  "results": [
    {
      "status": "Ongoing",
      "city": "Reading",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85645",
      "recalling_firm": "Arrow International Inc",
      "address_1": "2400 Bernville Rd",
      "address_2": "N/A",
      "postal_code": "19605-9607",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Domestic distribution to CA, FL, and MT. International distribution to AT, AU, DE, ES, FR, JP and SG.",
      "recall_number": "Z-2288-2020",
      "product_description": "Arrow FlexTip Plus Epidural Kit with NRFit Connector, catheter system, Product Codes (US)  OT19TAGSK  OT19TAGSKA  OT19TAPSK  OT19TCGLK    Product Codes (OUS)  AT-05501-NRF  OT19HBPSS  UM-05400-NRF  ASK-05500-NRON  EJ-05400-NRON  OU-05500-NRON  TU-05500-NRON    The Arrow Epidural Catheter kit permits access to the epidural space for administration of epidural anesthetic. The epidural catheter kit is intended for use up to 72 hours.",
      "product_quantity": "109 eaches (US), 23,021 (OUS)",
      "reason_for_recall": "Nine complaints reported that the rotatable collar on the filter had detached, causing a leak.",
      "recall_initiation_date": "20200508",
      "center_classification_date": "20200605",
      "report_date": "20200617",
      "code_info": "Lots:    (Dist'd in US)  23F19A0260  23F19M0224  23F19E0120  23F19L0014  23F19B0184    (Dist'd OUS)  71F19M0710  71F18K1579  71F18F1429  71F18J2111  71F20A1078  71F20A1614  71F19E2019  71F19L0269  71F19L2068  71F19K1049  71F19K2042  71F19L0977  71F19M1319  71F20A0805  71F20A0812  71F20A0813  71F19J0604  71F19K1653  71F19K2501  71F19M0326  71F20A1819  71F20A2040  71F20A0044  71F19K2043  71F19L0359"
    }
  ]
}