  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Mahwah",
      "address_1": "325 Corporate Dr",
      "reason_for_recall": "The recalling firm has received a higher than expected number of complaints documenting femoral head/hip stem disassociation for certain sizes of femoral heads manufactured prior to March 4, 2011.",
      "address_2": "",
      "product_quantity": "",
      "code_info": "Lots   10176601  10176602  10446901  10852601  10937801  11121001  11121002  11969801  11969802  12204501  12204502  12204503  12668301  12668302  12668303  13365101  13365102  13365103  14101701  14101702  14390201  14390202  15186701  15186702  15186703  15292201  15341301  15629301  15759301  16462501  16575502  16637001  16666801  16684301  17158901  17158902  17230301  17230302  18133801  18133802  18133803  18216101  18216102  18216103  18783601  18783602  19313601  19313602  19313603  19313604  19506101  19506102  19549501  19549502  19614801  19614802  19614803  19691301  19691302  19715701  19738101  19738102  19738103  19780201  19780202  19843901  19843902  20167601  20479101  20479102  35007304  35246401  35826901  35826902  35826903  35826904  35826905  36140401  36140402  36140403  36140404  36140405  36467201  36467202  36467203  36467204  36993801  37014601  37014602  37014603  37014604  37014605  37034301  37034302  37252501  37252502  37252503  37367101  37367102  37367103  37367104  37708901  37708902  37708903  37708904  38044501  38044502  38044503  38044504  38044505  38044506  38287401  38287402  38287403  38287404  87530201  87530202  87530203  87530204  87530205  87530501  87530502  87530503  87530504  87535501  87544601  90469601  90469602  92059101  92059102  92059103  92059104  92677301  92677302  93490501  93490502  93639401  93639501  93982601  93982602  36LMKA  374MKA  37KMMD  37PMTD  37RMTD  37TMTD  38449701C  38449702C  38449703X  38449704C  38449705C  394MKA  39JMMD  3AHMNE  3AVMNE  3AWMNE  3AWMNEX1  3JYMNA  3K3MNA  3KDMNA  3KDMNAX1  3KKMNA  3KKMNAX1  3MDMHD  3MJMMD  3NWMNE  3P2MHA  3P4MHA  3P5MHA  3P6MHA  3P7MHA  3PLMKD  3PPMKD  3T4MKD  3WJMMD  400MNE  401MNE  426MMD  42EMMD  438MHD  43LMTD  43MMTD  43NMTD  451MHD  47LMHD  48JMHD  48PMHD  49AMPD  4D6MHE  4D7MHE  4D8MHE  4V5MPE  4V5MPEX1  4VDMPD  4VDMPDX1  4WLMPE  4XJMKD  4XLMKD  4YPMKD  539MPE  53MMPD  53NMPD  57NMPE  59KMMD  59LMMD  59VMJE  5A1MKD  5A5MJE  5A7MJE  5K6MPE  DTVMJD  DYPMKE  E16MDD  E63MPD  E6HMND  E6LMPD  E6LMPDX1  E6WMPD  E6WMPDX1  E6YMND  E7HMDD  E7KMDD  E7KMDDX1  E7LMDD  E92MTA  E93MTA  E94MTA  E95MTA  E95MTAX1  EHRMNE  EJ0MAE  EJ4MAE  EVDMAE  EVLMNE  EWJMMD  H2XMAE  H30MAE  H31MAE  HARMTD  HARMTDX1  HHPMMD  HHTMMD  HHXMMD  HJLMTD  HJXMDD  HJXMDDX1  HJYMDD  HK3MLD  HNAMDD  HNAMDDX1  HNTMLE  HV2MLE  HV2MTD  J4JMLE  J5RMND  J5RMNDX1  J5XMND  J68MKE  J68MKEX1  J69MKE  J6HMND  J71MLE  J83MLE  J8LMLE  J9LMPA  J9MMPA  J9NMPA  JD1MDE  JE4MLE  JEJMKE  JJ8MLE  JK4MDE  JM4MDE  JM6MTD  JM7MTD  JM7MTDX1  JM8MTD  JMNMTD  JMPMTD  JMRMTD  MET5JR  MET5JRX1  