXAM  MJPY34  MJPY7X  MJPY9P  MJPYWP  MJR05N  MJR09T  MJR0EE  MJR1XH  MJR5N0  MJR5PT  MJR62W  MJR6R9  MJR6V6  MJR7L3  MJR7R1  MJR9E2  MJR9HA  MJR9K6  MJR9W7  MJRAEN  MJRAJ2  MJRDHT  MJRE79  MJRHHL  MJRHT8  MJRJ0H  MJRJ51  MJRJ8N  MJRJ8P  MJRL4D  MJRLJJ  MJRM11  MJRM4E  MJRN1X  MJRN5L  MJRN8D  MJRNHJ  MJRPK3  MJRPV0  MJRR4J  MJRT1A  MJRTNM  MKE8R8  MKE8TL  MKE9V2  MKEAT1  MKEPP4  MKER2J  MKERTM  MKERTN  MKEWEK  MKJMED  MR4MED  MR4MEDX1  MV2MMA  MV3MMA  MV4MMA  MV4MMAX1  MV4MMAX1X1  MV6MMA  MVKMMA  MVKMMAX1  MVRMDE  MWKMEE  MX4MEE  MX5MED  MX5MEDX1  MXVMMA  MXVMMAX1  N32MEE  N35MHE  N58MHE  N5XMMA  N61MEE  N85MHE  N9NMHE  NE1MDE  NE2MDE  NE3MDE  NJJMEE  NL1MAE  NLKMAE  NLXMEE  NM1MDE  NPPMAE  NT3MEE  NT4MEE  NTPMDE  NTRMDE  NWAMDE  P1YMKD  P1YMKDX1  P46MKD  P46MKDX1  P48MKD  P48MKDX1  P5MMAE  P8XMDD  P8YMDD  P90MDD  P91MDD  P9DMKD  PHNMDE  PNWMAE  PPYMMA  PPYMMAX2  PR0MMA  PR1MMA  PR1MMAX1  PRLMAE  PWTMJE  PWVMAE  X63MNE  X64MKD  X64MNE  X65MNE  XAVMKD  XENMHE  XENMHEX1  XEPMHE  XERMHE  XETMHE  XHWMJE  XHXMJE  XHYMJE  XJ0MJE  XLXMJE  XLYMJE  XM0MJE  XM1MJE  XYXMKD  Y4RMJE  Y6YMKD  Y7PMKD  Y8TMJE  YAEMJE  YDYMKD  YE0MKD  YH6MJE  YHYMKD  YM8MJE  YMNMJE  YMPMJE  00NMLE  00PMLE  00TMLE  00VMLE  033MRA  033MRAX1  034MRA  036MRA  037MRA  03WMEE  05JMEE  0DXMAE  0E0MAE  0KAMLE  0KDMLE  0M5MEE  0MWMPAX1  0MWMPAX1X1  0T4MKD  0T4MKDX1  0T5MKD  0T5MKDX1  0T6MKD  0T7MKD  0T8MKD  57EMPD  57JMPD  57TMTA  57XMME  57YMME  580MME  580MTA  580MTAX1  584MTA  584MTAX1  5AEMPA  5AEMPAX1  5AHMPA  5AJMPA  5AJMPAX1  5PTMRD  5THMRD  5WAMRD  5WTMNE  5YEMNE  60HMNE  60WMNE  60XMNE  618MNE  61DMNE  625MNE  64DMAD  64EMAD  64HMAD  64JMAD  64KMAD  64LMAD  69KMNE  6AAMTD  6ALMTD  6E0MND  6E1MND  6E2MND  6E3MND  6E5MND  6E6MND  6H7MND  6JAMNE  6KTMRD  6LLMNE  6NXMMD  6NYMMD  6P0MMD  6P1MMD  6TVMMD  6V7MMD  6W5MLD  6W6MLD  6XXMMD  71NMMD  753MTD  76JMAD  76KMAD  76LMAD  76MMAD  76NMAD  76PMAD  76PMADX1  770MTD  774MTD  778MRD  779MRD  77RMNE  77XMTA  77YMTA  DN3MDE  DPNMHE  DPPMHE  DPRMHE  DR2MHD  DRWMJE  DRWMJEX1  DV5MHD  DVVMJE  