      "country": "United States",
      "city": "Mahwah",
      "address_1": "325 Corporate Dr",
      "reason_for_recall": "The recalling firm has received a higher than expected number of complaints documenting femoral head/hip stem disassociation for certain sizes of femoral heads manufactured prior to March 4, 2011.",
      "address_2": "",
      "product_quantity": "",
      "code_info": "Lots  586MTA  587MTA  588MTA  588MTAX1  589MTA  5E8MMD  5E9MMD  5EAMMD  5HDMDE  5JKMMD  5JKMMDX1  5L9MPA  5LHMPA  5LJMPA  5LVMPA  5LVMPAX1  5LWMPA  5LXMPA  5MPMTD  5MRMTD  5MRMTDX1  5R0MND  5TDMMD  5V1MND  5V7MHD  5VHMHD  5VTMHD  5W0MHD  5WEMHD  5WKMHD  5WNMHD  5WTMHD  5WVMAE  5WXMAE  5WYMAE  5X0MAE  60RMAE  61DMME  61EMME  61HMME  61HMND  61JMME  61KMDE  61LMDE  61LMDEX1  61MMDE  62HMND  62HMNDX1  62LMMD  63YMAD  63YMADX1  640MAD  641MAD  642MAD  650MDE  65NMMD  662MMD  662MND  665MND  68PMMD  690MED  691MED  692MED  693MED  69DMTD  6A9MTD  6ARMAD  6HNMMD  6HYMAD  6J2MAD  EJAMHE  EJTMME  EJVMME  EK2MME  EKLMDD  EL3MME  ENNMRD  ETTMPA  ETXMPA  EVLMDD  EXRMRD  EXTMRA  EXVMRA  EXVMRAX1  EY1MRD  H0PMHE  H0RMDD  H0TMTD  H0VMTD  H0WMTD  H0XMTD  H0YMTD  H16MDD  H16MDDX1  H21MHE  H36MJE  H36MJEX1  H37MJE  H38MJE  H39MJE  H3HMHE  H50MRD  H51MRD  H52MRD  H79MHE  H90MTD  H91MTD  H92MTD  H9RMTD  H9TMTD  HH6MHE  HHLMMA  HHLMMAX1  HHMMMA  HHNMMA  HHPMMA  HHRMMA  HHRMMAX1  HJ4MAE  HJ5MAE  HJ6MAE  HJ7MAE  HL6MMA  HL7MMA  HL7MMAX1  HM7MHE  HMWMHE  HP9MJE  HPEMJE  HPHMJE  HPKMJE  HT6MLA  HT8MLA  HT9MLA  HV5MHE  HV7MME  HV8MME  HV9MME  HVAMME  HVEMRA  METHA2  METHAH  METHAHX1  METHAJ  METHAK  METHD3  METHD4  METHDD  METMPD  METMRL  METMRM  METMVP  METMWM  METMXW  METMY2  METMY6  METMYE  MHA754  MHA755  MHA756  MHA757  MHA8WA  MHA8WE  MHA8WH  MHA8WHX1  MHA8WJ  MHAHVH  MHAHW2  MHAJ1W  MHAJ89  MHAJM0  MHAJM2  MHAJP3  MHAJP4  MHAJP7  MHAJW9  MHAJX9  MHAK18  MHAK18X1  MHAK2L  MHAK4L  MHAK6R  MHANH1  MHANK3  MHANPT  MHANX5  MHAP0D  MHAP19  MHAP42  MHAP87  MHAPAY  MHAPDD  MHAPDE  MHAPDEX1  MHAPH4  MHDJR8  MHDK8H  MHDK8K  MHDK8L  MHDK8M  MHDK8N  MHDK8P  MHDK8R  MHDK8T  MHDK8V  MHDLAE  MHDLAW  MHDLDL  MHDLDX  MHDLEP  