  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Mahwah",
      "address_1": "325 Corporate Dr",
      "reason_for_recall": "The recalling firm has received a higher than expected number of complaints documenting femoral head/hip stem disassociation for certain sizes of femoral heads manufactured prior to March 4, 2011.",
      "address_2": "",
      "product_quantity": "",
      "code_info": "Lots  13461301  13461302  14097201  14097202  14390101  14390102  14814301  14934401  14941101  14973401  15029001  15029002  15186801  15186802  15186803  15341401  15341402  15341403  15341404  15629401  15629402  15759601  16462601  16551101  16551102  16575601  16575602  16644801  16666901  16684501  17159001  17159002  17230201  17230202  18077901  18110901  18110902  18110903  18216001  18216002  18216003  18298901  18298902  18744901  18748501  18748502  18783701  18783702  18783703  19122001  19122002  19362201  19362401  19362501  19362601  19506201  19506202  19549601  19549602  19634101  19634102  19634103  19634104  19691001  19691002  19696801  19696802  19696803  19716501  19719801  22400903  22400904  22402901  22407101  22648401  22648402  22648403  22740701  22740702  22740703  22740704  22740705  22775201  22775202  22775203  22775204  22775205  22817801  23434501  23434502  23434503  23471701  23516001  23516002  23516003  23542801  23542802  23542803  24109201  24122901  24142401  24142402  24205901  24205902  24268901  24268902  24282401  24297601  24806101  24806102  24867301  24867401  24941701  25290501  25290502  25344001  25344002  25360701  25369201  25369202  25369203  25447801  25447802  25447803  26147701  26147702  26195101  26195102  26195103  26205301  26205302  26205303  26290001  26290002  26306101  26380701  27257501  27304201  27304202  27304203  86149403  86149404  86149405  86296301  86296302  86296303  86375901  86375902  86375903  86375904  86375905  86466301  86466302  88966401  88966402  88966403  89044301  89446101  89840201  89840202  90405301  90405302  91124201  91124202  91797901  91797902  91797903  92501201  92501202  92501203  92501301  92993801  92993802  92993803  93418201  93418202  93418203  93982801  93982802  94499901  94499902  95178201  95178202  95178203  95501301  95501302  95821201  95821202  95821203  96419601  96419602  96419603  96419604  96837401  96837402  96837403  97136401  97136402  97338901  97338902  98094401  98426601  98426602  98883501  98883502  99063401  99063402  99350201  99350202  000MPA  20484207X1  20484207X1X1  20484207X2  20484208X1  20547701X1  20547702X1  20547703X1  20547703X1X1  20547703X2  20547704X1  20547705X1  20547705X2  20684401X1  20684403X1  20684405X1  20684405X1X1  20684406X1  20684407X1  20716803X1  20783204X1  20934403X1  21029805X1  21109603X1  21136001X1  21185902X1  21185903X1  21185906X1  21208601X1  21208604X1  21263802X1  21437503X1  21437504X1  21437507X1  21489405X1  21654005X1  21696101X1  21696102X1  21696103X1  21AMLE  21EMLE  21PMRD  22295206X1  22400902X1  225MPE  228MPE  22AMPE  255MPE  255MPEX1  25KMDE  25LMDE  25MMDE  26205302A  26WMTD  26XMTD  26YMTD  27TMLE  27VMLE  28EMHE  28HMHE  28JMHE  295MDE  296MDE  297MDE  297MDEX1  29KMPE  2A3MEE  2A4MEE  2A5MEE  2DKMEE  2EHMPE  63KMEE  63WMPD  63YMNE  648MEE  649MEE  64AMEE  64DMEE  67PMLE  67PMNE  684MLA  69XMJD  69YMJD  6A0MJD  6A1MJD  6A2MJD  6A3MJD  6A4MJD  6A5MJD  6A6MJD  6A7MJD  6D2MLA  6D6MLE  6DNMLE  6DWMLE  6E1MLE  6E7MPA  6E7MPAX1  6E8MPA  6E8MPAX1  6E9MPA  6E9MPAX1  6E9MPAX1X1  6EAMPA  6EAMPAX1  6EAMPAX1X1  6H1MEE  6H2MEE  6H3MEE  6HHMJE  6HHMJEX1  6HJMJE  6HKMJE  6HTMED  6HVMED  6HWMED  6HYMJD  6J0MED  6J5MNE  6J6MNE  6N4MEE  6PHMHA  6PJMHA  6V1MLA  6VJMHE  6VKMHE  6VLMHE  6VXMLD  707MLD  726MKD  726MKDX1  727MKD  728MKD  729MKD  729MKDX1  72AMKD  72EMKD  72HMKD  72JMKD  72NMEE  72WMNE  ALWMDE  ALWMDEX1  ALXMDE  ALYMAE  ALYMDE  ANMMTA  ANMMTAX1  APXMTA  APXMTAX1  APYMTA  AR2MTA  ARRMTA  AV1MHA  AV2MHA  AV3MHA  AV4MHA  AV5MHA  AV6MHA  AW8MHE  AXNMHA  AXPMHA  AXRMHA  AXTMHA  AXVMHA  AXWMHA  D4MMDE  D5HMLD  D6NMAE  D72MME  D7HMME  D7JMME  D7KMME  D85MDE  D9YMJD  DA0MJD  DA1MJD  DA6MJD  DA7MJD  DA9MDE  DE4MAE  DK0MKA  DL2MMD  DL4MMD  DLAMMD  DLAMMDX1  DMPMDE  DMRMDE  DMTMDE  DNLMEE  DNMMEE  DNNMEE  DPTMME  DR6MKA  DRTMAE  DTLMJD  DXVMLA  DXVMLAX1  DXVMLAX1X1  DXWMLA  E0XMHE  E0YMHE  E10MHE  E1YMKA  E24MKD  E26MHA  E4KMTD  E5DMLD  E5EMLD  E5EMLDX1  E65MME  K4XMJA  K4YMJA  K68MKD  K6AMEE  K8NMKD  K8NMKDX1  K8NMKE  K8YMJE  K8YMNE  K90MJE  K91MJE  K92MNE  K95MNE  K98MND  K9DMAD  K9DMADX1  K9EMAD  K9EMADX1  K9JMAD  K9KMAD  K9KMADX1  KDEMEE  KEHMHD  KH0MDE  KH1MDE  KH2MDE  KHHMHD  KHYMLE  KJJMLE  KJPMDE  KJRMDE  KJTMDE  KJVMAD  KJVMADX1  KJWMAD  KJYMAD  KK1MAD  KKHMLE  KKLMJE  KKPMND  KL6MLE  KLKMDE  KM4MRD  KMXMDE  KN8MEE  KNKMRD  KNWMDD  KNWMDDX1  KNXMDD  KNXMLE  KNYMDD  KNYMDDX1  KNYMLE  KP0MDD  KPVMKD  KPWMKD  KPXMKD  KPYMKD  KR0MKD  KR1MKD  KR2MKD  KRVMMD  KRXMRD  KV0MNE  KV1MNE  KV2MNE  KV6MNE  KV7MNE  KVMMLE  KVMMNE  MER21X  MER23W  MER245  MER245X1  MER269  MER26T  MER26TX1  MER26TX2  MER26W  MER26WX1  MER26Y  MER26YX1  MER272  MER279  MER27K  MER27KX1  MER29K  MER29KX1  MER29L  MER29LX1  MER29M  MER29P  MER29R  MER2DJ  MER3D3  MER3D3X1  MER3D3X2  MER3D4  MER743  MER743X1  MER7E6  MER7E6X1  MER7N7  MER7RM  MER7RMX1  MER858  MER87M  MER87P  MER890  MER8LV  MER8R5  MER8R6  MER8R6X1  MER8R6X2  MER93J  MER97K  MER983  MER986  MER99D  MER9A1  MER9A6  MERL3M  MERL3N  MERL3P  