MET5JV  MET63E  MET63H  MET63J  MET6PM  MET6YJ  MET724  MET727  MET7JY  MET7K7  MET7KL  MET7KLX1  MET7LH  MET7LN  MET80W  MET87X  MET8MR  MET9W3  MET9W3X1  MET9W5  MET9XT  MET9XV  META3H  META3HX1  META3HX2  META6V  META6X  METAA9  METAAE  METAE5  METAMN  METAR9  METARK  METATA  METAV5  METAV6  METAV7  METAV7X1  METAVE  METAWA  METAY7  METMR3  METMR3X1  METMVN  METMXV  METMY1  METMY5  MHA3WV  MHA4DT  MHA4JN  MHAHVL  MHAJ4L  MHAJLA  MHAJMX  MHAJP6  MHAJPR  MHAJRA  MHAJWR  MHAK0X  MHAK41  MHAKHD  MHAKM7  MHAKM7X1  MHAKW3  MHAKW7  MHAL07  MHAL0N  MHAL42  MHJH9X  MHJH9XX1  MHJHAN  MHJHKJ  MHJHTM  MHJHVW  MHJHVX  MHJJ63  MHJJ7X  MHJJ85  MHJJ8V  MHJJ8Y  MHJJR3  MHJJVL  MHJLAJ  MHJLEK  MHJLR6  MHJM24  MHJM38  MHJM4K  MHJM65  MHJMPR  MHJR7P  MHJRDY  MHJRE0  MHJRH6  MHJRKR  MHJRW2  MHJT0M  MHJT1R  MHJT24  MHK838  MHK86X  MHK8HE  MHK8NP  MHK8VH  MHK938  MHK93J  MHK94Y  MHK995  MHK9JN  MHK9L9  MHK9LA  MHK9LD  MHKA09  MHKA1J  MHKA1T  MHKA32  MHKA4A  MHKA4E  MHKA83  MHKA87  MHKA9E  MHKA9K  MHKAAT  MHKAAV  MHKAP4  MHKAP7  MHKD36  MHKD75  MHKD8M  MHKD8V  MHKD8W  MHKDA2  MHKDJL  MHKDM4  MHKDM5  MHKDWE  MHKDWH  MHKE3D  MHPM5H  MHPM5M  MHPM5N  MHPM69  MHPMAE  MHPMJK  MHPMNL  MHPMW5  MHPN94  MHPRNL  MHPVAX  MHR9LV  MHR9P7  MHR9PA  MHR9PL  MHR9PT  MHR9R4  MHR9RX  MHR9WR  MHRA14  MHRA9J  MHRAN1  MHRARK  MHRAT3  MHRAT4  MHRAVA  MHRD06  MHRD2E  MHRD2N  MHRD61  MHRDAW  MHRDEJ  MHRDW6  MHRDXR  MHRDXT  MHRDY7  MHRDY8  MHRE00  MHRE2E  MHRE32  MHRE4P  MHRE5H  MHRE8J  MHREJK  MHRELA  MHRETV  MHRH3V  MHRH5M  MHRH5T  MHRT87  MHRTA2  MHRTAK  MHRTDN  MHRTDX  MHRTDY  MHRTE0  MHRTLA  MHRTRN  MHRV49  MHT272  MHT2HM  MHT2LW  MHT2LWX1  MHT2Y8  MHT31P  MHT33E  MHT3EH  MHT3L2  MHT3L5  MHT3M5  MJHT9X  MJJ2HD  MJJ2HH  MJJ2HJ  MJJ305  MJJ307  MJJ308  MJJ39A  MJJ441  MJJ46A  MJJKA1  MJJKAA  MJJKAP  MJJKL1  MJJKX6  MJJKX7  MJJKX9  MJJLLN  MJJLNM  MJJLT6  MJK03M  MJK082  MJK0J9  MJK0WP  MJK0WT  MJK0WV  MJK112  MJK113  MJK114  MJK171  MJK1D6  MJK1MH  MJK1V9  MJK1VA  MJK1VD  MJK26W  MJK26X  MJK26Y  MJK2DW  MJK2DX  MJK2DY  MJK2TD  MJK2TE  MJK2TH  MJK78P  MJK7N1  MJK7TE  MJK8KE  MJK8M7  MJK8MA  MJK91M  MJKK0E  MJKKAM  MJKKJV  MJKKKL  MJKKKN  MJKKKP  MJKKWX  MJKKWY  MJKKX0  MJKL53  MJKL84  MJKLH2  MJKLJ6  MJKLLD  MJKLLE  MJKLLH  MJKYAY  MJKYMP  MJKYV7  MJPXRK  MJPXT6  MJPXV2  MJPY4J  MJPY59  MJPYR3  MJR1W0  MJR1X9  MJR218  MJR27W  MJR339  MJR47L  MJR4PK  MJR51K  MJR53K  MJR5XH  MJR6N2  MJR6NE  MJR7T6  MJR7YN  MJRAKP  MJRED9  MJREEX  MJRH90  MJRHK1  MJRHV3  MJRHX0  MJRKAK  MJRKW9  MJRLXY  MJRM6L  MJRMHJ  MJRMPH  MJRN3D  MJRNWT  MJRPYA  MJRRNV  MJRT3D  MJRTDM  MJRTM2  MJRV0V  MJRV5X  MJRVEY  MJRW56  MJRW7R  MJRWMR  MJRXY8  MJT02V  MJT03L  MJT09K  MJT156  MJT18J  MJT1Y7  MJT22Y  MJT29P  MJT36N  MJT3TW  MJT4JN  MJT4TK  MJT68D  MJT6NX  MJT6PE  MJT6T0  MJT8H5  MJT9N2  MJT9YN  MJTAVT  MJTAX5  MJTD4A  MJTD6X  PMEMHD  PMTMPD  PN4MHD  PNEMHD  PNVMKA  PNWMKAX1  PTKMEA  PTLMEA  PTMMEA  PTNMEA  PV4MJE  PVNMME  PVPMME  PVRMME  PXYMPD  PYAMKE  PYXMPD  R3YMKE  R5KMKE  R6TMKA  R6VMKA  R6VMKAX1  RAHMKD  