  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92399",
      "recalling_firm": "MEDLINE INDUSTRIES, LP - Northfield",
      "address_1": "3 Lakes Dr",
      "address_2": "N/A",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, China, Costa Rica, India, Lebanon, Panama, Qatar, Saudi Arabia, Singapore, United Arab Emirates.",
      "recall_number": "Z-2290-2023",
      "product_description": "Medline Probe Cover Kits containing Turkuaz Ultrasound Gel, Labeled as the following:     a. COVER, PROBE, W/GEL AND BANDS, 4X48 - Model Number DYNJE5900;     b. COVER, PROBE, W/GEL AND BANDS, 6X24 - Model Number DYNJE5910;     c. COVER, PROBE, W/GEL AND BANDS, 6X48 - Model Number DYNJE5920;     d. COVER, PROBE, W/GEL AND BANDS, 6X96 - Model Number DYNJE5930;     e. COVER, PROBE, W/GEL AND BANDS, 5X48 - Model Number DYNJE5940;     f. COVER, PROBE, W/GEL AND BANDS, 4X48 - Model Number DYNJEX5900",
      "product_quantity": "1,231,480 kits",
      "reason_for_recall": "Potential for the labeled sterile ultrasound gel component within packs to not meet sterility specifications.",
      "recall_initiation_date": "20230515",
      "center_classification_date": "20230803",
      "report_date": "20230809",
      "code_info": "a. Model Number DYNJE5900; UDI-DI Each: 10884389114124, Case: 40884389114125; Lot Numbers: 06920030012, 06920030021, 06920040022, 06920050011, 06920050011, 06920050011, 06920050011, 06920050011, 06920050011, 06920070011, 06920080012, 06920100011, 06920100011, 06920110011, 06920120011, 06920120011, 06921010011, 06921010011, 06921010022, 06921020011, 06921020021, 06921030021, 06921030031, 06921040011, 06921040011, 06921040011, 06921040011, 06921040011, 06921050011, 06921050011, 06921060011, 06921070011, 06921070011, 06921080011, 06921080011, 06921090011, 06921090012, 06921100011, 06921100012, 06921100012, 06921110011, 0692112H022, 0692112H022, 0692112H022, 0692201H011, 0692202H011, 0692202H011, 0692202H024, 0692202H024, 0692204H011, 0692204H011, 0692205H011, 0692205H011, 0692205H011, 0692205H011;   b. Model Number DYNJE5910; UDI-DI Each: 10884389114094, Case: 40884389114095; Lot Numbers: 06920120011, 06920120011, 06921010011, 06921030011, 06921040011, 06921040011, 06921040011, 06921040011, 06921040011, 06921040011, 06921060011, 06921070011, 06921070011, 06921080011, 06921080011, 06921090011, 06921110011, 06921120011, 0692112H022, 0692112H022, 0692202H011, 0692202H011, 0692202H024, 0692202H024, 0692204H011, 0692204H011, 0692205H011, 0692205H011, 0692208H014;   c. Model Number DYNJE5920; UDI-DI Each: 10884389114100, Case: 40884389114101; Lot Numbers: 06920030012, 06920030021, 06920040022, 06920040022, 06920050011, 06920050011, 06920050011, 06920070011, 06920080012, 06920090011, 06920100011, 06920100011, 06920110011, 06920120011, 06920120011, 06920120012, 06921010011, 06921020021, 06921030012, 06921030021, 06921030021, 06921040011, 06921040011, 06921040011, 06921040011, 06921040011, 06921050011, 06921060011, 06921070011, 06921070011, 06921080011, 06921080011, 06921090011, 06921100011, 06921100012, 06921100013, 06921110011, 0692112H022, 0692112H022, 0692112H022, 0692202H011, 0692202H024, 0692203H011, 0692204H011, 0692204H011, 0692205H011, 0692205H012, 0692205H013, 0692205H013, 0692206H011;   d. Model Number DYNJE5930; UDI-DI Each: 10884389114117, Case: 40884389114118; Lot Numbers: 06920030012, 06920030021, 06920040022, 06920040022, 06920050011, 06920050011, 06920050011, 06920050012, 06920070011, 06920080012, 06920100011, 06920110011, 06920110011, 06920110011, 06920120011, 06920120011, 06921010011, 06921010011, 06921010011, 06921020021, 06921030021, 06921040011, 06921040011, 06921040011, 06921050011, 06921060011, 06921070011, 06921070011, 06921080011, 06921080011, 06921090011, 06921090011, 06921090012, 06921100011, 06921100012, 06921100012, 06921120011, 0692112H022, 0692202H011, 0692202H011, 0692202H024, 0692202H024, 0692203H011, 0692204H011, 0692204H011, 0692205H011, 0692205H011, 0692205H011, 0692205H011, 0692206H011;   e. Model Number DYNJE5940; UDI-DI Each: 10888277723641, Case: 20888277723648; Lot numbers: 06920050011, 06920060012, 06920060012, 06920070021, 06920100011, 06920110011, 06920110011, 06920110011, 06920110012, 06920120011, 06921050011, 06921050011, 06921060011, 06921070011, 06921070011, 06921080011, 0692201H011, 0692202H011, 0692202H011, 0692202H024, 0692202H024, 0692203H011, 0692203H011, 0692204H011, 0692204H011;   f. Model Number DYNJEX5900; UDI-DI Each: 10884389114124, Case: 40884389114125; Lot Numbers: 06920040022, 06920080012.  **Lot format as follows: Example 069YYMM0001 where \"069\" is the last 3 digits of the vendor code for the probe cover kitting facility, \"YYMM\" is the year and month of manufacture of the probe cover kit, and \"0001\" is a sequential number for the month of production"
    }
  ]
}