MERMY4  MERN5D  MERN76  MERNJ3  MERNN2  MERPDH  MERPJY  MET0YR  MET0YRX1  MET12J  MET12K  MET24A  MET24E  MET24H  MET24J  MET24K  MHD595  MHD5ED  MHD60H  MHD62A  MHD62R  MHD630  MHD631  MHD631X1  MHD69P  MHD6AL  MHD6H2  MHD6H2X1  MHD6M3  MHD6N0  MHD6TE  MHD6X2  MHD72M  MHD7AH  MHD7AHX1  MHD7AN  MHD7K3  MHD7K5  MHD7MR  MHD7MRX1  MHD7RP  MHD7YV  MHD80A  MHD88A  MHD88K  MHD88KX1  MHD8M8  MHD8NH  MHD8NW  MHD8RX  MHD8T8  MHD8V7  MHD8Y9  MHD962  MHD962X1  MHDD4T  MHDD62  MHDD9W  MHDDK7  MHDDMW  MHDEHY  MHDEJA  MHDEW8  MHDEYM  MHDH13  MHDM73  MHDM8L  MHDM8LX1  MHDM9T  MHDMHD  MHDMJ1  MHDMN1  MHDNPR  MHDNPRX1  MHDP9R  MHDPLP  MHDR14  MHE1HA  MHE1JN  MHE1Y0  MHE1Y8  MHE221  MHE22T  MHE22TX1  MHE22TX2  MHE23H  MHJ0T5X1  MHJ0V7  MHJ0W2  MHJ0WX  MHJ16K  MHJ58D  MHJ597  MHJ5AM  MHJ5VA  MHJ5VAX1  MHJ61L  MHJ6DE  MHJ6DEX1  MHJ6EA  MHJ6H4  MHJ6HN  MHJ6NJ  MHJ7A5  MHJ7K5  MHJ7K5X2  MHJ7N2  MHJ7NK  MHJ7P5  MHJ7PM  MHJ84M  MHJ86D  MHJ8A6  MHJ8MM  MHJ8PK  MHJ8R6  MHJ92H  MHJ92HX1  MHJ97W  MHJ9JP  MHJ9N5  MHJ9PP  MHJ9RN  MHJ9V7  MHJ9V8  MHJA9L  MHJAJ3  MHJAM5  MHJAWM  MHJAXV  MHJAXY  MHJAYL  MHJAYN  MHJAYR  MHJD5T  MHJD5TX1  MHJD5V  MHJD5W  MHJDT8  MHK122  MHK123  MHK124  MHK125  MHK1K9  MHK1RW  MHK1VE  MHK1W2  MHK1W7  MHK21A  MHK22L  MHK251  MHK2H1  MHK2H6  MHK2M0  MHK2MX  MHK2N1  MHM426  MHM42P  MHM43V  MHM47R  MHM47V  MHM47W  MHM47Y  MHM480  MHM4VV  MHM50X  MHM512  MHM518  MHM51M  MHM52W  MHM52WX1  MHM566  MHM5HP  MHM5Y9  MHM62K  MHM64M  MHM6A5  MHM6DH  MHM6M0  MHM6VH  MHM70X  MHM72E  MHM72EX1  MHM742  MHM76W  MHM7AJ  MHM7HM  MHM7N1  MHM84T  MHM87J  MHM882  MHM890  MHM89P  MHMH9R  MHMHAX  MHMHAXX1  MHMHEM  MHMHJE  MHMHPH  MHMJAT  MHMJATX1  MHMJDR  MHMJH5  MHMJLT  MHMJMH  MHMJMV  MHMJMX  MHMJN0  MHMK92  MHMK93  MHMKEJ  MHMKEK  MHMKH0  MHMKH6  MHMKHJ  MHMKKM  MHMKMP  MHMKMT  MHMKNY  MHMKP0  MHMKP0X1  MHMKP4  MHMKP4X1  MHMKR9  MHMKXD  MHML54  MHPP0J  MHPP0K  MHPP0L  MHPP0M  MHPX6D  MHR02Y  MHR0A3  MHR0H1  MHR0P4  MHR19D  MHR1K6  MHR1M7  MHR1NL  MHR2KL  MHR2PE  MHR2T5  MHR2Y3  MHR31R  MHR3DE  MHR3H9  MHR3HA  MHR3JD  MHR3JH  MHR3K4  MHR3LL  MHR3RP  MHR3WX  MHR3XY  MHR43W  MHR48L  MHR4JY  MHR4MR  MHR4N8  MHR4PT  MHR4R6  MHR4RL  MHR4W3  MHR555  MHR5XN  MHR5YY  MHR600  MHR672  MHR6DX  MHR6L5  MHR733  MHR76E  MHR76Y  MHR770  MHR7AK  MHRDY5  MHRDY9  MHRDY9X1  MHRE01  MHRE2D  MHRE3K  MHRE4Y  MHRE62  MHREDT  MHREKR  MHRENY  MHRETW  MHRH3W  MHRH5L  MHRH5V  MHRH6H  MHRH76  MHRH82  MHRH93  MHRHHJ  MHRHJY  MJD7N7  MJD7N8  MJD7N9  MJD7NA  MJD7ND  MJD7NE  MJD7NH  MJD7NJ  MJD7NK  MJD7NL  MJD7NM  MJD8XN  MJD8XP  MJD8YR  MJD90M  MJD90N  MJD90P  MJD90R  MJD90T  MJD90V  MJD98K  MJD98L  MJD98N  MJD98R  MJD98T  MJD98V  MJDJVX  MJDK7W  MJDKE9  MJDL01  MJDL3D  MJDL6Y  MJDLJK  MJDLNR  MJDLW5  MJDMVP  MJDMVW  MJDMW0  MJDMW2  MJDR0K  MJDR17  MJDR5H  MJDR87  MJDT30  MJDT6M  MJDTHL  MJDTKV  MJE01Y  MJE078  MJE17X  MJE193  MJE1KE  MJE21P  MJE229  MJE6EW  MJE6W3  MJE703  MJE79V  MJE7DW  MJE7E0  MJE7HV  MJE7K9  MJE7KV  MJE7KW  MJE7LM  MJE7N1  MJE7W3  MJE808  MJE8NM  MJE8PA  MJJHTK  MJJJHN  MJJJJJ  MJJJTA  MJJJX9  MJJK1H  MJJK2T  MJJKYR  MJJL2M  MJJL51  MJJLTX  MJJLWA  MJJLXE  MJJLXH  MJJLXJ  MJJM0Y  MJJM2N  MJJM7M  MJJMAD  MJJMEH  MJJMEJ  MJJMEL  MJJMV3  MJJMYJ  MJJN0E  MJJN51  MJJN9E  MJJNEX  MJJNXJ  MJJP11  MJJP12  MJJP13  MJJP1M  MJJP2V  MJJP30  MJJP5N  MJJPDX  MJJPDY  MJJPE0  MJJPN3  MJJPTW  MJJPTY  MJJPV0  MJJR5M  MJJR5N  MJJR65  MJJR7E  MJJRAL  MJJRM0  MJJRX0  MJJTD2  MJJTDJ  MJJTEE  MJJTJA  MJJTR9  MJJTVH  MJJV2L  MJJV2N  MJJV2P  MJJV90  MJJVA9  MJJVAD  MJJVDW  MJJVDX  MJJVDY  MJKLXY  MJKLY0  MJKLY1  MJKLY8  MJKM0H  MJMRXE  MJMRY3  MJMT28  MJMT6H  MJMTLN  MJMV0L  MJMV24  MJMVA1  MJMVY2  MJMW3Y  MJMW6N  MJMW7N  MJMW9E  MJMWKA  MJMWLR  MJMX07  MJMX17  MJMXT6  MJMY7P  MJMY8A  MJMY9D  MJMYA7  MJMYAR  MJN03R  MJN07J  MJN08J  MJN0H9  MJN15K  MJN1NL  MJN1PT  MJN243  MJN2HD  MJN2LY  MJN2RR  MJN2RX  MJN339  MJN38T  MJN3N0  MJN3PM  MJN3RW  MJN4KL  MJN4NN  MJN4XE  MJN5ET  MJN5WW  MJN5X0  MJN62X  MJN6A5  MJN70P  MJN73L  MJN75J  MJN7J6  MJN7JN  MJN7M0  MJN7M5  MJN7WN  MJN89E  MJN8EA  MJN8PH  MJN9M5  MJN9W9  MJNA02  MJNA6R  MJNAAP  MJNAR8  MJNAW2  MJND50  MJNDAV  MJNDDP  MJNDTW  MJR9DV  MJR9T6  MJR9X9  MJRAPH  MJRD1D  MJRDDV  MJRDN7  MJRDV3  MJRDV4  MJRDW4  MJRDWX  MJRE20  MJRE23  MJRE4L  MJRE6Y  MJREER  MJREP9  MJRERH  MJRH68  MJRHD8  MJRHWT  MJRHY2  MJRHYL  MJRJ23  MJRJ6Y  MJRJHE  MJRJPT  MJRJTD  MJRK0E  MJRK94  MJRL04  MJRL21  MJRL5P  MJRL83  MJRL9W  MJRLD8  MJRLET  MJRLRX  MJRLXE  MJRMEK  MJRMK8  MJRMMA  MJRMV3  MJRMVJ  MJRN48  MJRN5N  MJRNH7  MJRNVA  MJRP40  MJRPLL  MJRPMY  MJRPNV  MJRR0W  MJRR3X  MJRR56  MJRR60  MJRR6P  MJRRMK  MJRT1D  MJRT9J  MJRT9P  MJRTAM  MJRTEM  MJRTLK  MJRV19  MJRVM2  MJRVYY  MJRW2A  MJRW4A  MJRW8M  MKD92M  MKD94E  MKDA3J  MKDAH2  MKDAHH  MKDAHY  MKDAJ8  MKDD0H  MKDE96  MKDHAJ  MKDHY4  MKDHYY  MKDK4L  MKDM8K  MKDMYX  MKDN1L  MKDN8E  MKDNT4  MKDNYA  MKDP13  MKDPJ9  MKDPPA  MKDR00  MKDRJ6  MKDRN5  MKDW9N  MKDWET  MKDWHR  MKDWKN  MKDWLW  MKDWMV  MKDWW9  MKDWWR  MKDWYH  MKDX10  MKDXJR  MKDXY4  MKDY4J  MKDY60  MKDYKE  MKE01Y  MKE0K4  MKE120  MKE17X  MKE1R0  MKE28W  MKE4VW  MKE5K8  MKE73A  MKE7YK  MKE825  MKE84P  MKE84Y  MKE8HE  MKE8HL  MKE8W9  MKEDLT  MKEK93  MKELD2  MKELKE  MKEM6T  MKEMJE  MKEN39  MKENPD  MKERV0  MKETHY  MKEV4Y  MKEWVA  MKEX2K  MKEXHD  TAPMNE  TDKMEE  TDLMEE  TDLMNE  TDMMEE  TH0MED  TL5MED  TL5MEDX1  TN4MAD  TRWMMA  TRXMMA  TRYMMA  TRYMMAX1  TW0MMA  TW0MMAX1  TW1MMA  TW2MMA  TWRMLE  V0RMHA  V1EMDE  V1HMDE  V1JMDE  V2EMHE  V2EMHEX1  V32MHE  V33MHE  V34MHE  