RJ5MKA  RJKMKA  RL4MKA  RL4MKAX1  RLPMKA  RLPMKAX1  RLRMKA  RM4MKA  RM5MKA  RM8MHE  RM9MHE  RM9MHEX1  RMAMHE  RMAMHEX1  RT3MAD  RT5MEA  RT6MEA  RT7MEA  RT8MEA  RV7MNA  RV8MAD  RV8MADX1  RV8MNA  RV9MAD  RV9MADX1  RV9MNA  RVAMAD  RVAMNA  RVDMAD  RWPMEA  RWTMEA  RYYMAD  T00MAD  T00MADX1  T05MAD  T05MADX1  T06MAD  T07MAD  T08MAD  T09MAD  T24MLD  T25MLD  T25MLDX1  T29MLD  T2DMLD  T59MLD  T5YMLD  20479103  20479104  20479105  20479106  20479107  20479108  20513301  20513302  20620701  20620702  20620703  20620704  20620705  20620706  20620707  20620708  20637401  20637402  20637403  20637404  20683501  20683502  20683503  20684301  20684302  20684303  20704801  20724301  20731701  20746401  20783101  20783102  20783103  20783104  20783105  20783106  20783107  20783108  20904301  20904302  20904303  20904304  20936201  20936202  20936203  20936204  21035201  21035202  21035203  21035204  21035205  21109701  21109702  21109703  21109704  21135001  21135002  21135003  21135004  21151601  21208301  21208302  21208303  21208304  21208305  21263901  21263902  21263903  21263904  21263905  95178501  95501401  95501402  96144101  96144102  96144103  96144104  96419901  96893801  96893802  97136501  97136502  98094301  98883401  99063601  99349901  99349902  00HMLE  00JMLE  00KMLE  00LMLE  00PMPE  00YMJA  02JMEE  02KMEE  02KMPE  02LMEE  038MPD  07HMPE  08WMED  08WMEDX1  095MJA  097MJA  09MMJA  0AWMRD  0KEMPE  0KRMPD  0L1MJA  0MPMNA  0MPMNAX1  0MRMNA  0N5MJA  0N6MJA  0NLMDD  0NMMDD  0NNMDD  0NPMDD  0PYMRD  0RTMED  0RTMEDX1  0RVMED  0RWMED  0RWMEDX1  0TAMPE  0V1MPD  0V2MRD  0V3MRD  0VMMDE  0VVMRD  0W2MRD  0WNMPA  0WNMPAX1  0WRMPA  0WRMPAX1  0WTMPA  0WTMPAX1  0XTMPE  0Y5MLE  0Y6MLE  0Y7MLE  5LDMKD  5PNMPE  5Y6MPE  5Y9MPE  5YAMPE  61XMHE  620MHE  621MHE  63DMPE  63KMPE  63KMPEX1  68YMPA  68YMPAX1  690MPA  690MPAX1  690MPAX1X1  690MPAX2  692MPA  692MPAX1  6KJMND  6KKMND  6KLMND  6MLMEE  6MMMEE  6MNMEE  6NEMED  6P0MED  6P5MJE  6P8MED  6P8MEDX1  6P9MED  6T3MPA  6T3MPAX1  6V5MDE  6VWMHE  6VXMHE  6VYMHE  6X0MDE  6X1MDE  700MPA  700MPAX1  70KMPD  72JMPA  72JMPAX1  72KMPA  734MJE  738MME  74NMDE  74PMDE  74PMDEX1  74RMDE  75RMDE  776MJE  795MNE  799MME  7E9MJE  7EAMJE  7EDMJE  7HAMNE  7JYMJE  7K4MJE  7K7MJE  7KXMEE  7KYMEE  7L0MEE  7MTMJD  7MVMJD  7MWMJD  7MXMJD  7MYMJD  JNMMLE  JNNMLE  JNPMLE  JNRMLE  JRDMHE  JREMHE  JRHMHE  JV1MND  JWKMND  JWYMAD  JWYMADX1  JYMMPA  JYNMPA  K0XMND  K0YMND  K0YMNDX1  K1AMMD  K36MND  K37MND  K3MMMD  K3NMMD  K42MEE  K5RMAD  K5RMADX1  K5TMAD  K5VMRD  K5VMRDX1  K6JMAE  K6JMAEX1  K6MMAE  K6RMAE  K7RMAD  K7TMAD  K7TMADX1  K7VMAD  K7XMAD  K7XMADX1  K7YMAD  KEHMRD  KEJMRD  KEKMRD  KELMRD  KEMMRD  KKLMKD  KKMMKD  KKPMKD  KNJMDD  KNKMDD  KNKMDDX1  KNKMKD  KNLMDD  KNLMKD  KNMMDD  KNMMDDX1  KNMMDDX1X1  KPDMKD  L10MEE  L11MEE  L12MEE  L1KMTD  L1LMTD  L1MMKA  L1MMTD  L3EMJE  L4XMJE  L54MEE  L55MJE  L6MMJE  L6NMJE  L6PMJE  MHAL7T  MHALEW  MHALR7  MHALY9  MHAVXL  MHAVY3  MHAW1A  MHAW6H  MHAW6HX1  MHAW9E  MHAW9EX1  MHAWMH  MHAWRW  MHAWT7  MHD012  