DX2MJE  DX9MHD  DYRMHD  E14MHD  E17MHD  E18MHD  E23MHD  E23MHDX1  E23MJE  E23MJEX1  E24MHD  E25MHD  E2VMHD  E2YMHD  E5DMRD  E5EMRD  E73MRD  E74MRD  E7EMNE  E7JMJE  E83MNE  E85MNE  E8WMNE  E90MJE  E9EMND  E9HMND  E9JMND  E9KMND  E9LMND  E9MMND  EA2MEE  EA3MEE  EA4MEE  EA5MEE  EA8MTD  EA9MTD  EAAMTD  ED9MME  ED9MTD  EDAMTD  EDDMTD  EE2MME  EH3MEE  EH4MEE  EH5MEE  EH6MEE  EJ3MRD  EK7MMA  EM8MNE  EM9MNE  EMAMNE  EMDMNE  EN1MLD  EN2MLD  EN4MLD  EV0MPA  EV1MPA  EV6MPA  EV7MPA  H07MLE  H08MLE  LEKMRA  LEVMNA  LEWMNA  LJ4MRAX1  LK7MLE  LLVMLE  LNEMLE  LNPMLE  LRYMRA  LT9MDD  LT9MDDX1  LTAMDD  LTDMDD  LX9MRA  LXXMNA  LXXMNAX1  LXYMNA  LXYMNAX1  LY0MKD  LY0MNA  LY1MNA  M2PMRA  M2PMRAX1  M69MMA  M6DMRA  M6DMRAX1  M6EMRA  M82MED  M82MKE  M9VMED  MA0MPD  MA1MPD  MA2MPD  MA2MPDX1  MANMDD  MAVMDD  MD3MDD  MD6MKE  MD7MKE  MDAMPD  MDKMPD  MDTMKE  MDXMKE  ME5MNA  ME5MNAX1  MELMDE  MEMMDE  MENMDE  MEP9NA  MEP9T2  MEP9WW  MEP9YH  MEPA3R  MEPAET  MEPAETX1  MEPAP4  MEPATJ  MEPATK  MEPAV1  MEPAWD  MEPAWH  MEPD8K  MEPD8P  MEPDEA  MEPDEAX1  MEPH79  MEPH7A  MEPH82  MEPH84  MEPH8T  MHDMMT  MHDMMV  MHDMMW  MHDNN7  MHDNN8  MHDNN9  MHDNNA  MHDR12  MHDR7Y  MHDR91  MHDRVK  MHDRVL  MHDRVM  MHDT73  MHDT7A  MHDT7AX1  MHDT7D  MHDT82  MHDT82X1  MHDT8E  MHDT8L  MHDT8V  MHDT8VX1  MHDT93  MHDT9W  MHDTDN  MHDTDP  MHDTDPX1  MHDTDR  MHDTDT  MHDTDTX1  MHDTPH  MHDTPJ  MHDTPK  MHDTPL  MHDV03  MHDV0Y  MHDV10  MHDV8R  MHDVE3  MHDVE4  MHDVE5  MHDVEW  MHDVJ0  MHDVJ1  MHDVL3  MHDVL7  MHDVL8  MHDVLD  MHDVM0  MHDVM4  MHDVM6  MHDVRK  MHDVRL  MHDVRM  MHDVRN  MHDVRR  MHDVRV  MHDWWE  MHDWX1  MHDWXH  MHDWXT  MHDY0M  MHE95E  MHEK1N  MHEK1P  MHEK3M  MHEK3MX1  MHEK61  MHEK61X1  MHK67X  MHK680  MHK681  MHK6TE  MHK704  MHK70Y  MHK710  MHK713  MHK718  MHK71L  MHK769  MHK78X  MHK78Y  MHK7K4  MHK7V7  MHK7V8  MHK7WH  MHK7WJ  MHK7XH  MHK7XJ  MHL15N  MHL15W  MHL15X  MHL15Y  MHL160  MHL161  MHL22L  MHL22M  MHL22N  MHL22P  MHL22R  MHL22T  MHL22V  MHL22W  MHL22X  MHL39N  MHL3D2  MHL3M4  MHL3Y7  MHL404  MHL40T  MHL412  MHL41M  MHL46L  MHL46M  MHL46N  MHL46R  MHL46T  