MHL76M  MHL76N  MHL76P  MHL7NA  MHL7ND  MHL7NE  MHL8AK  MHL8E4  MHL8E6  MHL8E7  MHL8EW  MHL8EX  MHL8HW  MHL8LN  MHL8V6  MHL8WR  MHL8WT  MHL8WV  MHL8WW  MHL97M  MHL9KV  MHL9V7  MHL9VP  MHL9VT  MHL9W2  MHL9WD  MHL9WE  MHL9X3  MHL9X5  MHL9X6  MHLA6Y  MHLA70  MHLA72  MHLA73  MHLA74  MHLAWP  MHLD04  MHLD06  MHLD13  MHLD14  MHLD15  MHLD23  MHLD24  MHLD2W  MHLD3E  MHLD3H  MHLD3J  MHLD3K  MHLD3L  MHLD3M  MHLDTL  MHLDXD  MHLDXE  MHLE23  MHMNM5  MHMNM5X1  MHMNV0  MHMNWL  MHMNXD  MHMNXE  MHMNXP  MHMNYA  MHMP1D  MHMP43  MHMP5W  MHMP9D  MHMP9DX1  MHMPA6  MHMPD7  MHMPHE  MJDAVY  MJDAX1  MJDD30  MJDD53  MJDD54  MJDDKJ  MJDDKK  MJDDKT  MJDDKV  MJDDKW  MJDDKX  MJDDMT  MJDDMV  MJDDMW  MJDDMX  MJDDPY  MJDEAM  MJDEJM  MJDEKJ  MJDENP  MJDEWP  MJE3M2  MJE3R4  MJE3V7  MJE3VL  MJE3XN  MJE44N  MJE46E  MJE4XY  MJE4Y1  MJE518  MJE51D  MJE548  MJE585  MJE597  MJE5A1  MJE5E4  MJE5HT  MJEMAE  MJEMKJ  MJEMKY  MJEML0  MJEML1  MJEML2  MJEP28  MJEP2P  MJEP31  MJEP3E  MJEP95  MJEP9D  MJEP9E  MJEPJY  MJEPKP  MJEPPM  MJEPX0  MJHD0M  MJHD21  MJHD71  MJHD73  MJHDH2  MJHDH3  MJHDH4  MJHDH5  MJHH92  MJHHYY  MJHJ0R  MJHJ35  MJHJ8L  MJHJHK  MJHJJA  MJNEM6  MJNETR  MJNJL7  MJNK9K  MJNL2V  MJNL6M  MJNLKM  MJNLY3  MJNM4X  MJNM6P  MJNNAW  MJNP0R  MJNPPM  MJNRD5  MJNRNX  MJNRP7  MJNV1M  MJNVHD  MJNVL8  MJNVM9  MJNVV4  MJNWE0  MJNWP7  MJNWPL  MJNX41  MJNXTL  MJP048  MJP0P1  MJP107  MJP17K  MJP27Y  MJP29W  MJP2YV  MJP4J2  MJP4NT  MJP54K  MJP5ML  MJP5T6  MJP650  MJP6P5  MJP6R2  MJP77R  MJP820  MJP8K6  MJP9PP  MJP9YH  MJPD1Y  MJPD37  MJPD6D  MJPDA7  MJPDDL  MJPDNE  MJPDWV  MJPEHY  MJPEN7  MJPHMV  MJPHVM  MJPJKK  MJPJT7  MJPJVT  MJPK3J  MJPKYH  MJPL9W  MJPMK7  MJPMLK  MJPN6M  MJPNA2  MJPNDT  MJPNW6  MJPP3K  P1AMKD  P1DMKD  P1DMKDX1  P92MAE  P9XMAE  PA1MAE  PA2MAE  PKEMEE  PM2MAE  PM3MAE  PM4MAE  PM5MAE  PNHMMA  PNKMMA  PNMMMA  PNNMMA  PPXMMA  R0XMEE  R11MEE  R12MEE  R4PMME  R4RMME  R4TMME  R4WMME  R61MPD  R6HMPD  R9RMPD  R9TMPD  R9VMPD  R9WMPD  RV1MPA  RV2MPA  RV2MPAX1  RV3MPA  T0KMDE  T0LMDE  T0MMDE  T0NMDE  T3DMEE  T50MDE  T51MDE  T52MDE  T53MDE  T9PMKA  T9PMKAX1  TNJMKA  TNMMKA  TNMMKAX1  TNNMKA  TPPMKA  TPPMKAX1  TPVMKA  TR9MAE  