V4RMDE  V6KMJD  V76MLE  V79MLE  V7DMHA  V7NMAD  V7PMAD  V7YMHE  V80MHE  V81MHE  V89MAD  VAEMEA  VAHMEA  VAJMEA  VAKMEA  VALMEA  VAMMEA  VEHMAE  VEJMAE  VEKMAE  VELMHE  VHWMLE  VL4MJD  VL9MMD  VLNMLE  VP0MHE  VPJMHE  VV3MMA  VV3MMAX1  VVLMEE  VVMMEE  VVPMEE  VW8MED  VW9MLE  VWPMJD  VX0MMA  VX0MMAX1  VXHMJD  VYMMAE  VYNMMD  VYRMMD  VYRMMDX1  VYTMMD  19719802  19719803  19780301  19780302  19791001  19791002  19791003  19791004  19842101  19842102  19925401  19925402  19929402  19929403  19929404  19972601  19972602  19972603  19995201  19995202  19995203  19995204  20081701  20081703  20105201  20105202  20155201  20155202  20155203  20155204  20155205  20155206  20167401  20167402  20167403  20199601  20199602  20199603  20199604  20199605  20200401  20484201  20484202  20484203  20484204  20484205  20484206  20484207  20484208  20547701  20547702  20547703  20547704  20547705  20620201  20620202  20620203  20620204  20620205  20684401  20684402  20684403  20684404  20684405  20684406  20684407  20684408  20684409  20716801  20716802  27304204  27327401  27327402  27327403  27327404  27347101  27347102  27347103  27347104  28238801  28287901  28312401  28312402  28312403  28312404  28312405  28365201  28365202  28365203  28365204  28365205  28875801  28875802  28875803  29206601  29206602  29209801  29209802  29209803  29273201  30032201  30032202  30032203  30077001  30077002  30077003  30112701  30201501  30201502  30201503  30264501  30264502  30264503  30311001  30501601  30501602  30501603  30536201  30648101  30648102  30648103  30648104  30887001  30887002  30887003  30887004  31309301  31309302  31309303  31309304  31309305  31337801  31337802  31989801  31989802  31989803  31989804  31989805  32664101  32664102  001MPA  002MPA  002MPAX1  003MPA  003MPAX1  005MNE  006MNE  00HMHD  00JMTD  00RMTD  00RMTDX1  010MMD  019MTD  01YMKD  027MTD  02YMHD  05JMNE  05LMNE  05MMNE  05TMTD  07NMAE  08RMKD  0H5MMD  0K0MMD  0KAMJD  0LYMKD  0M8MEE  0M8MHD  0M9MEE  0MTMNA  0MTMNAX1  0MVMNA  0T0MHD  0T3MKD  0T7MJD  0VWMEE  0WWMHD  0WWMHDX1  0WXMHD  0WXMHDX1  0X0MHD  0X1MHD  0Y9MED  0Y9MEDX1  0YAMED  0YEMRD  0YJMED  0YKMED  104MKD  105MKD  106MKD  107MKD  112MHD  112MHDX1  11HMHD  11HMHDX1  11VMPD  11XMKD  124MPD  125MPD  125MPDX1  134MEE  134MKD  135MEE  135MKD  136MEE  14390101A  149MDE  14AMDE  14DMDE  2EHMPEX1  2H0MEE  2HAMJE  2HHMJE  2HJMJE  2MDMNE  2NWMLE  2P9MPE  2VHMLE  2VHMPE  2Y4MLE  2YAMLE  30112701C  301MRD  30DMLE  317MPE  32KMKD  32PMKD  32RMKD  32RMKDX1  32TMKD  32VMKD  32WMKD  33NMTD  33PMTD  33RMTD  341MRD  342MRD  347MRD  348MRD  349MRD  34AMRD  34MMEA  34NMEA  34PMEA  34RMEA  34TMEA  34XMNE  350MHA  36MMDD  36NMDD  36PMDD  36RMDD  36YMPE  371MPE  372MPE  38449801C  38449802C  38449803C  38449804C  38449805C  38449806C  38449807C  38449808C  38449809C  385MNE  387MNE  38AMNE  392MNE  393MPE  394MPE  3AWMPE  3DHMPE  3H7MPE  3HKMED  3J0MPE  3J0MPEX1  3J7MME  3J8MME  3J9MME  741MLA  742MLA  744MKD  745MKD  745MKDX1  749MLA  74AMME  74JMLA  74KMLA  758MPA  758MPAX1  758MPAX1X1  759MPA  75AMPA  75HMLD  75HMLDX1  76AMNE  772MJE  773MJE  79EMJE  79EMJEX1  79KMJE  79MMND  79NMND  79NMNDX1  79PMND  7DHMRD  7E5MJE  7E6MJE  7E7MJE  7H3MED  7H3MLD  7J8MME  7L2MED  7NEMLD  7NKMME  7PLMHA  7PMMHA  7PWMED  7PYMED  7PYMEDX1  7R0MED  7R0MEDX1  7R1MED  7R2MED  7R3MED  7R3MEDX1  7R5MED  7R5MEDX1  7RAMJE  7RHMJE  7RJMJE  7TDMME  7YEMME  7YJMME  819MND  819MNDX1  81HMND  81HMNDX1  81JMND  81JMNDX1  82KMEA  82LMEA  82MMEA  82NMEA  82RMEA  83NMME  855MEA  85AMLD  85KMME  E67MRD  E6MMMD  E74MMD  E7DMME  E7MMPD  E9EMLD  E9EMLDX1  E9HMTA  E9HMTAX1  E9HMTAX2  E9JMLD  E9JMTA  E9JMTAX1  E9KMTA  E9KMTAX1  E9LMTA  E9VMDD  E9WMDD  E9XMDD  E9XMDDX1  EA0MDD  EA0MDDX1  EA1MDD  EA2MDD  EA4MDD  EA6MDD  EA6MDDX1  EA6MLD  EA8MKE  EA9MME  EATMLD  EATMLDX1  ED2MLD  EDLMLD  EDWMME  EE0MKA  EEAMME  EEHMKD  EH2MKE  EHAMLD  EMPMKA  EN7MKD  ENTMKE  EP0MED  EP2MED  EP4MED  EP5MED  ERAMMD  ERLMME  ERMMLD  ETEMND  EV4MKA  EW6MLD  EWKMKA  EWLMKA  EWMMKA  EX0MLD  EX9MMD  EY5MPD  EYVMKE  H0NMHE  H11MND  H1WMPD  H1WMPDX1  H38MME  H3AMME  H3TMPD  H5DMND  H5DMNDX1  H5EMND  KVWMLE  KVXMDE  KW0MMD  KW8MND  KW9MND  KWDMND  KWWMHE  KWXMHE  KWYMHE  KX4MRD  KX5MME  KX6MME  KX7MME  L0VMHD  L13MEE  L13MJA  L14MEE  L15MEE  L1KMRD  L39MLA  L3DMLA  L3EMLA  L3EMLAX1  L3HMLA  L3XMAE  L3YMAE  L40MAE  L41MAE  L45MJA  L4KMPD  L4KMPDX1  L4LMMD  L4RMDE  L4VMHD  L56MEE  L5DMJA  L83MPD  L83MPDX1  L84MPD  L84MPDX1  L8AMJA  L8JMPD  L8MMPD  L8MMPDX1  L94MEE  L9JMDE  L9LMDE  L9VMJA  LA4MJA  LA8MRD  LA9MRD  LAAMRD  LADMRD  LAEMJA  LAHMJA  LAJMJA  LALMJA  LAMMJA  LAMMJAD  LAPMJA  LARMHD  LARMJA  LAVMHD  LAVMHDX1  LAXMHD  LD2MHD  LERMHD  LETMHD  LH3MHE  LHKMPD  MET24M  MET24N  MET24P  MET24R  MET24RX1  MET24T  MET24V  MET24VX1  MET3DM  MET3DN  MET43M  MET43N  MET43P  MET460  MET461  MET462  MET46J  MET48X  MET490  MET4EL  MET4L0  MET4LJ  MET4LL  MET4Y2  MET55E  MET55H  MET560  MET560X1  MET561  MET5K7  MET60J  MET60L  MET66K  MET6H4  MET6H7  MET6H7X1  MET6L6  MET6MX  MET6WT  MET6WY  MET6X4  MET6X9  MET72X  MET72Y  MET72YX1  MET72YX1X1  MET7DL  MET7J5  MET7J7  MET7JW  MET7JWX1  MET7K3  MET7K4  MET7LK  MET7NR  MET7X3  MET800  MET801  MET802  MET8MK  MET8ML  MET8PD  MET8X7  MET8X7X1  MET8XD  MET93W  MET96D  MET9E7  MET9HM  MET9L4  MHE257  MHE26R  MHE2AK  MHE2L3  MHE2LY  MHE31M  MHE34E  MHE34EX1  MHE34EX2  MHE36P  MHE37H  MHE389  MHE38N  MHE3AV  MHE3KJ  MHE3MA  MHE3PJ  MHE3WP  MHE3WR  MHE3XM  MHE3XMX1  MHE44W  