MHL46W  MHL59A  MHL59E  MHL59J  MHL5ND  MHL5W4  MHL5X1  MHL614  MHL636  MHL63K  MHL63V  MHL63Y  MHL64V  MHL67R  MHL694  MHL6AW  MHL6DR  MHL6EN  MHL6TD  MHL6WH  MHL707  MHL709  MHNV4H  MHNV4J  MHNV4K  MHNV4L  MHPAPY  MHPAR0  MHPAR1  MHPDDX  MHPDDY  MHPDE0  MHPDE1  MHPE6H  MHPE6J  MHPE6K  MHPE6L  MHPEKY  MHPEL1  MHPEL2  MHPEL3  MHPH4T  MHPH4V  MHPH4W  MHPH4X  MHPHDJ  MHPHDN  MHPHDT  MHPHDX  MHPHDY  MHPHE0  MHPHE1  MHPHE2  MHPTR6  MHPTR8  MHPTR9  MHPTRA  MHPTTW  MHPTTY  MHPTV1  MHPTV2  MHPVJE  MHPVJH  MHPVJJ  MHPVJK  MHPVVR  MHPVVT  MHPVVV  MHPVVW  MHPWV5  MHPWV6  MHPWV7  MHPWV8  MHR13T  MHR14T  MHR154  MHR157  MHR159  MHR3TK  MHR3TL  MHR3TM  MHR3TN  MHR3TR  MHR3TV  MHR3TW  MHR3TX  MHR3TY  MHR9YM  MHR9YN  MHR9YP  MHR9YR  MHRAAP  MJDJM9  MJDJMA  MJDJMD  MJE83L  MJED4P  MJED74  MJED87  MJED98  MJEDHD  MJEDXE  MJEDXH  MJEE5T  MJEE94  MJEEJ5  MJEEL3  MJEER3  MJEET8  MJEHAA  MJEHAT  MJEHD5  MJEHK7  MJEHTD  MJEHW2  MJEHWJ  MJEJ24  MJEJ4V  MJEJKW  MJEJM0  MJEJRK  MJEJWK  MJEK0L  MJEK22  MJEKPT  MJEKRV  MJEKYX  MJEL56  MJEL7D  MJEL8X  MJELEK  MJELMM  MJELMN  MJELMP  MJELMT  MJELXT  MJELXV  MJEM3A  MJHDJH  MJHDJJ  MJHDM3  MJHDNX  MJHDT3  MJHDV0  MJHDV9  MJHDYX  MJHE6N  MJHE8P  MJHK1K  MJHK1N  MJHK1P  MJHK1T  MJHK27  MJHK28  MJHK45  MJHK6Y  MJHKJP  MJHMH4  MJHMJD  MJHMJN  MJHP2E  MJHP5D  MJLWYW  MJLXV2  MJLYP9  MJM04Y  MJM06D  MJM097  MJM0HY  MJM171  MJM1N6  MJM1PT  MJM2EH  MJM2H9  MJM2HD  MJM2HM  MJM30N  MJM406  MJM436  MJM442  MJM4H6  MJM4XJ  MJM59N  MJM5AL  MJM5TK  MJM5TL  MJM6KK  MJM6MW  MJM721  MJM740  MJM76W  MJM7JW  MJM7M0  MJM870  MJM8DY  MJM96L  MJM9E0  MJM9T1  MJM9TM  MJM9TY  MJM9YK  MJMA0H  MJMA8T  MJMATM  MJMATN  MJMD3H  MJMD4H  MJMDD6  MJMDDM  MJMDDN  MJMDTL  MJME5R  MJMEJ5  MJMHTH  MJMJD8  MJMK8X  MJMKHH  MJMKXD  MJMM2R  MJMM6W  MJMMTW  MJMNH2  MJMNKY  MJMNX4  MJMP7H  MJMP89  MJMPVH  MJMR9T  MJMRL5  MJMTHW  MJMTMP  MJMTW7  MJRVHL  MJRVYJ  MJRWRH  MJRXP1  MJRXW5  MJRYJJ  MJRYVT  MJT0EH  MJT231  MJT2X4  MJT2XY  MJT300  MJT308  MJT3KM  MJT3R4  MJT465  MJT4LT  MJT58K  MJT5KA  MJT5KY  MJT69W  