TRDMKA  TREMKA  TREMKAX1  TYRMJE  V0HMAE  V20MJE  V27MJE  V28MJE  V2HMAD  V2JMAD  V2KMAD  V2LMAD  V2MMAD  V2NMAD  V2PMAD  V2PMADX1  V36MJE  6J6MAD  6K2MKE  6MVMMD  6MVMPD  6N8MPD  6N9MPD  6NAMPD  6NDMPD  6NHMPD  74JMDE  74KMDE  74LMDE  74MMDE  76TMAD  76TMADX1  76WMAD  76XMAD  76YMAD  771MAD  771MADX1  772MAD  77XMKD  77YMKD  780MKD  781MKD  7A5MKD  7A8MKD  7ADMKD  7AHMKD  7D5MTD  7D6MNE  7DVMKD  7HHMNE  7JTMKE  7JWMKE  7KDMKE  7KDMKEX1  7L6MJD  7N6MJD  7N7MJD  7NMMJD  7XTMDE  818MDE  87YMPE  888MPE  898MPE  899MDE  899MPE  89AMDE  8ADMTA  8AEMTA  8AHMTA  8AJMTA  8AJMTAX1  8DJMPE  8DKMPE  8L7MJE  8L8MJE  8L9MJE  8LAMJE  8LVMME  8PHMMA  8PJMMA  8PKMMA  8PKMMAX1  8RVMME  8THMPE  8WEMME  8WKMME  8WLMME  HVHMRA  HVHMRAX1  HVHMRAX2  HVKMLA  HVLMLA  J02MHD  J0LMHD  J1DMLE  J1PMHD  J1PMHDX1  J1RMHD  J1VMHD  J1VMHDX1  J1YMHD  J1YMHDX1  J20MHD  J23MLE  J2NMHD  J2WMLE  J3PMHD  J4AMHD  J4HMLE  J72MLE  JE2MTA  JE6MHD  JH0MHD  JMDMRD  JMMMRD  JMNMRD  JPRMRD  JR5MRD  JRJMHD  JTWMJE  JW6MHD  JXPMHD  JYKMHD  K28MJE  K2JMME  K2KMME  K2LMME  K2MMME  K5KMJE  K63MJE  K6KMJE  K6MMJE  K7DMJE  K7VMJE  KLEMKE  KPHMKE  KPJMKE  KRHMME  KRMMME  KXWMKE  L00MED  L00MEDX1  L1RMED  L1VMED  L1WMED  L1XMED  L1YMED  L20MED  L22MED  L4HMTA  L4JMTA  L4KMTA  L4LMTA  L4MMTA  L52MPA  L53MPA  L54MPA  MHDLEV  MHDM6V  MHDM7L  MHDMH5  MHDMH6  MHDMJ4  MHDMJ5  MHDMJE  MHDMJEX1  MHDVW5  MHDVW6  MHDVW7  MHDVW8  MHDWW2  MHDWY3  MHDWY4  MHDWY5  MHDX25  MHDX3N  MHDX7N  MHDXKA  MHDY16  MHDY24  MHDY25  MHE87P  MHE87Y  MHE882  MHE884  MHE91J  MHE938  MHE93L  MHE93N  MHE953  MHE98R  MHE9E2  MHE9HX  MHEHYA  MHEHYAX1  MHEJ06  MHEJ06X1  MHEJ0L  MHEJ0LX1  MHEJ55  MHEJ55X1  MHEJ56  MHEJ56X1  MHEJ6Y  MHEJ70  MHEJ71  MHEJ72  MHEJ72X1  MHEJP4  MHETNM  MHETX0  MHETY8  MHEV0R  MHEV20  MHEVVJ  MHEVWK  MHEVWL  MHH2Y8  MHH32P  MHH32X  MHH330  MHH330X1  MHH34R  MHH34T  MHH34W  MHH34WX1  MHH34X  MHMPP4  MHMPV3  MHMPWH  MHMPWR  MHN343  MHN348  MHN349  MHN34D  MHN3EM  MHN3EN  MHN3ER  MHN3EV  MHN40T  MHN40W  MHN40Y  MHN410  MHN446  MHN44M  MHN44N  MHN44P  MHN44R  MHN4AX  MHN4AY  MHN4D0  MHN4D1  MHN4NL  MHN4P7  MHN4P8  MHN4P9  MHNKHX  MHNKJV  MHNKJX  