  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88229",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "PO Box 610",
      "postal_code": "18034-0610",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution   Foreign:  Canada\t  Australia\t  Japan\t  Hong Kong\t  Europe\t  Singapore",
      "recall_number": "Z-2294-2021",
      "product_description": "Soltive Premium Super Pulsed Laser System (TFL-PLS) & Pro Super Pulsed Laser System (TFL-SLS). Intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in urology, lithotripsy, gastroenterological surgery and gynecological surgery",
      "product_quantity": "408 units",
      "reason_for_recall": "Thermal injury following dusting and fragmenting treatment of ureteral stones when user exceeded the 20W standard presets",
      "recall_initiation_date": "20210630",
      "center_classification_date": "20210815",
      "termination_date": "20240126",
      "report_date": "20210825",
      "code_info": "Software Version 2.0  (all serial numbers)"
    }
  ]
}