MJT6X0  MJT73R  MJT77D  MJT7RH  MJT91E  MJT91H  MJTA7N  MJTD06  MJTD44  MJTD70  MJTD96  MJTDEH  MJTDYJ  MJTE9A  MJTH43  MJTH9E  MJTHAP  MJTJ4R  MJTJL2  MJTK6V  MJTL6R  MJTL7V  MJTL80  MJTL9D  MJTLKY  MJTN62  MJTNHJ  MJTNKV  MJTP2P  MJTPN2  MJTPY2  MJTR51  MJTRN8  MJTRX7  MJTTHE  MJTTYA  MJTTYV  MJTV12  MJTV87  MJTV8X  MJTVTL  MJTVWJ  MJTWWM  MJTX3W  MJTY2X  MJTY80  MJTYDL  MJTYRK  MKA0KT  PWVMJE  PWWMJE  PWXMJE  PWYMJE  PX0MJE  PX1MJE  PX2MJE  R13MEE  R14MEE  R15MEE  R16MEE  R5JMAE  R72MAE  R7LMAE  RPEMAE  RPKMAE  RR2MPD  RR4MPD  RR5MPD  RVJMPA  RVKMPA  RVLMPA  RVLMPAX1  RWRMPD  RWTMPD  RWVMPD  T04MAE  T05MAE  T06MAE  T07MAE  T24MJE  T24MJEX1  T25MJE  T29MJE  T2AMJE  T2DMJE  T2EMJE  T4JMJE  T4MMJE  T4PMJE  T4WMJE  T5LMJE  T5NMJE  T5RMJE  T64MJE  T7JMAE  T86MJE  T8PMDE  T8TMAE  T8TMDE  T8VMDE  T8WMDE  TA4MAE  TAAMJE  TD8MJE  THVMNE  THWMNE  THXMNE  THYMNE  TJ6MJE  TJHMJE  TL7MLE  TL8MLE  TL9MLE  TLDMLE  TLKMAE  TPYMJE  TTDMJE  TWEMJD  TWEMJDX1  0T8MKDX1  0TNMLE  0TXMMD  0TYMMD  0WAMMD  0WDMMD  0WDMMDX1  0WJMNE  0WLMMD  0WNMMD  0WNMMDX1  0X0MNE  0XJMNE  0XWMMD  0Y0MNE  0Y9MLE  0YAMLE  106MMD  10HMED  112MLE  113MLE  11TMKD  11TMKDX1  11VMKD  11VMMD  11WMKD  11YMLE  12RMPE  135MJD  139MKD  13AMKD  13AMKDX1  13DMJE  13DMKD  13DMKDX1  13EMKD  13EMKDX1  13HMKD  13JMKD  13JMKDX1  13KMKD  13KMKDX1  13LMED  13LMKD  13MMED  13NMED  13TMED  13TMNE  13XMED  140MED  142MED  148MNE  14KMPE  14RMJE  15PMLE  165MJE  18YMJE  19AMMD  1J3MTD  1J4MTD  1J6MTD  1J7MTD  1J8MTD  1JNMMD  1LKMMD  1T5MMD  1T8MMD  1VMMHE  1X5MMD  1YYMJD  780MTA  781MTA  781MTAX1  78TMNE  793MNE  79AMMD  7A6MTD  7D9MMD  7DAMDE  7DDMTD  7DMMJE  7EHMJE  7EJMDE  7H9MDE  7HXMJE  7J1MMD  7K8MJE  7LHMDE  7MEMDD  7MJMDD  7MJMJE  7MVMDD  7MYMDD  7N1MDD  7N6MDD  7N8MDD  7NAMHD  7NAMHDX1  7NDMHD  7NEMHD  7NEMHDX1  7NHMHD  7NHMHDX1  7NJMHD  7NNMMD  7NVMDD  7NVMDDX1  7RAMPA  7RAMPAX1  7TMMPA  7TPMMD  7TPMPA  7TPMPAX1  7TRMMD  7TTMMD  7TWMMD  7VXMJE  7X2MMD  7XYMPD  7Y1MMA  82NMPD  82WMPD  845MME  849MME  852MME  853MME  854MME  855MME  856MME  856MMEX1  