MHNKJY  MHNKK1  MHNKRP  MHNKRT  MHNKT9  MHNKV7  MHNKVR  MHNKVT  MHNKVV  MHNKVW  MHNL3V  MHNL41  MHNL42  MHNL43  MHNL5L  MHNL5M  MHNL5N  MHNL5R  MHNLX3  MHNM04  MHNM05  MHNM26  MHNM27  MHNM2N  MHNM2R  MHNM42  MHNM43  MHNM44  MHNM45  MHNN20  MHNN28  MHNN2P  MHNN2R  MHNN65  MHNN66  MHNN7A  MHP8R7  MHP8R8  MJHJK3  MJHJK8  MJHK05  MJHK06  MJHR4K  MJHR59  MJHR5A  MJHR5D  MJJ21N  MJJ21T  MJJ220  MJJ221  MJJ234  MJJ235  MJJ243  MJJ24L  MJJ26L  MJJ273  MJJ289  MJJ29P  MJJ2HM  MJJ2M0  MJJ2M1  MJJ2M2  MJJ2M3  MJJ2V5  MJJ2V6  MJJ2V7  MJJ2V8  MJK40D  MJK40E  MJK40H  MJK40J  MJK72A  MJK762  MJK79T  MJK7HJ  MJK7M7  MJK7RV  MJK7TD  MJK8JT  MJK8L8  MJK8LX  MJK8N2  MJK8RT  MJK907  MJK91L  MJKEEM  MJKEEN  MJKEEP  MJKEER  MJKEKK  MJKEKL  MJKEKM  MJKEMJ  MJKENH  MJKETL  MJKETN  MJKETP  MJKEVJ  MJKH0J  MJKH2M  MJKH6Y  MJKHM8  MJKHM9  MJKHMD  MJKHMK  MJKHMX  MJKHN3  MJKHTY  MJPP6D  MJPPWA  MJPR5D  MJPRM4  MJPT88  MJPT90  MJPW9K  MJPWWT  MJPX30  MJPXLX  MJPYD7  MJPYVV  MJR0NP  MJR0YK  MJR1VE  MJR322  MJR4VY  MJR5VY  MJR627  MJR7L7  MJR7NV  MJR7Y5  MJR86T  MJR86W  MJR91D  MJR9RM  MJR9WL  MJRA1V  MJRAXE  MJRD6X  MJRDRV  MJRE8N  MJRHX8  MJRJ6A  MJRJJM  MJRJRR  MJRJWJ  MJRK9N  MJRKE6  MJRLLE  MJRLLK  MJRLR4  MJRNDH  MJRNX1  MJRP32  MJRP7Y  MJRPL6  MJRPPN  MJRR7J  MJRRDX  MJRT2D  MJRTJR  MJRW7V  MJRX4T  MJRXE6  MJRXKJ  MJRYWA  MKA0MM  MKA142  MKA1JT  MKA2V6  MKA34A  MKA357  MKA3JJ  MKA3MM  MKA4LM  MKA571  MKA5NX  MKA5PK  MKA69J  V40MJE  V44MAE  V47MAE  V5EMJE  V5HMJE  V5JMJE  V5KMJE  V9KMMD  V9KMMDX1  V9YMJE  VAKMJE  VARMLE  VATMLE  VAVMLE  VAWMLE  VE5MMD  VH3MJE  VL0MMD  VL8MMD  VL8MMDX1  VNKMNE  VNLMNE  VNMMNE  VNNMNE  VWDMNE  VWRMEE  VWXMNE  VY1MNE  VYDMMD  VYDMMDX1  VYHMMD  W6MMEE  WA6MAD  WJ8MEE  WKTMEE  WV1MEE  WV1MEEX1  WV2MEE  WV3MEE  WV4MEE  WV5MEE  XDJMDE  XDKMDE  XDLMDE  XDMMDE  XH9MHE  XHAMHE  XHDMHE  XKMMHE  YHNMHE  YHTMHE  YHVMHE  YHVMHEX1  YNJMJE  YNKMJE  YNLMJE  YNMMJE  000MNE  001MNE  002MNE  003MNE  03TMEE  040MEE  043MEE  075MPE  077MPE  079MPE  0DJMNE  0DKMNE  0DLMNE  0DMMNE  0EKMJE  0LEMJE  0RHMPE  0RJMPE  0RKMPE  10AMJD  11EMJE  11HMJE  11MMJD  11NMJD  11PMJD  11RMJD  11TMJD  12XMJD  