MHD012X1  MHD013  MHD014  MHD11L  MHD125  MHD12A  MHD154  MHD1A9  MHD1LM  MHD991  MHD9HY  MHD9TN  MHD9VW  MHD9XL  MHD9Y3  MHD9YD  MHD9YDX1  MHD9YDX2  MHD9YH  MHDA2V  MHDA2VX1  MHDA6J  MHDAL5  MHDAL5X1  MHDATR  MHDD0R  MHDD1M  MHDD3J  MHDD47  MHDXXH  MHDXXHX1  MHDXXHX1X1  MHDY0D  MHDY1A  MHDY2A  MHDY3P  MHDY4D  MHDY4DX1  MHE0N3  MHE0TV  MHE10L  MHE10W  MHE11T  MHE126  MHE146  MHE18N  MHE79W  MHE7M4  MHE7MT  MHE7NR  MHE7R9  MHE7RE  MHE7RT  MHE7TN  MHE7V3  MHKE6P  MHKE7M  MHKE8A  MHKE8K  MHKE91  MHKEDM  MHKEY8  MHKH29  MHKH29X1  MHKH7M  MHKH86  MHKH94  MHKHER  MHKHTN  MHKHTR  MHL73P  MHL73R  MHL73T  MHL8EH  MHL8EJ  MHL8H0  MHL8LP  MHL8WX  MHL97E  MHL97K  MHL9KR  MHL9V8  MHL9VW  MHL9W1  MHL9X7  MHL9X8  MHLA79  MHLA7A  MHLA7E  MHLA7H  MHLAWR  MHLD05  MHLD18  MHLD1E  MHLD3P  MHLD3R  MHLD3T  MHLD3W  MHLD3X  MHLE43  MHLXH7  MHLXLP  MHLY68  MHLYAP  MHLYD2  MHLYD4  MHM002  MHM006  MHM016  MHM057  MHM06N  MHM07M  MHM07MX1  MHM0E8  MHM0JL  MHM0V0  MHM0VW  MHM0YE  MHM102  MHM103  MHM103X1  MHM116  MHM137  MHM138  MHM139  MHT3R0  MHT3R1  MHT3R3  MHT69A  MHT6KD  MHT6PD  MHT6VW  MHT750  MHT754  MHT78P  MHT78R  MHT9H5  MHT9H7  MHT9H9  MHT9HA  MHT9HD  MHT9HE  MHTA8X  MHTA8Y  MHTAA6  MHTANR  MHTAPD  MHTAWY  MHTD18  MHTD7P  MHTD9K  MHTD9Y  MHTDLE  MHTDLK  MHTDMJ  MHTDP3  MJA2PR  MJA2PV  MJA2TL  MJA2TM  MJA2YL  MJA2YP  MJA336  MJA337  MJA357  MJA359  MJA3DV  MJA3JR  MJA3K4  MJA3LD  MJA3LN  MJA3N0  MJA3N1  MJA3P7  MJA3RW  MJA41N  MJA43D  MJA456  MJA49X  MJA4ER  MJA4R2  MJA4WL  MJA4Y2  MJA4YP  MJA507  MJA51M  MJA55M  MJA560  MJA6D4  MJA6D5  MJA6D6  MJA6RA  MJA70E  MJA76P  MJAMYA  MJL005  MJL10J  MJL10V  MJL3XR  MJL48D  MJL4YT  MJL698  MJL7J0  MJL7KM  MJL7LD  MJL7XR  MJL7Y9  MJL956  MJL9AJ  MJL9L2  MJLD92  MJLDT0  MJLEE2  MJLEE3  MJLEE4  MJLEP5  MJLHDX  MJLHE9  MJLHM1  MJLHM2  MJLHW4  MJLJY1  MJLKDX  MJLKX3  MJLLV1  MJLMKR  MJLN81  MJLNA4  MJLNW5  MJLR8X  MJLTMK  MJLV7X  MJLW0J  MJLW3J  MJLX0E  MJLXPH  MJLY7D  MJLYEK  MJM1T7  MJM2M1  MJM2P6  MJM34P  MJM388  MJM3A3  MJM3EH  MJM3ME  MJM46Y  MJM47V  MJM54X  MJM56P  MJM5WL  MJM6Y2  MJM76J  MJM774  MJM8HR  MJM8PJ  MJM8WK  MJM967  MJMAHV  MJMD70  MJMDP0  MJMDP1  MJME06  MJME0N  MJME34  MJTD7L  MJTDNX  MJTEKD  MJTEKN  MJTEYL  MJTJ1X  MJTJ4M  MJTJ4Y  MJTJAH  MJTK4X  MJTKL7  MJTPMV  MJTT78  MJTTK5  MJTVKY  MJTW3D  MJTXDW  MJTXH3  MJTXKP  MJTXY9  MJTY5D  MJTY88  MJTY94  MJTYKT  MJTYMV  MJTYX9  MK9MHD  MKA03H  MKA057  MKA0L6  MKA1M7  MKA23L  MKA2RT  MKA5JN  MKA5KD  MKA65Y  MKA6YE  MKA6YT  MKA817  MKA8H9  MKA95A  MKA95N  MKAEYL  MKAH41  MKAJEW  MKAJWV  MKAK69  MKAL75  MKALL3  MKAM8K  MKAM9Y  MKAMHD  MKAN1V  MKAN3H  MKANLV  MKANYM  MKAR2Y  MKARNA  MKARNK  MKARVW  MKAV50  MKAWHV  MKAX5T  MKAY32  MKAYET  MKAYT3  MKD0LV  MKD0RK  MKD10X  MKD12L  T67MLD  T6AMLD  TD6MLE  TEPMLE  TJ2MLE  TL0MEA  TLMMEA  TLNMEA  TPYMNE  TRMMJE  TRPMJE  TTYMMA  TV0MMA  TVHMEE  TVJMEE  TVKMEE  TX0MED  V0EMME  V27MED  V5TMJE  V5VMJE  V5WMJE  V79MMD  V8RMED  V9LMMD  VAKMLE  VE3MEE  VE5MEE  VE6MEE  VHNMMD  VKWMNE  VL1MNE  VL3MEE  VLAMMD  VLHMED  VM7MED  