858MME  862MMA  89VMPD  8ANMMA  8APMMA  8ARMMA  8E0MMA  8E1MLD  8E2MMA  8E2MMAX1  H0LMMA  H0NMLE  H0PMLE  H4PMRA  H4PMRAX1  H4RMRA  H4RMRAX1  H4TMRA  H6LMLD  H6MMLD  H6PMLD  H8NMDD  H93MDD  H9TMPD  HADMPD  HAEMPD  HD3MPD  HD5MPD  HDMMPD  HE1MLD  HE2MLD  HE2MLDX1  HE3MLD  HE4MLD  HH3MME  HH7MLE  HHJMLE  HHLMLE  HHNMLE  HJ0MDD  HJ0MDDX1  HJHMDD  HJHMDDX1  HJVMDD  HMWMME  HR2MME  HR6MME  HR8MME  HRYMME  HT0MME  HT1MME  HV6MHE  HV7MHE  HV8MHE  HV8MRA  HV9MHE  HV9MRA  HVAMRA  HXPMLE  J15MRD  J1EMRD  J1LMRD  J20MRD  J25MTA  J26MRD  J27MTA  J28MTA  J28MTAX1  J29MTA  J2DMRD  J2DMTA  J2JMTA  J2KMTA  J36MRD  J46MHE  J70MHE  J96MLE  J99MLE  J9DMLE  JH0MHE  MEPHV2  MEPHV3  MEPHV5  MEPJ6X  MEPJ70  MEPJ7L  MEPJ8H  MEPJ8K  MEPM70  MEPMDE  MEPMLA  MEPMPD  MEPMRY  MEPMT1  MEPMTX  MEPMTXX1  MEPMV3  MEPNX6  MEPNXA  MEPNXD  MEPNXDX1  MEPNXE  MEPR0T  MEPR2J  MEPTY6  MEPTY7  MEPTY8  MEPTY9  MEPVHM  MEPVJ9  MEPVLJ  MEPVMR  MEPVN7  MEPVN8  MER1ET  MER1EV  MER1EW  MER1EX  MER22W  MER22WX1  MER22WX2  MER23P  MER274  MER27H  MER297  MER298  MER299  MER29A  MER29AX1  MER29E  MER29EX1  MER29H  MER29HX1  MER36E  MER36H  MER3D6  MER3K4  MER3K4X1  MER3LH  MER5YR  MER601  MER608  MER6H4  MER6H7  MER6HR  MER6HT  MER6XJ  MER6XK  MER6XL  MER6XY  MHEK62  MHEK87  MHEK89  MHEK89X1  MHEKEY  MHEKH0  MHEKKL  MHEKKT  MHEKVW  MHEKVWX1  MHEL21  MHEL3T  MHEL42  MHEL42X1  MHEL43  MHEL50  MHEL50X1  MHELT7  MHELT8  MHELT8X1  MHELT9  MHELT9X1  MHELTA  MHEM77  MHEM78  MHEM81  MHEM8K  MHEMP7  MHEMP7X1  MHEMV0  MHEMV0X1  MHEMYT  MHEMYTX1  MHEN0J  MHEN0JX1  MHEN0K  MHEN0KX1  MHEN2R  MHEN8D  MHEN8DX1  MHEN8J  MHEN8JX1  MHEN8K  MHEN8M  MHEN8MX1  MHENMT  MHENR1  MHENX0  MHENX4  MHENX4X1  MHEP15  MHEP2W  MHEP82  MHEP82X1  MHEPAX  MHEPAXX1  MHEPAY  MHEPAYX1  MHEPD0  MHEPKJ  MHEW70  MHEW70X1  MHEWD8  MHEWD8X1  MHEWJE  MHEX07  MHEX07X1  MHEX08  MHEX29  MHEX29X1  MHL70A  MHL70N  MHL710  MHL712  MHL72K  MHL72L  MHL72M  MHL72N  MHL72P  MHLE5K  MHLE5L  MHLE5M  MHLE5N  MHLE5P  MHLE5R  MHLHR3  MHLHR4  MHLHR6  MHLHRK  MHLHRL  MHLHRR  MHLHRX  MHLJ4X  MHLJ50  MHLJ9M  