MHE457  MHE458  MHE459  MHE463  MHE463X1  MHE47D  MHE489  MHE48K  MHE48V  MHE49L  MHE4KJ  MHE4KP  MHE4KPX1  MHE4M9  MHE4ML  MHE4MN  MHE4PN  MHE4R1  MHE4Y3  MHE53M  MHE55P  MHE55Y  MHE565  MHE57A  MHE57R  MHE57Y  MHE58E  MHE58W  MHE59K  MHE5A6  MHE5JR  MHE5KH  MHE5KHX1  MHE5N4  MHE5P4  MHE5W3  MHE5Y8  MHE624  MHELT4  MHELT5  MHELT6  MHEM75  MHEM76  MHEM80  MHEM8E  MHEMAV  MHEMAVX1  MHEMD7  MHK2P5  MHK2W0  MHK2W1  MHK35V  MHK3A9  MHK3EN  MHK3R9  MHK3WE  MHK3Y5  MHK3YX  MHK434  MHK4D8  MHK53L  MHK544  MHK54M  MHK5A9  MHK5AE  MHK5JK  MHK5JR  MHK67T  MHK67W  MHK6M2  MHK6TA  MHK705  MHK711  MHK719  MHK71M  MHK768  MHK78A  MHK790  MHK7K5  MHK7V9  MHK7X4  MHK7XK  MHK7YH  MHK8LP  MHKLYP  MHKM5V  MHKM6K  MHKM97  MHKMDV  MHKMEE  MHKMMD  MHKMMH  MHKMMV  MHKMMW  MHKMX9  MHKMXT  MHKMXV  MHKMXW  MHKN2J  MHKN4V  MHKN5R  MHKN5V  MHKN74  MHKND7  MHKNEA  MHKNNP  MHKNNT  MHKNPT  MHKNPX  MHKNRA  MHKNRV  MHKNRW  MHKNTR  MHKNTT  MHKNV4  MHKNV5  MHKNV5X1  MHKNXX  MHML81  MHML83  MHMLA2  MHMLA3  MHMLA4  MHMLA6  MHMLA6X1  MHMLA7  MHMLA8  MHMLA9  MHMM2Y  MHMM2YX1  MHMM6H  MHMM9H  MHMMKW  MHMMM2  MHMMMK  MHMMML  MHMMRH  MHMMRJ  MHMMRJX1  MHMN04  MHMN2J  MHMN8R  MHMNHE  MHMNM4  MHMNM4X1  MHMNV1  MHMNXH  MHMNXL  MHMNXN  MHN224  MHN24D  MHN27P  MHN2JN  MHN2T3  MHN2X4  MHN3E7  MHN3H3  MHN3MY  MHN3XM  MHN3XN  MHN408  MHN409  MHN414  MHN53D  MHN577  MHN5AL  MHN5D9  MHN5DV  MHN5LX  MHN5PA  MHN631  MHN6K8  MHN6LR  MHN6M0  MHN6MJ  MHN6NM  MHN8JL  MHN922  MHN939  MHN94P  MHN95E  MHN95H  MHN95J  MHN991  MHN992  MHN9E0  MHN9E3  MHN9E5  MHRHPL  MHRHY0  MHRJ81  MHRJDD  MHRJEK  MHRJKN  MHRJPL  MHRJX9  MHRJYP  MHRK1Y  MHRKJ4  MHRKK3  MHRKKP  MHRKM0  MHRKN4  MHRKN5  MHRKPX  MHRKPY  MHRKVX  MHRKW0  MHRL1L  MHRL1T  MHRL2N  MHRL2P  MHRL9P  MHRLJX  MHRLKE  MHRLL2  MHRLL9  MHRLLK  MHRLM2  MHRLNP  MHRLNR  MHRLNT  MHRLNV  MHRM2L  MHRM2M  MHRMT8  MHRMTN  MHRMVT  MHRMVX  MHRMYT  MHRMYW  MHRMYX  MHRMYY  MHRNAT  MHRNAV  MHRNAW  MHRNAX  MHRNAY  MHRPKY  MHRPL1  MHRPL2  MHRPL4  MHRR46  MHRR8V  MHRR99  MHRR9A  MHRRD6  MHRRD7  MHRRD8  MHRRD9  MHRRDA  MHRV8D  MHRVE3  MHRVE4  MHRVE5  MHRVE7  MHT00Y  MHT01P  MJE8R8  MJE8TH  MJE8TX  MJE90N  MJE910  MJE93E  MJE94Y  MJE9NM  MJE9NN  MJE9NR  MJE9P0  MJE9TH  MJE9Y7  MJE9Y8  MJEA4K  MJEA8J  MJEAVR  MJEAW1  MJED3V  MJED63  MJED89  MJEDEA  MJEDW9  MJEH6Y  MJEH7V  MJEH8X  MJEHDM  MJEJJN  MJEJL9  MJEJP1  MJEJVN  MJELXR  MJEM03  MJEML3  MJEML4  MJEML5  MJEMY8  MJEN0M  MJEP79  MJEXAJ  MJEXD4  MJEXMW  MJEXNY  MJEXRJ  MJEY1R  MJEYT8  MJEYT9  MJEYTA  MJH007  MJH01M  MJH0A2  MJH172  MJH173  MJH174  MJH175  MJH176  MJH177  MJH1LY  MJH2J9  MJH2JA  MJH2JD  MJH2JE  MJH2JH  MJH850  MJH851  MJH85A  MJH85H  MJH85J  MJH85K  MJH8K4  MJKM0J  MJKM0K  MJKM86  MJKM8H  MJKM8T  MJKMP9  MJKMPM  MJKMPN  MJKN56  MJKN58  MJKN5D  MJKN9E  MJKN9H  MJKN9J  MJKNK7  MJKNK8  MJKNK9  MJKPNN  MJKPRY  MJKPY5  MJKR14  MJKR39  MJKW3N  MJKWR4  MJKWR5  MJKWR6  MJKWR7  MJKWR8  MJKWY0  MJKWY2  MJKWY3  MJKWYX  MJKWYY  MJKX48  MJKX70  MJKXE8  MJKXKV  MJL0PD  MJL2W9  MJL2X0  MJL37R  MJL455  MJL48E  MJL4H0  MJL4M4  MJL59D  MJL5AV  MJL5YM  MJL62R  MJL6NT  MJL6NX  MJL7AH  MJL7D5  MJL7DT  MJL7DX  MJL7JR  MJL7W3  MJL84T  MJL8RE  MJL92J  MJL94H  MJL9KX  MJL9L3  MJL9PA  MJL9RK  MJLAD4  MJLAEL  MJLAL0  MJLALW  MJLAML  MJNDV6  MJNDWN  MJNDX9  MJNDXD  MJNDYW  MJNE3D  MJNE6R  MJNE92  MJNEJN  MJNH75  MJNHHW  MJNHKK  MJNHP0  MJNJ4N  MJNK11  MJNK22  MJNK6T  MJNKER  MJNKH4  MJNKP9  MJNKXP  MJNL0P  MJNLJN  MJNM2A  MJNM51  MJNM5Y  MJNM92  MJNMMA  MJNMX5  MJNN6W  MJNN8J  MJNNHR  MJNNJ0  MJNNK5  MJNNY5  MJNPA9  MJNRH0  MJNRKJ  MJNTLK  MJNTNH  MJNTT2  MJNTX5  MJNV3X  MJNVA7  MJNVNH  MJNW77  MJNWPV  MJNX8H  MJNX9K  MJNXNY  MJNY6E  MJNY7T  MJNY9H  MJNYJX  MJNYV8  MJNYVA  MJP0J4  MJP0TR  MJP0VD  MJP0XH  MJP108  MJP1M5  MJP1WW  MJP1Y9  MJP2N5  MJP365  MJP3JN  MJP41R  MJP46X  MJP48W  MJRWJY  MJRWRN  MJRX2A  MJRX7D  MJRXJ1  MJRXKA  MJRXXK  MJRYJD  MJT095  MJT0V1  MJT2R7  MJT321  MJT3EW  MJT3HA  MJT3N0  MJT3T2  MJT47P  MJT48X  MJT4LK  MJT4V6  MJT54M  MJT54X  MJT5R6  MJT675  MJT7D3  MJT7R9  MJT88N  MJT8PP  MJT904  MJT988  MJT9YJ  MJTAEE  MJTDL2  MJTE0V  MJTE1X  MJTEA1  MJTEEL  MJTHHN  MJTHJN  MJTHNL  MJTJ13  MJTJ9P  MJTJM1  MJTJMY  MJTK0P  MJTKHM  MJTKV3  MJTL08  MJTL0W  MJTLJ4  MJTLWT  MJTLYL  MJTNL6  MJTP6A  MJTP6M  MJTP98  MJTR2P  MJTRM3  MJTRNY  MJTT0E  MJTT0T  MJTTD2  MJTTLM  MJTTRT  MJTTTL  MJTW75  MJTWDE  MJTWYA  MJTX2H  MJTX4X  MKHMJE  MKJMJE  MKJMJEX1  MKVMKE  MLHMKE  MLHMKEX1  MLJMLE  MLLMEE  MLLMLE  MLMMEE  MLMMLE  MLNMEE  MMLMEA  MMMMEA  MMNMEA  MNTMME  MP3MLE  MR3MRD  MR4MRD  MR5MRD  MRKMAE  MTTMRD  MTVMRD  MTWMKE  MTWMRD  MTXMKE  MV0MKE  MV0MLE  N02MDD  N05MDD  N07MDD  N07MDDX1  N08MKE  N09MKE  N24MAE  N29MDD  N29MDDX1  N2KMDD  N2KMDDX1  N2LMDD  N2MMDD  N2PMPD  N2RMME  N45MNE  N49MPD  N4EMNE  N4HMPD  N4XMNE  N74MAE  N79MKA  N7KMKA  N8NMPD  N8NMPDX1  N8PMJE  N91MKE  N9LMKA  N9RMKA  N9VMKA  N9VMME  N9XMKA  N9YMKA  NA3MKA  NAXMPD  NDYMKE  NH5MKE  NH7MKE  NH8MKE  NHXMKE  NK5MKA  NKAMAE  VYWMMD  VYWMMDX1  VYYMMD  W00MMD  W1YMAE  W20MMD  W2HMMD  W2JMMD  W2JMMDX1  W2TMMD  W2XMED  W2XMEDX1  W3RMMD  W4EMJD  W4HMJD  W4JMJD  W4KMJD  W4LMJD  W4XMJD  W71MHE  W71MHEX1  W9TMEA  W9VMEA  W9WMEA  WALMHE  WANMHA  WAPMHA  WARMHA  WD8MHE  WDLMED  WDLMEDX1  WK5MED  WK5MEDX1  WKTMED  WKTMEDX1  WLDMHE  WLKMME  WM1MHE  WMDMAE  WMRMKE  WP0MAE  WP9MED  WP9MEE  WRHMED  WRJMED  WRMMED  WT5MED  WTXMME  WVJMNE  WVLMME  WVTMEE  WVVMEE  WVWMEE  WVXMEE  WVYMEE  WW0MEE  WW5MME  WXWMMA  WXWMMAX1  WYLMHA  WYMMEA  WYNMEA  WYPMEA  WYRMEA  X3LMKE  X3MMKE  