VMLMNE  VTKMME  VTPMNE  VTWMNE  VTXMNE  VV0MME  VV0MMEX1  VV4MMA  VV4MMAX1  VWKMMA  VWKMMAX1  VWVMMA  VYKMMD  VYMMMD  W02MMD  W03MMD  W04MMD  W12MNE  W13MNE  W4EMJE  W7EMJE  W7LMLE  W83MJE  W89MLE  W8KMNE  W8PMNE  WDKMHE  WKJMJA  WKNMJA  WL1MHE  WL2MHE  WL4MHE  WL6MHE  WL7MHE  21451901  21451902  21451903  21451906  21473701  21473702  21473703  21473704  21473705  21473706  21491501  21491502  21491503  21491504  21491505  21696001  21743801  21743802  21743803  21743804  21762304  22562601  22610301  22610302  22621701  22621702  22829501  22829502  22829503  23550701  23550702  24123001  24123002  24142501  24142502  24295101  24295102  24844701  24844702  24876401  24941801  25259901  25259902  25259903  25467301  25467302  25467303  25467304  26115701  26115702  26115703  26205801  26205802  26381001  26381002  27258001  27294101  27294102  27294103  27377201  27377202  27377203  28125401  28125402  28125403  28125404  28125405  28789801  28789802  28789803  0YMMNE  103MTD  105MTD  106MTD  108MTD  109MTD  10AMTD  10WMPE  10XMPE  10YMPE  117MLE  118MLE  119MLE  12204503A  122MHD  12KMNE  12LMNE  12TMDE  12VMDE  12WMDE  12WMPE  13VMNE  14DMNE  14VMPE  15VMKD  17LMHD  19313601X1  19313602X1  19313602X1X1  19313602X2  19313603X1  19313604X1  19506102X1  19549502X1  19549502X2  19614801X1  19614802X1  19614802X2  19614803X1  19738101X1  19738102X1  19738102X1X1  1E2MHD  1EP01X  1EP02J  1EP02W  1EP02WX1  1EP033  1EP088  1EP08A  1EP08D  1EP08E  1EP08J  1HEMEE  1JK0DY  1JK0L7  1JK0NA  1LVMKD  1P1MHD  1PRMEE  1T7MLE  1X4MMD  1XHMMD  1XHMMDX1  20479101X1  20479101X2  20479102A  20479102X1  20479103X1  20479104X1  7N0MJD  7N1MJD  7N2MJD  7N3MPD  7P7MME  7R1MME  7Y9MME  80JMCA  80KMCA  80LMCA  80MMCA  81JMME  81RMND  81TMND  81TMNDX1  81VMND  822MME  83DMME  83EMME  841MME  848MME  85PMME  87530202A  8D7MLD  8D7MLDX1  8HRMJE  8KPMDE  8KRMJE  8L2MEE  8LYMJE  8M6MJE  8MAMLD  8NRMJE  8P0MJE  90MMJE  91XMLD  92059103A  92LMLD  93490502A  96893801A  9H1MDA  9H2MDA  9H3MDA  9H4MDA  9H5MDA  9H6MDA  9HRMRD  9JDMRD  9JNMRD  9JTMNE  9KRMLD  9KXMRD  9M4MLD  9M9MND  9M9MNE  9MDMND  9MEMNE  9MHMND  9MMMDE  9MNMDE  9MPMDE  9MPMDEX1  9R9MNE  9T1MRD  9XWMRD  A0AMED  A0JMPD  A0NMPD  A0VMPD  A10MRD  L8NMEE  L92MEE  LD6MKE  LD8MKE  LD8MKEX1  LDAMKE  LDLMEE  LHRMPD  LKYMPD  LL1MEE  LMMMHA  LMNMHA  LMPMHA  LP9MHA  LPAMHA  LPDMHA  LPEMHA  LPHMHA  LPYMHA  LR0MHA  LR1MHA  LRKMAD  LV0MPD  LV0MPDX1  LXMMKA  LXNMKA  LXRMKA  M22MHE  M22MHEX1  M23MHE  M24MHE  M2LMPD  M65MPD  M79MAE  M9HMKA  M9HMPD  M9JMKA  M9JMPD  M9JMPDX1  M9KMPD  M9LMPD  M9NMKA  M9PMKA  M9RMKA  MA2MAE  MA2MJA  MA3MAE  MA4MAE  MEEMHD  MEJMHD  MEKMHD  MELMHD  MEPJHY  MEPJT3  MEPL22  MEPL58  MEPLA7  MEPLA7X1  MEPLAH  MEPLKD  MEPLKDX1  MEPLKDX2  MEPLLT  MEPLLY  MEPLN1  MEPMV6  MEPMW3  MEPMW4  MEPMW4X1  MEPMW5  MHE91E  MHE92D  MHE93D  MHE956  MHE956X1  MHEEA6  MHEEA6X1  MHEENM  MHEENMX1  MHEEWP  MHEH46  MHEHE2  MHEHE2X1  MHEHWE  MHEHWEX1  MHEHWV  MHEHWVX1  MHEHXY  MHEHXYX1  MHEJ00  MHEJ4N  MHEJ4NX1  MHEJ7H  