MHLMKE  MHM07R  MHM07T  MHM07V  MHM07W  MHM07X  MHM13V  MHM14P  MHM14R  MHM14T  MHM14V  MHM158  MHM1AN  MHM1VN  MHM1VW  MHM1VX  MHM1VY  MHM4N4  MHM4N5  MHM4N7  MHM4N8  MHM4N8X1  MHM4NA  MHM771  MHM772  MHM775  MHM778  MHM7V8  MHM87A  MHM87D  MHM88W  MHM88Y  MHM8A3  MHM8AL  MHM8HY  MHM8LK  MHM8LL  MHM8LP  MHM8LR  MHME14  MHME21  MHME22  MHME23  MHME24  MHME8T  MHRAAR  MHRAAT  MHRAAV  MHRAAW  MHRAVY  MHRAW0  MHRAW1  MHRAW2  MHRAW3  MHRAW4  MHRAW5  MHRAW6  MHRAW7  MHRDYM  MHRDYN  MHRDYP  MHRDYR  MHRE3L  MHRE3M  MHRE3N  MHRE3P  MHRKWY  MHRKX1  MHRKX2  MHRKX3  MHRKX4  MHRKX6  MHRKX7  MHRKX8  MHRLME  MHRLMH  MHRLMJ  MHRLMK  MHRLML  MHRLMM  MHRLMN  MHRLMP  MHRLMR  MHRMAH  MHRMAJ  MHRMAK  MHRMAP  MHT4LH  MHT4P5  MHT4PL  MHT4PM  MHT4PX  MHT4RH  MHT4TJ  MHT4TK  MHT4TL  MHT514  MHT541  MHT547  MHT574  MHT5AT  MHT5DN  MHT5HM  MHT5MJ  MHT5RD  MHT5T4  MHT60R  MHT60V  MHT62T  MHT62V  MHT63Y  MHT640  MHT647  MHT648  MHT9ET  MJHP6M  MJHP78  MJHP9A  MJHP9D  MJHP9E  MJHP9H  MJHPDJ  MJHPDL  MJHPDM  MJHPDN  MJHPJX  MJHPJY  MJHPN3  MJHPP8  MJJ0XP  MJJ0XR  MJJ0XT  MJJ0XV  MJJ12D  MJJ12E  MJJ14V  MJJ173  MJJ19D  MJJ1E3  MJJ1JM  MJJ1L2  MJJ1M5  MJJ1PT  MJJ1XX  MJJ1XY  MJJ1Y2  MJJVEL  MJJVEM  MJJVEN  MJK0TV  MJK0VK  MJK0VW  MJK0W6  MJK11V  MJK163  MJK170  MJK1A2  MJK1AL  MJK1L6  MJK1N8  MJK1TV  MJK1V5  MJK1V6  MJK1V7  MJK1V8  MJK26P  MJK26R  MJK26T  MJK26V  MJK2T6  MJK2T7  MJK2T8  MJK2T9  MJK3N4  MJK3N5  MJK3N6  MJK3N7  MJK8M1  MJK8M2  MJK8M3  MJK8M4  MJK8M5  MJK91Y  MJK921  MJK922  MJMV10  MJMVD2  MJMVTX  MJMWD6  MJMWX6  MJMX0J  MJN031  MJN0J0  MJN12A  MJN15Y  MJN194  MJN19T  MJN1D7  MJN1MK  MJN1R1  MJN1TM  MJN267  MJN2X2  MJN36T  MJN3L5  MJN3YX  MJN4AA  MJN4TY  MJN57D  MJN5A5  MJN5DT  MJN5JR  MJN67D  MJN6DT  MJN6JR  MJN6T0  MJN714  MJN7JA  MJN7TR  MJN842  MJN8TN  MJN9ED  MJN9XP  MJNA68  MJNAD9  MJNAY1  MJND5V  MJNDK7  MJNDN3  MJNETN  MJNEXK  MJNH54  MJNH59  MJNHD0  MJNHY5  MJNJ8V  MJNJEX  MJNJYM  MJNK88  MJNKNN  MJNKT8  MJNLAH  MJNLRJ  MJNM3V  MJNM96  MJNMD8  MJNML2  MJNN43  MJNN9N  MJNP1V  MJNP7P  MJNP99  MJNPRR  