12YMJD  13YMED  141MED  143MED  144MED  145MED  15PMKE  190MKE  1A1MKE  1A8MKE  1AHMKE  1AYMKE  1D5MKE  1DJMKE  1EXMKE  1HAMKE  1HLMJE  1KPMJE  21PMPA  23EMRD  23HMRD  23JMRD  23KMRD  23LMRD  23MMRD  248MJE  24DMJE  251MLE  8WRMRA  8WRMRAX1  8WTMRA  8WTMRAX1  90DMME  90KMEE  910MME  911MEE  913MME  91PMME  92MMEE  951MPA  952MPA  97RMMA  97RMMAX1  97TMMA  97VMMA  996MEE  9DEMNE  9KNMEE  9L3MNE  9MHMNE  9N5MEE  9R5MNE  9R7MNE  9RAMNE  9RRMNE  9VTMEE  9WEMEE  9WEMPA  9WHMPA  9WHMPAX1  9WJMPA  9WJMPAX1  9XTMMA  9XTMMAX1  A1TMTA  A1TMTAX1  A1VMTA  A1WMTA  A1WMTAX1  A4AMTA  A4AMTAX1  A84MMA  A9VMDD  A9VMDDX1  A9WMDD  A9WMDDX1  A9XMDD  A9YMDD  AA0MDD  ADVMMA  ADVMMAX1  ADVMMAX1X1  ADWMMA  ADXMMA  ADYMMA  AE0MMA  AE1MMA  AE2MMA  AKEMMA  AKEMMAX1  AKHMMA  ALYMMA  ANMMMA  AT8MMD  AT8MMDX1  ATTMJD  AV0MJD  AW5MMD  L5HMKE  L80MKE  L9RMED  L9RMEDX1  LD9MKE  LDEMRA  LDHMRA  LDJMRA  LDPMRA  LDRMRA  LDTMKE  LEDMKE  LEEMKE  LH4MKE  LHKMKE  LHLMHD  LJ5MKE  LJ9MKE  LK3MKE  LM0MPD  LM1MPD  LM2MKE  LM2MPD  LM3MKE  LM4MKE  LT6MDD  LT8MDD  LT8MDDX1  LTNMPA  LTPMPA  LTTMPA  LVMMDD  LVWMDD  M1TMNA  M1VMNA  M1VMNAX1  M2TMPA  M2YMPA  M34MNA  M34MNAX1  M35MNA  M35MNAX1  M49MPD  M4AMPD  M4AMPDX1  M4DMPD  M7NMKE  M7PMKE  M8AMKE  MEPE22  MEPE23  MEPE25  MEPE26  MEPENN  MEPEX8  MEPH52  MEPH53  MEPH6N  MEPKK8  MEPKK8X1  MEPKK9  MEPKKA  MEPKKD  MEPLNX  MEPLP2  MEPM1M  MEPM2W  MEPM54  MEPN9K  MEPNEN  MHH34Y  MHH40J  MHH47M  MHH47N  MHH4LM  MHH4LN  MHH4PN  MHH4PP  MHH54K  MHH551  MHH55K  MHH55L  MHH55M  MHH55MX1  MHH560  MHHDJY  MHHDTY  MHHDV4  MHHE4Y  MHHE54  MHHEA8  MHHEA9  MHHEE2  MHHEE4  MHHEMX  MHHEN9  MHHEPA  MHHET0  MHHET1  MHHET2  MHHET3  MHHET4  MHHET5  MHHET6  MHHET7  MHHET8  MHHHK0  MHHHK1  MHJ291  MHJ2KN  MHJ2R6  MHJ313  MHJ34H  MHJ36L  MHJ37P  MHJ38T  MHJ38X  MHJ397  MHJ39K  MHJ3HY  MHJ3J1  MHJ3J2  MHJ412  MHJ416  MHJ419  MHJ4MA  MHJ4MD  MHJ4ME  MHJ4MH  MHJ4MJ  MHK839  MHK86Y  MHK8HJ  MHK8LN  MHK8NM  MHK8NN  MHK8Y0  MHK91A  MHK937  MHK93H  MHP8R9  MHP8RA  MHP90W  MHP90X  MHP911  MHP912  MHP9RV  MHP9RY  MHP9T0  MHPA19  MHPA1H  MHPA1L  MHPA1M  MHPA1N  MHPLJW  MHPLJX  MHPLJY  MHPLK0  MHPLK2  MHPLK3  MHPLK4  