MHEJLW  MHEJLWX1  MHEJWN  MHEJWNX1  MHEK0X  MHEK0XX1  MHEK27  MHEK2N  MHEK32  MHEK32X1  MHEK85  MHEK85X1  MHETH0  MHETH0X1  MHETHD  MHETLR  MHETLT  MHETNN  MHETNT  MHETNTX1  MHETW1  MHETWP  MHEVE1  MHEVKT  MHEVVH  MHEVVHX1  MHEVWN  MHEVWNX1  MHEVX7  MHEY67  MHEY67X1  MHH2XK  MHH2Y2  MHH2Y2X1  MHH2Y7  MHH2YH  MHH32T  MHH34N  MHH34P  MHH34PX1  MHH3KV  MHH3XE  MHH3XX  MHH3XXX1  MHH40V  MHH41D  MHH41J  MHM13A  MHM13AX1  MHM13D  MHMA91  MHMA92  MHMA92X1  MHMA93  MHMA94  MHMA95  MHMA95X1  MHMA95X2  MHMA96  MHMA97  MHMA98  MHMA99  MHMA9A  MHMA9D  MHMA9E  MHMA9H  MHMA9J  MHMA9K  MHMA9L  MHMA9M  MHMA9N  MHMA9NX1  MHMA9P  MHMA9R  MHME7T  MHMTNY  MHMTW7  MHMTW7X1  MHMV4N  MHMV4NX1  MHMV5A  MHMV5AX1  MHMVD0  MHMVD3  MHMVJ1  MHMVJ1X1  MHMVJN  MHMVJNX1  MHMVKK  MHMVLH  MHMVM9  MHMVM9X1  MHMVR4  MHMVY7  MHMVY7X1  MHMW15  MHMW4V  MHMW7T  MHMWD7  MHMWD8  MHMWM2  MHMWM3  MHMWME  MHMWMP  MHN00K  MHN02N  MHN08M  MHN0EX  MHN0TJ  MHN0VL  MHN0VM  MHN0VN  MHN0XN  MHN10M  MHN1L1  MHN1TD  MHN1VR  MJAMYD  MJAMYE  MJAN7W  MJAN7X  MJAN7Y  MJANAY  MJANDJ  MJANDK  MJANDL  MJANDM  MJANDN  MJANDP  MJAP61  MJAP65  MJAP68  MJAP6E  MJAP6K  MJAPT8  MJAPT9  MJAPTA  MJE0AP  MJE1HH  MJE1LW  MJE1P3  MJE295  MJE2H7  MJE2W2  MJE34D  MJE3M3  MJE3R6  MJE3VE  MJE3XP  MJE45D  MJE4XX  MJE4Y4  MJE527  MJE586  MJE7AK  MJE7AM  MJE7H9  MJE7HL  MJE7HM  MJE87X  MJE896  MJE89J  MJE8LN  MJE8LR  MJEHW7  MJEJ91  MJEJEY  MJEJHW  MJEKE0  MJEKKA  MJEKND  MJEKNE  MJEP1K  MJEP1M  MJEP1Y  MJEP9A  MJEPK0  MJEPKN  MJEPPN  MJERA1  MJERA2  MJERA4  MJERJH  MJERJJ  MJERJK  MJERRE  MJERXA  MJMEMR  MJMEXL  MJMH3X  MJMJ2M  MJMJA7  MJMJD6  MJMJHT  MJMJTX  MJMN0L  MJMP41  MJMPM1  MJMR7T  MJMRKA  MJMT48  MJMW49  MJMW4T  MJMW8W  MJMWA3  MJMX7P  MJMXR4  MJMY6E  MJMYE7  MJMYXP  MJN0R2  MJN1LJ  MJN1X3  MJN294  MJN2N8  MJN2T6  MJN386  MJN3M6  MJN3XY  MJN413  MJN44H  MJN479  MJN5A9  MJN5J3  MJN5NR  MJN5R2  MJN6RK  MJN89A  MJN8DA  MJN9AH  MJNA24  MJND2P  MJND54  MJNDED  MJNEJ3  MJNEVV  MJNHW1  MJNHWN  MJNJA1  MJNJDR  MJNJJE  MJNK3D  MJNKDR  MJNKHW  MJNKTR  MJNMWN  MJNN2Y  MJNN4T  MJNNJ9  MJNNT8  MJNNWJ  MJNPNE  MJNPNR  MJNR3E  MJNRP8  MJNRPY  MJNRTM  MKD1NJ  MKD1VH  MKD340  MKD39E  MKD3MM  MKD45M  MKD47H  MKD4XH  MKD5XE  MKD6R9  MKD84K  MKD8N1  MKD8XP  MKD94R  MKD9XX  MKDDH0  MKDDJD  MKDE6X  MKDHXJ  MKDJ6L  MKDJMD  MKDK51  MKDLK8  MKDM7W  MKDM96  MKDMW8  MKDN32  MKDNY3  MKDPJY  MKDR5K  MKDRJY  MKDT2V  MKDTX3  MKDV84  MKDWVP  MKDXPA  MKDYEJ  MKDYLW  MKE506  MKED9P  MKEJY1  MKEKYW  MKER0L  MKEVMK  MLEMEE  MLHMEE  MLJMEE  MM3MHA  MM4MHA  MM5MHA  MM6MHA  MM7MHA  MM8MHA  MMPMHA  MR8MKD  MTRMJA  MTVMJA  MTWMJA  MVPMDE  MW0MEA  MW1MEA  MW6MLE  MW7MLE  MW8MLE  MWPMNA  MY5MKD  MY7MKD  MYTMHE  MYVMHE  N00MHE  WXXMMA  WXXMMAX1  X0DMPD  XE4MAE  XEDMDE  XEEMDE  XEHMDE  XNVMAE  XNYMHA  XP2MHA  XP7MAE  YNRMJE  YNTMJE  YNVMJE   28857701  28857702  29265201  29265202  29265203  29265204  29966501  29966502  29966503  29966504  29969301  29969302  29969303  30287101  30287102  30309301  30597001  30597002  30597003  30700401  30700402  