  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Austin",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81420",
      "recalling_firm": "Luminex Corporation",
      "address_1": "12212 Technology Blvd",
      "address_2": "N/A",
      "postal_code": "78727-6101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution. US nationwide, Canada, Ireland, Netherlands, Singapore, Thailand, and United Kingdom",
      "recall_number": "Z-2292-2019",
      "product_description": "Luminex Aries M1 System  Model: Aries M6V1  Aries Software v1.0     For In Vitro Diagnostic Use in clinical laboratories.      Scanner Label/Labeling:  JADAK  JDK-2330 REV B 170512-017",
      "product_quantity": "153 (140 U.S. and 13 O.U.S.)",
      "reason_for_recall": "An incorrect control board and firmware installed in some of the Hand Held Barcode Scanners could result in sample/patient mismatch or duplicate results (from other patients specimens) for a patient order.",
      "recall_initiation_date": "20180629",
      "center_classification_date": "20190819",
      "termination_date": "20210428",
      "report_date": "20190828",
      "code_info": "ARIES System UDI:  00840487101537   ARIES M1 System UDI:  00840487100080    ARIES-M12V1-IVD, ARIES System   ARIES-M6V1-IVD, ARIES M1 System  CN-0321-01, ARIES System Handheld Barcode Scanner   All scanners manufactured after April 2017 and including Part Number JDK-2330 Rev B."
    }
  ]
}