  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Pacheco",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85359",
      "recalling_firm": "Biocare Medical, LLC",
      "address_1": "60 Berry Dr",
      "address_2": "N/A",
      "postal_code": "94553-5601",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: TX, MA, TN, MO, IN",
      "recall_number": "Z-2291-2020",
      "product_description": "4plus Streptavidin HRP Label,  Conjugated Streptavidin Horseradish Peroxidase, Detection Component, Catalog Number: HP604H",
      "product_quantity": "8",
      "reason_for_recall": "Firm has identified product has no staining. If used, may result in invalid test results when used with controls in the clinical setting. Positive control would observe no staining. The clinical impact may result in the consumption of patient tissue sample and a delay in diagnosis.",
      "recall_initiation_date": "20191010",
      "center_classification_date": "20200605",
      "termination_date": "20201202",
      "report_date": "20200617",
      "code_info": "Lot Number: 041619-4, Expiration Date: 2020/10"
    }
  ]
}