MHPLK5  MHPX1P  MHPX1R  MHPX1T  MHPX1V  MHPX1W  MHPXMW  MHPXNP  MHR01N  MHR01R  MHR01T  MHR01V  MHR01W  MHR06P  MHR06Y  MHR073  MHR0NW  MHR194  MHR195  MHR196  MHR198  MHR1J6  MHR1J9  MHR1JA  MHR1JE  MHR210  MHR211  MHR266  MHR267  MHR268  MHR26D  MHRHL4  MHRHL7  MHRHLD  MHRHLH  MHRJE2  MHRJE3  MHRJE4  MHRJE5  MHRJE6  MHRJE7  MHRJE8  MHRJE9  MHRMX9  MHRMXD  MHRMXE  MHRMXJ  MHT668  MHT6EV  MJKHV0  MJKHV1  MJKHV2  MJKJ2W  MJKJ7P  MJKKX1  MJKKX3  MJKKX4  MJKKX5  MJKL0D  MJKL0E  MJKL28  MJKL43  MJKLA7  MJKLDN  MJKLHK  MJKLJ5  MJL0Y9  MJL11M  MJL1EA  MJL2L0  MJL33A  MJL3RE  MJL45W  MJL4LR  MJL8T6  MJL94K  MJL983  MJL9D2  MJL9H7  MJLA1A  MJLAVT  MJLE4N  MJLE7N  MJLE8W  MJLE8X  MJLE8Y  MJLE90  MJLED2  MJLED3  MJLED4  MJLEKK  MJLEKN  MJLEKV  MJLEKX  MJLELW  MJLEM4  MJLEN2  MJLH24  MJLH60  MJLHT5  MJLJ1K  MJLJDL  MJLLE0  MJLLWD  MJLMED  MJLMER  MJLMM0  MJLMRD  MJLN2H  MJLNM9  MJLNYD  MJLPHL  MJLRA7  MJLRJA  MJLTKR  MJLVV2  MJLW8Y  MJLWV6  MJLXKH  MKA6KL  MKA7VT  MKA81K  MKA91N  MKA969  MKA9ET  MKA9TL  MKAA9P  MKAARX  MKAAXT  MKADDP  MKADMM  MKADMT  MKAE3M  MKAE5V  MKAE7X  MKAEM3  MKAHX1  MKAJ7D  MKAJHY  MKAJY3  MKAJYE  MKAJYJ  MKAK7A  MKAKV8  MKAKVD  MKAKYD  MKAL3L  MKAL8J  MKALMT  MKALRR  MKALVE  MKAN94  MKANPX  MKAP07  MKAP51  MKAP6W  MKAP75  MKAPMN  MKAPN7  MKAPRE  MKARVT  MKATM4  MKAV09  MKAV62  MKAVJP  MKAVV1  MKAWJT  MKAXD4  MKAXRK  MKAYEA  MKAYPY  MKAYY0  MKD18E  MKD1J0  MKDAPT  MKDD2E  MKDD9D  MKDDMH  MKDHEA  MKDHNX  MKDMPM  MKDMY4  MKDT6E  MKDXJW  MKE075  MKE08M  MKE0AE  MKE17E  MKE34V  253MLE  254MLE  255MDE  255MLE  25DMDE  25HMDE  25JMDE  29DMHE  29WMHE  2AXMHE  2DMMHE  2E7MHE  2JHMKD  2JHMKDX1  2JJMKD  2JKMKD  2JLMKD  2JLMKDX1  2JMMKD  2JNMKD  2JRMHE  2LTMLE  2LVMLE  2LVMLEX1  2NXMLE  2VEMLE  2Y3MLE  2Y6MLE  2Y7MLE  35LMRD  35MMRA  35NMRA  3RKMAE  3RLMAE  3RMMAE  3RNMAE  3WKMAE  41AMME  455MED  455MEDX1  456MED  457MED  458MED  459MED  45AMED  45JMTD  45XMTD  461MTD  47KMHE  4E8MHE  4EDMHE  4EEMHE  4K0MTD  4K7MTD  4KAMTD  4R5MME  4R6MME  4R7MME  4XYMPA  4Y0MPA  4Y0MPAX1  4Y2MPA  4YXMTD  53LMEE  53MMEE  53NMEE  53PMEE  53YMTD  540MTD  585MTA  AW8MND  AWHMND  AWHMNDX1  AXAMJD  