MJNPRX  MJNR2R  MKA0YP  MKA10V  MKA1AL  MKA1AX  MKA1HV  MKA1HX  MKA1N0  MKA1T8  MKA1X6  MKA289  MKA2L4  MKA2MX  MKA30J  MKA35A  MKA35P  MKA3KJ  MKA486  MKA4K9  MKA5JT  MKA5K5  MKA5LK  MKA5V7  MKA623  MKA62W  MKA6HN  MKA6VL  MKA6WP  MKA71X  MKA7XM  MKA834  MKA93X  MKA9JN  MKAA01  MKAA62  MKAAHV  MKAAJJ  MKAAR7  MKAD6R  MKADEM  MKADH9  MKADJ9  MKADK8  MKAE24  MKAHK9  MKAK58  MKAKJX  MKAL0J  MKAL8L  MKAL8Y  MKALJ7  MKALWW  MKALYW  MKAM74  MKAMMH  MKAMR7  MKAN3Y  MKANP5  MKAPLN  MKAPP5  MKAPYV  MKARDN  MKARMA  MKARP3  MKARWK  MKATMV  MKAV3Y  MKAV5E  MKAWNY  MKAXE1  MKAXTX  TWJMJD  TWJMJDX1  TWKMJD  TWKMJDX1  TWNMNE  TWRMNE  TWTMNE  TXXMNE  V28MHE  V29MHE  V29MHEX1  V2AMHE  V2VMDE  V8EMJD  V8EMJDX1  V8HMJD  V8KMJD  V8LMJD  V8LMJDX1  V8MMJD  V8MMJDX1  V8WMJD  V91MJD  V91MJDX1  V92MJD  V92MJDX1  V93MJD  V93MJDX1  V93MJDX1X1  VAJMLE  VE8MAE  VE9MAE  VE9MME  VEAMAE  VEAMME  VEDMAE  VEDMME  VEEMME  VK6MJE  VKHMLE  VKJMLE  VKKMLE  VKLMLE  VM2MJE  VM3MJE  VM5MJE  VMAMKA  VMDMKA  VMDMKAX1  VMDMKAX1X1  VMDMNE  VMEMNE  VMRMNE  VMTMNE  VR7MKA  VR9MKA  VR9MKAX1  VRMMKA  VV2MME  VV3MME  VV6MKD  VV6MKDX1  VVYMKD  VW2MKD  VWTMKD  VWTMKDX1  VX3MME  VY2MME  VY5MME  VYVMKD",
      "center_classification_date": "20180702",
      "distribution_pattern": "Distributed nationwide. Foreign distribution to Australia, Canada, Chile, China, Brazil, France, Germany, India, Italy, Japan, Malaysia, Netherlands, New Zealand, Poland, Portugal, Romania, Singapore, South Africa, Spain, Sweden, Switzerland, and United Kingdom.",
      "state": "NJ",
      "product_description": "LFIT Anatomic CoCr Femoral Heads,  Catalog No. 6260-9-140",
      "report_date": "20180711",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Howmedica Osteonics Corp.",
      "recall_number": "Z-2296-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80059",
      "termination_date": "20200508",
      "more_code_info": "",
      "recall_initiation_date": "20180522",
      "postal_code": "07430-2006",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}