30877401  30877402  30877403  31309401  31309402  31309403  31829101  31829102  31829103  31829104  32217101  32242001  32280701  32658101  32658102  32658103  32658104  32773701  32773702  32798501  32798502  33187901  33187902  33234301  33235801  33602501  33602502  33602503  33602504  33994701  33994702  33994703  33994704  33994705  34354801  34354802  34354803  34354804  34354805  34793901  34793902  34793903  34793904  34793905  34838601  34838602  35007301  35007302  35007303  20479104X1X1  20479105X1  20479105X1X1  20479105X2  20479106X1  20479107X1  20479108X1  20479108X2  20513301X1  20513302X1  20513302X2  20620707X1  20620708X1  20783105X1  20936201D  20JMEE  20TMKD  21109701X1  21208301X1  21208305X1  21263901X1  21263904X1  21263905X1  21473702X1  221MEE  221MEEX1  226MMD  238MEE  23DMEE  23DMHD  23EMHD  23HMHD  23TMLE  24844701D  24DMLE  24EMLE  24MMLE  25KMHD  25KMHDX1  26115701D  26205801D  26205802A  269MKD  29AMTD  29DMTD  29EMTD  2HEMEE  2J3MKD  2J8MKA  2LVMMD  2MYMHE  2N0MHE  2N1MHE  2WDMKD  2WLMMD  30LMKA  315MKD  325MKD  34AMEA  34DMEA  34DMKD  34EMEA  34HMEA  34JMEA  34KMEA  34LMEA  34VMKA  354MHA  35LMKA  36KMKA  A17MEE  A18MEE  A19MEE  A1AMEE  A1DMEE  A1EMEE  A2JMRD  A2PMRD  A2PMRDX1  A31MTA  A33MTA  A34MTA  A34MTAX1  A68MME  A6TMME  A6VMME  A6WMME  A6XMME  A6YMME  A85MLD  AH4MLD  AH4MLDX1  AL3MED  ALTMLD  ANTMED  AT9MME  AWHMHE  AWMMHE  AWPMHE  AX5MED  AXDMED  AXMMED  D0XMAE  D0YMAE  D10MAE  D1JMJD  D22MHD  D2NMHD  D38MHD  D59MME  D5NMME  D5YMME  D6DMKE  D9DMTA  D9EMTA  D9EMTAX1  D9KMTA  D9MMTA  DA4MJD  DA8MLA  DAHMLA  DAHMLAX1  DAVMJD  DAWMJD  DAWMJDX1  DAXMJD  DELMKE  DHDMHE  DHEMHE  DHHMHE  DMNMKE  DNWMKA  DNYMKA  DP0MKA  DT2MED  DT3MED  DT4MED  DT6MED  DT7MED  DTRMJD  MEPNE8  MEPNE9  MEPNE9X1  MEPNEA  MEPNEAX1  MEPVT4  MEPWP7  MEPWRY  MEPWT0  MEPWTL  MEPWVD  MEPWVE  MEPWVEX1  MEPX2N  MEPX2NX1  MEPY7P  MEPY92  MEPYEV  MEPYM7  MEPYVD  MEPYVDX1  MEPYXM  MER0NW  MER0NX  MER0NXX1  MER12A  MER161  MER3DP  MER3M3  MER3M3X1  MER3WK  MER41X  MER474  MER4D4  MER4D8  MER4D8X1  MER4K2  MER4K2X1  MER4KD  MER4KL  MERALP  MERALPX1  MERALR  MERDA4  MERL3J  MERL3K  MERL3KX1  MERL3L  MERL3LX1  MERL8T  MET0HA  MET0HN  MET0HP  MET0T6  MET0T7  MET11H  MET11P  MET11R  MET11RX1  MET11T  MET11V  MET11W  MET22E  MET22L  MET22M  MET47N  MET47NX1  MET47P  MET47R  MET5JP  MHH41JX1  MHH47P  MHH4LR  MHH4PR  MHH53D  MHH53DX1  MHH547  MHH548  MHH55R  MHH55RX1  MHH55T  MHHDJV  MHHDLE  MHHDV0  MHHE4X  MHHE52  MHHEA5  MHHEA7  MHHEMY  MHHEN8  MHHEPE  MHHERJ  MHHERK  MHHERL  MHHERM  MHHERR  MHHERW  MHHERX  MHHHK3  MHHHKE  MHHHT0  MHHJ39  MHHJ4X  MHHPR0  MHHR48  MHHR49  MHHR4A  MHHRR6  MHHRTY  MHHRWW  MHHT42  MHHT43  MHJ34E  MHJ37R  MHJ38W  MHJ39M  MHJ3JL  MHJ3JM  MHJ3JN  MHJ3JP  MHJ4ML  MHJ4MM  MHJ4YL  MHJ4YM  MHJ56K  MHJDWA  MHJE03  MHJE04  MHJE14  MHJE1W  MHJE6V  MHJE6VX1  MHJE6W  MHJEE6  MHJEPX  MHJH11  MHJH44  MHJH77  MHJH7E  MHJH8V  MHN1VT  MHN1XE  MHN1XJ  MHN1XX  MHNKHW  MHNKJP  MHNKJR  MHNKJT  MHNKRY  MHNKV6  MHNKVX  MHNKVY  MHNKVYX1  MHNKW0  