AXHMND  AXHMNDX1  AXNMKD  AXNMKDX1  AXTMMA  AYNMKD  D0YMMD  D11MMD  D19MMD  D1EMND  D1LMJD  D1LMJDX1  D2AMKD  D2AMKDX1  D30MKD  D4TMND  D74MEE  D75MEE  D76MEE  D77MEE  D7PMJD  D7PMJDX1  D8WMND  D9NMND  D9YMLE  DA0MLE  DA1MLE  DA2MLE  DD2MEE  DDMMMA  DDNMMA  DDPMMA  DE1MND  DEEMEE  DHJMKD  DLWMND  DM3MEE  DNYMKD  DPMMEE  DT8MKD  E13MLE  E16MLE  E22MLE  E2DMLE  E36MPA  E38MNE  E3KMND  E3LMND  E3MMND  E3YMNE  E49MPA  E49MPAX1  E4XMNE  E5AMPD  E5VMND  E5WMND  E5XMND  E67MPD  E68MPD  E6JMPD  E6WMNE  E6YMPD  EADMHE  EAMMDD  EEDMDD  EEEMME  MEPNER  MEPNET  MEPNETX1  MEPNETX2  MEPNXN  MEPTA6  MEPTD7  MEPTHN  MEPTMR  MEPTN5  MEPTN6  MEPTYW  MER3DA  MER3DN  MER3M0  MER3M1  MER3WE  MER3WEX1  MER3WH  MER41V  MER48R  MER4DY  MER4E0  MER4E2  MER4K1  MER4K3  MER85M  MER85MX1  MER88M  MER88T  MER88TX1  MER893  MERH5P  MERH5PX1  MERH6L  MERH6M  MERH6N  MERHPP  MERHPR  MERHRD  MERJVV  MERJVVX1  MERK3A  MERK6W  MERK6Y  MERL38  MERL39  MET7MP  MET7MPX1  MET7MR  MET7MRX1  MET7MT  MET7MV  MET7MVX1  MET80N  MET80P  MET80R  MET80RX1  MET80T  METD4V  METD4W  METDD0  METDD1  METDD1X1  METEHM  METEHR  METEHT  METEHTX1  METEHV  METHA1  MHK93K  MHK946  MHK947  MHK996  MHK9DE  MHK9JT  MHK9LH  MHK9LJ  MHK9LK  MHK9YN  MHKA07  MHKA08  MHKA10  MHKA1D  MHKA28  MHKA2V  MHKA31  MHKA44  MHKA46  MHKA47  MHKA97  MHKA98  MHKAAN  MHKAAP  MHKAPD  MHKAPE  MHKAPH  MHKAPJ  MHKD35  MHKD76  MHKD8N  MHKD8P  MHKD8X  MHKD90  MHKD90X1  MHKD96  MHKDA1  MHKDJM  MHKDM1  MHKDM2  MHKDM3  MHKDWA  MHKDWD  MHKE38  MHKE69  MHKE7L  MHKE85  MHKE86  MHKE89  MHKE8D  MHKE90  MHKE92  MHKEDL  MHKEVX  MHKH2A  MHKH71  MHKH7J  MHKH7L  MHKH7N  MHKH84  MHKH85  MHKH8R  MHKH8T  MHKH9D  MHKHRR  MHKHT4  MHKHTE  MHKHXD  MHL76K  MHL76L  MHT6KE  MHT6PE  MHT6TM  MHT6WE  MHT74X  MHT74Y  MHT76Y  MHT774  MHT775  MHT78M  MHT78N  MHTAPH  MHTAR1  MHTAYR  MHTD14  MHTD3Y  MHTD7Y  MHTD9L  MHTD9X  MHTDD5  MHTDDD  MHTDL6  MHTDL9  MHTDME  MHTDMH  MHTDMW  MJA2TP  MJA2TR  MJA2TV  MJA2TW  MJA2TY  MJA2V0  MJA2V1  MJA2V2  MJA2V3  MJA3T6  MJA3T7  MJA3T8  MJA3T9  MJA45N  MJA45P  MJA45R  MJA45T  MJA4J7  MJA4J9  MJA4JA  MJA4JD  MJA502  MJA503  MJA504  MJA505  MJD00L  MJD010  MJD05P  