X4JMAE  X5TMLD  X68MNE  X6DMNE  20716803  20716804  20716805  20716806  20716807  20716808  20722401  20722402  20722403  20722404  20722405  20722406  20722407  20722408  20783201  20783202  20783203  20783204  20783205  20783206  20783207  20783208  20819901  20819902  20819903  20819904  20819905  20848301  20848302  20848303  20848304  20848305  20848306  20850301  20850302  20850303  20922901  20922902  20922903  20934401  20934402  20934403  20934404  20967601  20967602  20967603  20967604  21029801  21029802  21029803  21029804  21029805  21109601  21109602  21109603  21109604  21109605  21136001  21136002  21136003  21136004  21136005  21136006  21136007  21136008  21185901  21185902  21185903  21185904  21185906  32664103  32664104  32664105  32729301  32729302  32729303  32729304  32729305  32799001  32799002  32799003  33117301  33117302  33117303  33117304  33234001  33234002  33234003  33234004  33640201  33640202  33640203  33640204  33640205  33968401  33968402  33968403  33968404  33968405  33968406  34015901  34015902  34280901  34280902  34280903  34280904  34280905  34280906  34304901  34363001  34363002  34711501  34711502  34711503  34711504  34711505  34850801  34850802  34850803  34850804  34850805  35041901  35041902  35041903  35041904  35066801  35066802  35234901  35234902  35234903  35234904  35617201  35617202  35617203  35661601  35661602  35661603  35661604  35661605  35724901  188MPE  189MTD  19362201X1  19362401X1  19362501X1  19362601X1  19362601X1X1  19506201X1  19506201X1X2  19506201X2  19549601X1  19549602X1  19634101X1  19634102X1  19634103X1  19634103X2  19634104X1  19696801X1  19696802X1  19696802X1X1  19696803X1  19696803X1X1  19719801X1  19719802X1  19719803X1  19791001X1  19791002X1  19791002X3  19791003X1  19791003X2  19791004X1  19929403X1  19929404X1  19972601X1  19972601X2  19972602X1  19972602X3  19972603X1  19995202X1  19995203X1  19995203X2  19995204X1  19995204X1X1  19995204X2  19TMEE  19VMEE  19WMEE  1EJMHE  1EKMHE  1ELMHE  1ELMTD  1EP02X  1EP03A  1EP03AX1  1EP090  1EP091  1EP091X1  1EP092  1EP093  1EP094  1EP17T  1EP17W  1EP17X  1HAMTD  1HJMJD  1HRMEE  1HXMJD  1J5MJD  1J6MJD  1JE06R  3NTMLE  3PRMKD  3PTMKD  3PWMPE  3PWMPEX1  3TPMED  3TPMLE  3W3MLE  3WPMDD  3WXMDD  3XDMDD  3XHMDD  3XKMDD  3XKMDDX1  3Y8MPE  419MKA  42WMLE  43KMPE  452MLE  461MPE  463MED  46YMTD  479MTD  47AMTD  485MMD  486MMD  487MMD  49NMKA  49XMMD  49XMMDX1  4ADMPE  4AJMED  4AYMLD  4AYMLDX1  4DEMHD  4DHMHD  4DJMHD  4DJMKA  4DKMKA  4E2MKA  4E3MKA  4ENMKA  4ETMKA  4ETMKAX1  4J7MNE  4J8MNE  4KDMDD  4KRMKA  4M8MPE  4MAMPE  4MMMPE  4MVMMD  4MYMED  4N0MED  4NRMLD  4T3MPE  4THMHE  4TJMHE  4TKMHE  4XJMHD  4Y2MED  4Y5MHE  4Y6MHE  4Y7MHE  4Y8MKA  505MEE  506MEE  507MEE  517MLD  54RMLD  85LMME  85MMME  86149403A  86149404A  87EMRD  8A5MRD  8A5MRDX1  8AKMRD  8AMMRD  8J7MHA  8J8MHA  8KTMJE  8LEMDE  8LEMED  8LKMJE  8M8MJE  8N3MME  8N7MME  8NMMEE  8NPMEE  8NRMEE  8NTMEE  8NTMJE  8NVMEE  8NWMEE  8P1MJE  8RHMDE  8RVMJE  8TPMPD  8TRMPD  8TRMPDX1  8V8MPD  8VKMME  8VWMPD  8VYMME  8WJMLA  8WKMJE  8WXMLA  8WYMLA  8X1MLA  8X1MLAX1  8X2MLA  8X2MLAX1  8X6MPD  8XDMJE  8XEMJE  8XLMJE  8XRMHA  8XTMHA  902MDE  907MRD  90JMJE  90WMEE  91DMRD  91XMJE  92993802A  92RMND  92RMNDX1  92YMND  930MND  931MND  931MPD  937MDE  949MDA  94DMDA  94EMDA  94HMDA  96TMJA  96VMJA  96WMJA  H69MRD  H7HMHE  H7WMND  H7XMND  H8RMTD  H8RMTDX1  H8TMTD  H8VMTD  H8VMTDX1  H8WMTD  H8WMTDX1  H8XMND  H8XMTD  H8YMTD  H93MKE  H9MMTD  H9WMKE  HA7MHE  HALMAE  HAMMAE  HANMAE  HAXMKE  HDKMPD  HDKMPDX1  HJ0MPD  HK5MDA  HK6MDA  HK7MDA  HK8MDA  HMPMLD  HMVMND  HR4MKE  HRRMND  HT6MPD  J01MMD  J0EMEE  J0HMEE  J0JMEE  J0KMEE  J0LMEE  J0MMEE  J18MKE  J1AMPD  J26MLD  J33MTD  J34MTD  J35MTD  J35MTDX1  J36MTD  J4WMKE  J55MKE  J56MKE  J69MLD  J6HMMD  J6RMMD  J7LMLE  J7PMLE  J85MLE  JAKMMD  JEHMKD  JEPMTD  JJ0MAE  JJ1MAE  JJ2MAE  JJ3MLE  JJ9MLD  JJDMLD  JKJMLD  JKLMAE  JKMMAE  LHKMPDX1  LN7MEE  LN8MEE  LN9MEE  LNNMLE  LP0MLE  LP5MHE  LP5MHEX1  LPRMMD  LRHMAD  LRHMADX1  LRJMAD  LRPMHE  LT1MLE  LT5MLE  LV4MME  LV7MME  LV8MME  LVDMMA  LVDMMAX1  LX0MME  LX1MME  LX2MME  LXJMJE  LXKMJE  LXLMJE  LYWMEE  M20MLE  M24MLE  M8TMJA  M8WMJA  M8XMJA  M99MMA  M9AMJA  M9VMJA  M9WMJA  M9XMJA  M9YMJA  MA0MJA  MEPE1X  MEPE1Y  MEPE20  MEPE21  MEPH28  MEPH2P  MEPH6P  MEPH6PX1  MEPH6R  MEPH7T  MEPH8V  MEPHLK  MEPHPK  MEPHV1  MEPHV4  MEPHV4X1  MEPJ5X  MEPJ5Y  MEPJ60  MEPJ84  MEPJ85  MEPJ9Y  MEPJD5  MEPJH6  MEPJM5  MEPJWY  MEPKWK  MEPKWL  MEPKWM  MEPLHD  MEPLHDX1  MET9N1  MET9VN  META00  META3E  META6T  META96  METD0R  METD23  METD23X1  METD23X2  METD38  METD38X1  METD38X2  METD4P  METD4R  METDD3  METDKW  METDKWX1  METDKX  METE03  METE43  METE4E  METE5K  METE68  METE6R  METE86  METE87  METE8A  METE9P  METEHX  METEHY  METEJ0  METHDA  METLLR  METN07  METN17  METN4J  MH8MAE  MH8MJE  MH9MAE  MH9MAEX1  MH9MJE  MHA4LD  MHA4M9  MHA4M9X1  MHA4MA  MHA5PA  MHA5PD  MHA5PE  MHA5R5  MHA626  MHA63V  MHA686  MHA698  MHA698X1  MHA6E7  MHA6V5  MHA6VD  MHA6VDX1  MHA711  MHA71R  MHA71RX1  MHA768  MHA774  MHA77J  MHA7AK  MHA7PK  MHA7XN  MHA7YM  MHA812  MHEMEE  MHEMEEX1  MHEMH8  MHEMHL  MHEMHLX1  MHEMHLX2  MHEMJY  MHEML3  MHEMNR  MHEMPT  MHEN1N  MHEN1NX1  MHEN82  MHEND9  MHENDA  MHENDAX1  MHENE4  MHENE5  MHEW53  MHEW71  MHEWD4  MHEWD4X1  MHEWJD  MHEX0M  MHEX28  MHEX28X1  MHEX31  MHEX32  MHEX3M  MHEXD2  MHEXD2X1  MHEXJH  MHEXJHX1  MHEXNV  MHEY90  MHEY91  MHEYDE  MHH18K  MHH18KX1  MHH1AJ  MHH1LK  MHH1VV  MHH564  MHH5HN  MHH5M0  MHH5RK  