MHNKX8  MHNL59  MHNL5D  MHNL5H  MHNLWT  MHNLWY  MHNM03  MHNM24  MHNM25  MHNM3Y  MHNM40  MHNM41  MHNN0P  MHNN18  MHNN1J  MHNN67  MHNN79  MHNN7J  MHNN8K  MHNN8L  MHNN8M  MHNNJV  MHNNJW  MHNNJX  MHP1V7  MHP1V8  MHP1V9  MHP244  MHP245  MHP246  MHP247  MHP278  MHP279  MHP27D  MHP298  MHP29E  MHP29H  MHP29J  MHP29K  MHP29L  MHP29M  MHP29N  MHP29P  MHP29R  MHP29W  MHP29X  MHP29Y  MHP2A0  MHP2A1  MHP2A2  MHP2A3  MHP2A4  MHP2A5  MHP2A6  MHPM2V  MHPM3Y  MJET31  MJET4E  MJH50R  MJH50T  MJH50V  MJH6XD  MJH6Y0  MJH6Y1  MJH7JV  MJHAAJ  MJHAAK  MJHAAL  MJHAAX  MJHAH3  MJHAMD  MJHAT4  MJHAT8  MJHAT9  MJHATA  MJHAYD  MJHAYJ  MJHD74  MJHD76  MJHD77  MJHDHH  MJHDHJ  MJHDHK  MJHDLE  MJHDLX  MJHDLY  MJHDM0  MJHDPK  MJHDT4  MJHDWX  MJHE0J  MJHE0P  MJHE98  MJHEE6  MJHEEA  MJHEER  MJHJK7  MJHK07  MJHK08  MJHK09  MJHK29  MJHK2A  MJHK4E  MJHK72  MJHK8L  MJHKH5  MJHN07  MJHN56  MJHN57  MJHN75  MJHN76  MJHNDR  MJHNER  MJHNKE  MJHNLH  MJHNP1  MJHNP3  MJHNP4  MJHNW0  MJHNW1  MJHNW3  MJHP2D  MJHP52  MJHP7M  MJHT9V  MJHT9W  MJNRVW  MJNTA1  MJNV1D  MJNV4E  MJNVKE  MJNVPE  MJNVYP  MJNW57  MJNWK2  MJNYER  MJP0EL  MJP0TV  MJP1N7  MJP20J  MJP24X  MJP2P4  MJP2PM  MJP2TM  MJP34K  MJP3K9  MJP3RD  MJP43A  MJP4A7  MJP4YP  MJP5DJ  MJP5VV  MJP6LT  MJP73W  MJP74M  MJP7AE  MJP7ND  MJP7VA  MJP832  MJP90E  MJP90W  MJP9EL  MJPA65  MJPD50  MJPDJJ  MJPED4  MJPEVT  MJPEVW  MJPHWJ  MJPJME  MJPK4X  MJPKA9  MJPKHL  MJPKJM  MJPL8E  MJPLKL  MJPM5E  MJPM8R  MJPM9D  MJPMT2  MJPN4N  MJPNV5  MJPNV9  MJPP0V  MJPP2T  MJPPPK  MJPPR2  MJPR53  MJPRDX  MJPRRH  MJPT4P  MJPTJJ  MJPV89  MJPW5J  MJPWDL  MJPXNP  N1XMNA  N1XMNAX1  N1XMNAX2  N24MRD  N4KMNA  N4KMNAX1  N4LMNA  N4MMNA  N4MMNAX1  N4XMME  N5KMDD  N5KMDDX1  N5LMDD  N6MMDE  N7NMJA  N7PMJA  NA1MRD  NA4MME  NA8MME  NE2MED  NE2MEDX1  NE7MRD  NKRMAE  NKTMAE  NKVMAE  NLJMAE  NN1MDE  NN2MDE  NN3MDE  NNAMRD  NNEMRD  NNLMJE  NPHMKE  NPJMKE  NPMMKE  NPTMED  NPXMED  NR3MHE  NW5MDE  NXKMKD  NXXMKE  NYXMKD  NYYMKD  P0NMLE  P0PMKD  P0PMLE  P0RMLE  P0TMKD  P32MKE  P3HMJE  P3JMJE  P3KMJE  P3TMHD  P3VMEA  P3WMEA  P8MMED  P8TMKE  P8TMKEX1  P93MDD  PE7MPD  PEHMMD  PEJMHD  PJJMHD  PJKMHD  PJMMHD  PJNMHD  PJRMHD  PKMMEA  PMAMHD  PMDMHD",
      "center_classification_date": "20180702",
      "distribution_pattern": "Distributed nationwide. Foreign distribution to Australia, Canada, Chile, China, Brazil, France, Germany, India, Italy, Japan, Malaysia, Netherlands, New Zealand, Poland, Portugal, Romania, Singapore, South Africa, Spain, Sweden, Switzerland, and United Kingdom.",
      "state": "NJ",
      "product_description": "LFIT Anatomic CoCr Femoral Heads,  6260-9-036",
      "report_date": "20180711",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Howmedica Osteonics Corp.",
      "recall_number": "Z-2292-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80059",
      "termination_date": "20200508",
      "more_code_info": "",
      "recall_initiation_date": "20180522",
      "postal_code": "07430-2006",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}