MJD08D  MJD0HY  MJD0KA  MJD0P7  MJD10J  MJD10K  MJD133  MJD135  MJD162  MJD17L  MJD17M  MJD17N  MJD17Y  MJD18D  MJD192  MJDAV6  MJLY34  MJLYVA  MJM0YW  MJM1KD  MJM1KM  MJM1MV  MJM1NH  MJM1W5  MJM1WY  MJM292  MJM2WW  MJM380  MJM504  MJM5R7  MJM5YJ  MJM624  MJM637  MJM63H  MJM7A7  MJM7J6  MJM8W9  MJM93R  MJM9L8  MJMAP6  MJMENH  MJMEP5  MJMJ11  MJMK65  MJMN3K  MJMNTD  MJMP2X  MJMP48  MJMP7D  MJMPN7  MJMPPJ  MJMR4J  MJMRTP  MJMT95  MJMV43  MJMVRT  MJMW8A  MJMWH6  MJMXW5  MJN174  MJN224  MJN28Y  MJN60P  MJN68D  MJN6D0  MJN6T8  MJN75Y  MJN7MH  MJN8MV  MJN93A  MJN978  MJN9D8  MJN9HD  MJN9P6  MJN9TP  MJN9V3  MJN9W6  MJNA43  MJNA7V  MJNATL  MJNAV9  MJND42  MJND9P  MJNDKK  MJNEE0  MJNEJH  MKE40W  MKE41Y  MKE469  MKE6YT  MKE7LV  MKE8LA  MKE970  MKEA7M  MKEJ33  MKEKNA  MKEMPX  MKEND4  MLDMEE  MN9MEE  MP0MKD  MPEMEE  MPHMEE  MPJMEE  MPKMEE  MPLMEE  MPMMEE  MVHMLE  MW0MLE  MW1MLE  MW2MLE  MW3MLE  MX1MKD  MX7MED  MX8MKD  MXRMED  MY2MED  MYRMKD  MYTMKD  N0NMKD  N0XMKD  N10MKD  N1MMHE  N1PMHE  N1RMHE  N1VMHE  N4LMKD  N4LMKDX1  N4RMPA  N4RMPAX1  N4YMPA  N52MPA  N55MPA  N55MPAX1  N9AMMA  NEHMMA  NKKMMA  NKLMMA  NKNMMA  NKVMMA  NR0MKD  NW8MMA  NW8MMAX1  NXHMKD  NXJMKD  NYPMKD  NYRMKD  P02MKD  P03MKD  P03MKDX1  P0HMLE  P0MMKD  P0NMKD  P0TMLE  P0VMLE  P0WMLE",
      "center_classification_date": "20180702",
      "distribution_pattern": "Distributed nationwide. Foreign distribution to Australia, Canada, Chile, China, Brazil, France, Germany, India, Italy, Japan, Malaysia, Netherlands, New Zealand, Poland, Portugal, Romania, Singapore, South Africa, Spain, Sweden, Switzerland, and United Kingdom.",
      "state": "NJ",
      "product_description": "LFIT Anatomic CoCr Femoral Heads,  Catalog No. 6260-9-040",
      "report_date": "20180711",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Howmedica Osteonics Corp.",
      "recall_number": "Z-2293-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80059",
      "termination_date": "20200508",
      "more_code_info": "",
      "recall_initiation_date": "20180522",
      "postal_code": "07430-2006",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}