MHH5VV  MHH62A  MHH62AX1  MHH65A  MHH65V  MHH663  MHH67Y  MHH68E  MHH68EX1  MHH704  MHH730  MHH751  MHHAN6  MHHD69  MHHD6D  MHHDD0  MHHDED  MHHDEE  MHHDEEX1  MHHHKV  MHHJ1M  MHHJ55  MHHJJH  MHHJKT  MHKNYP  MHKP0L  MHKP0M  MHKP0N  MHKP10  MHKP1M  MHKP8L  MHKP9N  MHKPJA  MHKPJE  MHKPPD  MHKPR2  MHKPR8  MHKPRA  MHKPRD  MHKPRY  MHKPT3  MHKPVD  MHKPVM  MHKPYD  MHKPYE  MHKPYH  MHKR24  MHKR42  MHKR9P  MHKRJT  MHKRJTX1  MHKRPA  MHKRYW  MHKT3T  MHKT3V  MHKT3W  MHKT3Y  MHKT40  MHKTRD  MHKTRE  MHKV43  MHKV44  MHKV7T  MHKW2A  MHKW42  MHL009  MHL06K  MHLE5T  MHLE5V  MHLE5W  MHLE5X  MHLE5Y  MHLE61  MHLE61X1  MHLH73  MHLHH3  MHLHH7  MHLHH8  MHLHJA  MHLHJV  MHLHKD  MHLHLP  MHLHM1  MHLHM2  MHLHM4  MHLHM5  MHLHNH  MHLHP3  MHLHP4  MHLHRE  MHLHRH  MHLHRN  MHLHRP  MHLJ42  MHN9E7  MHN9ML  MHN9MM  MHN9MN  MHNAL9  MHNALA  MHNALD  MHNATV  MHNATW  MHNATX  MHNDEM  MHNDPH  MHNDPJ  MHNDPP  MHNDWN  MHNDX7  MHNE0N  MHNE0P  MHNE0R  MHNE80  MHNE85  MHNE86  MHNH54  MHNH56  MHNH7E  MHNH8L  MHNP44  MHNP6Y  MHNP70  MHNP9Y  MHNPA1  MHNPAA  MHNPAP  MHNPDE  MHNPDJ  MHNPDK  MHNPM4  MHNPM5  MHNPM6  MHNPM7  MHNRED  MHNREE  MHNREH  MHNRNA  MHNRND  MHNRNE  MHNT15  MHNTEH  MHNTKT  MHNTKX  MHNTMX  MHNTNP  MHNTPT  MHNTPV  MHNTPW  MHNTXY  MHNV2A  MHNV4M  MHNV4N  MHNV4P  MHNV88  MHNV89  MHNV8A  MHP418  MHP419  MHP41A  MHP41D  MHP41E  MHP41H  MHP41J  MHT07H  MHT0DY  MHT0T1  MHT0YN  MHT0YR  MHT124  MHT125  MHT12V  MHT153  MHTDTD  MHTE42  MHTEDD  MHTEE0  MHTEHK  MHTEPM  MHTEX2  MHTEYX  MHTH0N  MHTJNY  MHTK94  MHTK95  MHTKA4  MHTKHM  MHTKLV  MHTKPH  MHTKT8  MHTKVV  MHTLNW  MHTLR0  MHTLR1  MHTLYL  MHTM2Y  MHTM9N  MHTMKL  MHTMKM  MHTMR2  MHTMR3  MHTN07  MHTN2L  MHTNAD  MHTNDM  MHTNPW  MHTP96  MHTP9D  MHTPA9  MHTPAA  MHTPJP  MHTPLN  MHTPW9  MHTPWP  MHTR1J  MHTR27  MHTR3D  MHTR80  MHTRJ8  MHTRKJ  MHTRKP  MHTRKX  MJA011  MJA01L  MJA064  MJA0AE  MJA0T0  MJA15X  MJA17K  MJA1DH  MJA1K4  MJA1TK  MJA1W4  MJA22L  MJH8MK  MJH8NT  MJH8RJ  MJH915  MJH916  MJH917  MJH97V  MJH97W  MJH97X  MJH98W  MJH9DT  MJH9KA  MJH9R0  MJH9T0  MJHA2P  MJHA2T  MJHA2V  MJHA8E  MJHA8H  MJHA8J  MJHELE  MJHELH  MJHER7  MJHER8  MJHER9  MJHEYW  MJHEYX  MJHEYY  MJHH8N  MJHK3E  MJHK3H  MJHK3J  MJHKET  MJHKEW  MJHKEX  MJHKK7  MJHKNV  MJHKPR  MJHL14  MJHL15  MJHL16  MJHL28  MJHL4R  MJHL51  MJHL85  MJHLJN  MJHLJP  MJHLJT  MJHLMJ  MJHLW7  MJHLW8  MJHLW9  MJHM0L  MJHM0M  MJHM0N  MJHM2N  MJHM2P  MJHM2R  MJHN27  MJHN8A  MJHND8  MJHNEK  MJHNYK  MJHNYL  MJHNYM  MJHP6D  MJHP9J  MJHP9L  MJHP9M  MJHPDR  MJLARY  MJLATY  MJLAXK  MJLD0J  MJLD61  MJLD9M  MJLD9P  MJLD9R  MJLD9T  MJLD9V  MJLD9W  MJLD9X  MJLE71  MJLEA4  MJLEA5  MJLEA6  MJLH22  MJLJ01  MJLJEL  MJLJJM  MJLJL8  MJLJXR  MJLJYJ  MJLK77  MJLKDR  MJLL2X  MJLL61  MJLL8D  MJLLHY  MJLLRV  MJLM4M  MJLM5Y  MJLMHY  MJLMMA  MJLMMAX1  MJLMXW  MJLN55  MJLNLJ  MJLPE7  MJLPR0  MJLPVN  MJLPY7  MJLR6J  MJLRTL  MJLT1A  MJLT57  MJLT6H  MJLTH1  MJLTTN  MJLTW5  MJLV0M  MJLV8J  MJLVH0  MJLW05  MJLW13  MJLWR0  MJLWYD  MJLX9E  MJLXMV  MJLXN6  MJLXW0  MJLXY9  MJLY8J  MJLYDV  MJLYE6  MJLYE9  MJLYTK  MJM0DW  MJM0RP  MJM151  MJP4AR  MJP4JR  MJP4RR  MJP4W9  MJP51E  MJP54L  MJP578  MJP595  MJP5L5  MJP5LP  MJP5YD  MJP68T  MJP6DH  MJP6EA  MJP6JV  MJP6PP  MJP6PV  MJP6TJ  MJP6WX  MJP76A  MJP85R  MJP8KP  MJP8PM  MJP8T3  MJP922  MJP935  MJP96T  MJP98T  MJP9HM  MJP9WA  MJPA84  MJPAX9  MJPAYA  MJPD0K  MJPD30  MJPD3T  MJPD6R  MJPDYT  MJPEK6  MJPEK9  MJPEX4  MJPH1R  MJPH1Y  MJPH29  MJPH55  MJPH91  MJPHKX  MJPHL5  MJPHVL  MJPJ25  MJPJJM  MJPJWE  MJPK94  MJPK9R  MJPKP2  MJPL0R  MJPL8D  MJPLE7  MJPLLL  MJPM1J  MJPM2R  MJPM64  MJPM75  MJPMKW  MJPMMA  MJPN1T  MJPN4P  MJPN5D  MJPN7T  MJPNXN  MJTX53  MJTX7W  MJTXD2  MJTXVN  MJTY8W  MJTY9D  MJTYMM  MJTYR2  MJTYV6  MK4MMA  MK6MJE  MKA06H  MKA0DR  MKA0V4  MKA12V  MKA1H5  MKA2DM  MKA2VD  MKA303  MKA32P  MKA34J  MKA388  MKA3AE  MKA3E6  MKA4R2  MKA53Y  MKA56R  MKA59J  MKA5VA  MKA5YY  MKA64N  MKA668  MKA6PY  MKA6YN  MKA7J2  MKA7L6  MKA7L8  MKA7P0  MKA7R9  MKA7Y1  MKA84V  MKA85L  MKA8ML  MKA8R0  MKA8X6  MKA9HV  MKAA8W  MKAA93  MKAAL0  MKAATX  MKAD2M  MKAD9D  MKAJ55  MKAJH6  MKAJHL  MKAJKM  MKAJL9  MKAJLP  MKAJN6  MKAJVL  MKAK6J  MKAK80  MKAKJE  MKAKPH  MKAKTE  MKAL82  MKALKR  MKALMP  MKALWH  MKAMHW  NKAMJE  NKDMAE  NKDMPD  NKEMAE  NKHMAE  NLRMJE  NLVMJE  NN1MKE  NNWMEA  NP4MEA  NPEMKE  NPMMKA  NPNMKA  NPPMKE  NPXMKE  NPXMPD  NXAMHE  NY2MHE  NY8MHE  NYHMKE  NYJMKE  NYKMKE  P08MKE  P0AMPD  P14MPD  P17MHE  P1NMHE  P1RMHE  P1TMHE  P1VMHE  P21MKD  P21MKDX1  P37MME  P39MME  P3AMME  P3VMKE  P3VMKEX1  P3XMEA  P3YMEA  P52MKD  P62MPD  P89MPD  P89MPDX1  P8XMKA  P8YMKA  P95MRD  P9EMKD  PE9MKE  PK9MRD  PK9MRDX1  PKHMNE  PL3MKD  PL9MKE  PNEMKE  PPLMMD  PPNMMD  PPVMMD  PRLMHE  PRMMHE  PRMMRD  PRPMRD  PRRMRD  PRTMRD  PTPMEA  PTRMEA  PTTMEA  PTVMEA  PVDMKE  PWHMNA  PXXMHD  X6RMNE  X6TMNE  X8DMNE  X95MAE  X9TMLD  XA1MKE  XE7MEA  XE8MEA  XE9MEA  XEWMJE  XH9MJE  XHAMJE  XJWMLD  XJWMLDX1  XJXMLD  XKJMKE  XKXMLD  XKYMLD  XL0MLD  XL1MLD  XL4MKE  XL8MLD  XN0MNE  XR2MHA  XR3MHA  XR4MHA  XR5MHA  XR6MHA  XYEMNE  Y69MNE  Y6KMNE  Y7XMNE  Y84MNE  Y8HMNE  Y8MMNE  Y8PMJE  YA5MJE  YH4MJE  YJDMJE  YJVMJE  YJYMJE  YJYMJEX1  YK0MJE  YK1MJE  YK1MJEX1  10176801  10176802  10353401  10353402  10852701  10852702  10937101  10937102  11521901  11521902  11970001  11970002  11970003  12178201  12178202  12178203  12664601  12664602  12664603  12664604  12864101  12864102  13433701  21208601  21208602  21208603  21208604  21208605  21218501  21218502  21218503  21220101  21220102  21220103  21220104  21220105  21220106  21263801  21263802  21263803  21263804  21263805  21296801  21296802  21296803  21296804  21296805  21437501  21437502  21437503  21437504  21437505  21437506  21437507  21489401  21489402  21489403  21489404  21489405  21489406  21654001  21654002  21654003  21654004  21654005  21668301  21668302  21668303  21668304  21688101  21688102  21696101  21696102  21696103  21762301  21762302  222455",
      "center_classification_date": "20180702",
      "distribution_pattern": "Distributed nationwide. Foreign distribution to Australia, Canada, Chile, China, Brazil, France, Germany, India, Italy, Japan, Malaysia, Netherlands, New Zealand, Poland, Portugal, Romania, Singapore, South Africa, Spain, Sweden, Switzerland, and United Kingdom.",
      "state": "NJ",
      "product_description": "LFIT Anatomic CoCr Femoral Heads,  Catalog No. 6260-9-136",
      "report_date": "20180711",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Howmedica Osteonics Corp.",
      "recall_number": "Z-2295-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80059",
      "termination_date": "20200508",
      "more_code_info": "01  22245502  22245503  22245504  22245505  22245506  22295201  22295202  22295203  22295204  22295205  22295206  22349302  22349303  22349304  22400901  22400902  35724902  35724903  35724904  35855301  35855302  35855303  35922001  35922002  35922003  35922004  36219401  36219402  36219403  36219404  36219405  36219406  36219501  36219502  36219503  36219504  36305801  36305802  36957301  37020301  37020302  37020303  37020304  37020305  37020306  37063601  37064701  37064702  37064703  37073301  37073302  37221301  37221302  37221303  37221304  37221305  37381101  37381102  37381103  37381104  37381105  37657101  37657102  37657103  37657104  37657105  37657106  37657107  38004801  38004802  38004803  38004804  38004805  38004806  38044601  38044602  38044603  38272801  38272802  38272803  38272804  38327001  38327002  38327003  86149401  86149402  1JE09A  1KVMTD  1L7MNA  1L7MNAX1  1L8MNA  1L8MNAX1  1LEMPE  1LEMPEX1  1M2MNA  1M2MNAX1  1M6MDE  1M8MDE  1M9MDE  1NAMPE  1TXMLE  1VYMTD  1W3MTD  1W5MTD  1WAMPE  1WLMRD  1YRMPE  20081701X1  20081701X1X1  20081701X2  20081702X1  20081702X1X1  20081702X2  20081703X1  20081703X1X1  20081703X3  20105201X1  20105201X2  20105202X1  20105202X2  20105202X2X1  20155201X1  20155201X1X1  20155201X2  20155201X2X1  20155202X1  20155203X1  20155204X1  20155204X1X1  20155204X2  20155205X1  20155205X2  20155206X1  20155206X2  20199601X1  20199601X1X1  20199601X2  20199602X1  20199603X1  20199604X1  20199604X1X1  20199605X1  20199605X2  20200401X1  20200401X1X1  20484201X1  20484201X1X1  20484201X2  20484202X1  20484203X1  20484203X1X1  20484204A  20484204X1  20484205X1  20484206X1  20484206X1X1  54RMLDX1  55JMPD  561MPD  56MMPD  574MHD  57TMLE  580MLE  587MLE  58NMLE  58TMLE  5ADMED  5ANMND  5ANMNDX1  5D5MHD  5D6MHD  5DAMHE  5DDMHE  5DEMHE  5ENMLD  5ENMPD  5H5MLD  5HMMLD  5HRMLD  5J2MPD  5J2MPDX1  5J9MDD  5JXMHE  5JXMHEX1  5KJMND  5KKMND  5KKMNDX1  5LTMME  5LVMME  5LWMME  5LXMME  5MKMHA  5MLMHA  5MMMHA  5MNMHA  5MVMED  5N1MPD  5N1MPDX1  5N2MKD  5NAMJA  5NAMND  5NDMJA  5NDMND  5NEMJA  5R1MHE  5R1MND  5R2MND  5R3MND  5R9MJD  5RVMJD  5RYMJD  5VEMED  5VJMEE  5VMMEE  5VTMEE  5VXMPD  5VYMEE  5XPMME  5XRMME  5YVMPD  60JMED  60XMKD  60XMKDX1  60YMKD  60YMKDX1  616MKD  973MJA  992MRD  9AAMDE  9AHMDE  9DEMPD  9EHMDA  9EWMDA  9EYMDA  9H0MDA  9H0MJD  9H0MJDX1  9H8MDA  9H9MDA  9HAMDA  9HEMPD  9J2MJD  9J4MJD  9JJMRD  9LAMAE  9M3MAE  9MAMAE  9MAMAEX1  9T4MPD  9T6MPD  9W8MPD  A02MPD  A03MPD  A03MPDX1  A04MPD  A0XMRD  A1EMAE  A43MAE  A4JMHE  A4KMHE  A4LMHE  A4MMHE  A4NMHE  A4PMHE  A56MJA  A9NMJA  A9RMJA  AA3MRD  AA6MHA  AA9MHA  AADMHA  AARMRD  AAYMJA  AD2MJA  AD3MJA  AD5MRD  AD5MRDX1  AD6MRD  AD6MRDX1  ADEMJA  ADHMJA  ADLMRD  ADNMAE  AE0MRD  AHHMJD  AHHMJDX1  AHKMJD  AHMMJD  AHNMJD  AHPMJD  AHTMJD  AHVMJD  ALTMAE  ALTMDE  ALVMAE  ALVMDE  JKNMAE  JL3MLD  JL4MAE  JL6MDD  JL7MDD  JL8MDD  JLPMLE  JM9MNE  JMAMNE  JMDMDD  JMDMNE  JMEMDD  JMEMDDX1  JMHMDD  JMHMDDX1  JMJMDD  JMKMDD  JN4MMD  JN6MPA  JN6MPAX1  JN6MPAX2  JN7MPA  JN7MPAX1  JN7MPAX1X1  JN8MPA  JN8MPAX1  JN9MPA  JNAMPA  JNAMPAX1  JRHMLE  JRKMLE  JRLMLE  JRNMTD  JRNMTDX1  JTKMMD  JX1MPA  JX2MPA  JX3MPA  JX4MPA  JX4MPAX1  JX4MPAX2  JXVMDE  JY0MKD  JYXMTD  K14MTD  K16MTD  K17MTD  K29MKE  K2DMKE  K2EMKE  K2MMTD  K38MND  K3HMLD  K3HMLDX1  K3JMTD  K3LMND  K3LMTD  K3MMTD  K3NMTD  K3PMND  K3PMTD  K3RMTD  K3TMTD  K3VMTD  K3WMTD  K3XMTD  K44MEE  K45MND  K4VMJA  K4WMJA  MEPLHT  MEPLHV  MEPLLM  MEPLM2  MEPLP0  MEPLP0X1  MEPLP4  MEPLP4X1  MEPLP4X2  MEPLP5  MEPLP6  MEPLP6X1  MEPMME  MEPMTP  MEPRHP  MEPRN6  MEPRN6X1  MEPRN6X1X1  MEPRY3  MEPRY3X1  MEPT6W  MEPT78  MEPT78X1  MEPT7R  MEPTNA  MEPTYD  MEPV1J  MEPVD6  MEPVDD  MEPVDK  MEPX8W  MEPXX0  MEPXYE  MEPXYEX1  MEPXYEX2  MEPY33  MEPY35  MEPY3E  MEPYHW  MEPYK0  MEPYK4  MEPYK4X1  MEPYK6  MER03W  MER053  MER053X1  MER080  MER081  MER081X1  MER0MT  MER0MTX1  MER0MV  MER0MY  MER15W  MER15X  MER15Y  MER15YX1  MER1D9  MER1DA  MER1DD  MER1DDX1  MER1DH  MER1DHX1  MER1EK  MER1JE  MER1LJ  MER1VE  MER213  MER21T  MER21TX1  MHA81A  MHA81X  MHAEEA  MHAEK4  MHAEW0  MHAEW0X1  MHAH2E  MHAH5N  MHAH9V  MHAHAW  MHAHDT  MHAHEK  MHAHEKX1  MHAHKL  MHAHP9  MHAHVE  MHAHWR  MHAJ47  MHAJ8Y  MHAJLH  MHAJNL  MHAJP5  MHAJRD  MHAJW5  MHAK0W  MHAK3T  MHAKH4  MHAKH4X1  MHAKM6  MHAKWE  MHAKYT  MHAL08  MHAL0M  MHAL12  MHAL3T  MHAL7L  MHAL7P  MHALM5  MHAMAE  MHAMJE  MHAPNX  MHAPT9  MHAR1W  MHAR34  MHARWD  MHARY7  MHAT3W  MHATDA  MHATVL  MHATVLX1  MHATY1  MHAV1V  MHAV8M  MHAVND  MHAVPE  MHAVR1  MHAVR5  MHAVRV  MHD010  MHD036  MHD1WN  MHD20K  MHD20KX1  MHD26W  MHD272  MHD281  MHD28P  MHD4YK  MHD53P  MHD568  MHHJNT  MHHJP9  MHHJPA  MHHJPE  MHHJPH  MHHJTV  MHHJTW  MHHJY9  MHHK3A  MHHK3M  MHHK54  MHHK5P  MHHK8J  MHHKJE  MHHKK3  MHHKM2  MHHKMM  MHHKND  MHHKT1  MHHKVR  MHHKW3  MHHL08  MHHL2V  MHHL56  MHHL58  MHHL7H  MHHL7X  MHHLKK  MHHLNR  MHHLPD  MHHLPT  MHHLVY  MHHLX5  MHHLYP  MHHM1J  MHHMA0  MHHMA6  MHHMAH  MHHMJY  MHHMT8  MHHMTA  MHHMX8  MHHMXX  MHHN0T  MHHN0X  MHHN12  MHHN14  MHHN1N  MHHN1P  MHHN21  MHHN21X1  MHHN4V  MHHN4X  MHHN4Y  MHHNJK  MHHNLP  MHHNLR  MHHP9Y  MHHPA0  MHHPA1  MHHPMW  MHHRKD  MHHRKH  MHHRKHX1  MHHTR8  MHHTTE  MHHTVW  MHHTWH  MHJ0NR  MHJ0T5  MHLJ4T  MHLJ4W  MHLJ4WX1  MHLJ7X  MHLJA8  MHLJA9  MHLJAA  MHLJAD  MHLJAH  MHLJKH  MHLJX4  MHLK19  MHLK1V  MHLK21  MHLK23  MHLK2L  MHLK2P  MHLK2Y  MHLK34  MHLK4M  MHLK5M  MHLK5N  MHLK7R  MHLKHX  MHLKHY  MHLKJ0  MHLKJ0X1  MHLKJ6  MHLKNW  MHLKP1  MHLKRK  MHLL0R  MHLL3E  MHLL3H  MHLL3J  MHLL40  MHLL49  MHLL4D  MHLL59  MHLNEN  MHLNER  MHM1L7  MHM1L7X1  MHM1L8  MHM1PH  MHM1R5  MHM1RA  MHM1RJ  MHM1RV  MHM1XJ  MHM1Y9  MHM1YA  MHM1YD  MHM26W  MHM26X  MHM2T0  MHM2VR  MHM2WP  MHM2X4  MHM2XD  MHM2YA  MHM30W  MHM30X  MHM30Y  MHM310  MHM311  MHM3TV  MHM40E  MHM419  MHM41M  MHP41L  MHP41M  MHP41N  MHP41P  MHP41T  MHP41TX1  MHP7KE  MHP7KH  MHP7NT  MHP7NV  MHP7NW  MHP7NX  MHP7P0  MHP7P1  MHP7P2  MHP8RD  MHP8T4  MHP8X5  MHP8X6  MHP8X7  MHPDE2  MHPDE3  MHPDE4  MHPE67  MHPE68  MHPE69  MHPE6A  MHPE6D  MHPE6E  MHPH4Y  MHPH50  MHPH51  MHPH52  MHPH53  MHPH54  MHPH55  MHPH8W  MHPJ76  MHPJ77  MHPJ78  MHPJ79  MHPJ7A  MHPJ7D  MHPJWH  MHPJWK  MHPJX4  MHPKV3  MHPKV5  MHPKV7  MHPL59  MHPL5A  MHPL5D  MHPLRD  MHPLRE  MHPLRH  MHPLRJ  MHPLRK  MHPMDD  MHPMDE  MHPMDH  MHPMDJ  MHPMDL  MHPMDM  MHPN76  MHPN77  MHPN78  MHPN78X1  MHPN79  MHPN7D  MHPN7E  MJA279  MJA2HX  MJA2K6  MJA2KV  MJA2KW  MJA2M0  MJAAH9  MJAAJV  MJAAP5  MJAAWM  MJAD0A  MJAD25  MJAD9P  MJADEW  MJADH4  MJAEWM  MJAHVH  MJAHWT  MJAHX7  MJAJ0L  MJAJ2A  MJAJDJ  MJAJN9  MJAJTM  MJAK32  MJAK3T  MJAK55  MJAK68  MJAK8X  MJAKDA  MJAKKD  MJAKM0  MJAKVJ  MJAKXM  MJAL1P  MJAL3R  MJAL5D  MJAL93  MJALJR  MJALN9  MJALT2  MJAM3W  MJAM5N  MJAM6D  MJAM6T  MJAMAV  MJAMDK  MJAMMA  MJAMME  MJAW37  MJAW38  MJAW39  MJAW3A  MJAW3D  MJAW3E  MJAW3H  MJD0JJ  MJD0JK  MJD0JL  MJD6MK  MJD6MM  MJD6MR  MJD6MV  MJD6MX  MJD6MY  MJD7KJ  MJD7L4  MJD7LA  MJD7LK  MJD7LP  MJHPDT  MJHPDW  MJHPLL  MJHPLM  MJHPLP  MJHPP7  MJHPVE  MJHR1W  MJHR21  MJHR22  MJHR29  MJHR4J  MJHR5E  MJHR5H  MJHR5N  MJHR5P  MJHTPH  MJHTPJ  MJHTPK  MJHV51  MJHV52  MJHV53  MJHVAX  MJHVAY  MJHVD1  MJJ039  MJJ03D  MJJ03H  MJJ07A  MJJ0AJ  MJJ0AW  MJJ0KP  MJJ0MW  MJJ13V  MJJ13W  MJJ13X  MJJ1LN  MJJ1LR  MJJ1XR  MJJ1XT  MJJ1XW  MJJ233  MJJ23D  MJJ23X  MJJ23Y  MJJ240  MJJ251  MJJ277  MJJ5T1  MJJ682  MJJ696  MJJ6H8  MJJ6JN  MJJ6LH  MJJ6PM  MJJ6XH  MJJ7JR  MJJ7M1  MJJ7T5  MJJ7X7  MJJ82X  MJJ83N  MJJ8L2  MJJ8R6  MJJDMX  MJJDTM  MJJDTN  MJJHDK  MJJHJN  MJJHNT  MJM15E  MJM17M  MJM1DY  MJM1L3  MJM2WV  MJM3DP  MJM3P4  MJM4EA  MJM4P1  MJM4XH  MJM4XW  MJM52V  MJM59V  MJM5ER  MJM5R8  MJM613  MJM62D  MJM642  MJM67D  MJM6X1  MJM744  MJM82R  MJM891  MJM8MW  MJM8N2  MJM8VE  MJM8Y2  MJM904  MJM984  MJM98D  MJM99H  MJM9KN  MJMA3W  MJMAJP  MJMD55  MJMD6W  MJMDAT  MJMDM0  MJME0A  MJME3X  MJME8P  MJMEE8  MJMH4D  MJMHAE  MJMHLN  MJMHW5  MJMHX7  MJMJ8M  MJMJJX  MJMJK7  MJMJMN  MJMJRR  MJMK2T  MJMK2Y  MJMK31  MJMK70  MJMKLP  MJMKN3  MJMKXX  MJML5J  MJMLDW  MJMMMA  MJMMNV  MJMMY1  MJMN4T  MJMPP3  MJMR26  MJMRNW  MJMRPD  MJMRX3  MJPP0T  MJPP29  MJPP69  MJPP76  MJPP7X  MJPP88  MJPPLV  MJPR1L  MJPR6K  MJPRD0  MJPRLX  MJPRVN  MJPT7Y  MJPTML  MJPTW3  MJPV3R  MJPVDV  MJPVKM  MJPVRR  MJPW0H  MJPWP6  MJPX3N  MJPXL0  MJPXL7  MJPYYA  MJR06D  MJR0MV  MJR13V  MJR187  MJR19J  MJR1EA  MJR1PH  MJR1Y1  MJR1YY  MJR204  MJR243  MJR25E  MJR2ET  MJR2N4  MJR2R1  MJR2YM  MJR30D  MJR32V  MJR39V  MJR45K  MJR475  MJR48H  MJR4PE  MJR512  MJR57V  MJR595  MJR5EK  MJR5PM  MJR5TP  MJR5X7  MJR670  MJR6AH  MJR6D8  MJR6PW  MJR6Y3  MJR703  MJR70H  MJR71W  MJR749  MJR7X1  MJR7Y2  MJR802  MJR821  MJR8AJ  MJR8N1  MKAMYA  MKAN71  MKAN8M  MKANH1  MKANL8  MKANNA  MKANT2  MKANV2  MKANX5  MKAP3A  MKAP3W  MKAPM3  MKAPM4  MKAPR0  MKAPW2  MKAR43  MKAR98  MKARHY  MKARP6  MKAT6X  MKAT91  MKAT94  MKATHE  MKAVAW  MKAVDE  MKAW26  MKAW8R  MKAWJ4  MKAWYM  MKAXAH  MKAXP0  MKAXT7  MKAY35  MKAYLT  MKAYYJ  MKD06P  MKD0JW  MKD0KV  MKD0NP  MKD0RW  MKD16R  MKD19R  MKD1LJ  MKD1R9  MKD1XT  MKD2HV  MKD2K9  MKD2NW  MKD2VN  MKD32E  MKD33T  MKD3R7  MKD406  MKD4R1  MKD5AM  MKD5PE  MKD5VL  MKD5YN  MKD65V  MKD757  MKD762  MKD7D5  MKD7PP  MKD89H  MKD8A1  MKD8E3  MKD8K6  MKD8RV  MKD8T7  MKD8Y9  PY0MNE  PY2MKE  R01MKD  R0TMNA  R0VMNA  R0VMNAX1  R0WMNA  R0WMNAX1  R3MMHE  R3RMHE  R3VMHE  R4PMLD  R4XMKD  R4XMLD  R67MLD  R6DMLD  R85MLD  R90MEA  R91MEA  R92MEA  R93MEA  RA3MLD  RAEMHD  REVMJE  REVMLD  REWMJE  REXMJE  RHYMED  RKNMED  RLDMKA  RPMMHD  RPNMHD  RPRMHD  RR2MLD  RRHMHD  RRJMHD  RT3MHD  RT6MED  RT7MED  RT8MED  RT9MED  RTLMHD  RTMMHD  RTTMKA  RTVMKA  RV6MHD  RVDMHD  RWNMLD  RYJMMA  RYJMMAX1  T3NMJE  T4HMHE  T4HMHEX1  T4KMHE  T4LMHE  T4TMJE  T5JMJE  T5MMJE  T5PMJE  T5TMJE  T63MJE  T7TMJE  T8WMMA  T91MKA  T91MKAX1  T92MKA  TAEMJE  TAEMJEX1  TAHMJE  TAJMJE",
      "recall_initiation_date": "20180522",
      "postal_code": "07430-2006",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}