  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Athlone",
      "state": "",
      "country": "Ireland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80279",
      "recalling_firm": "Teleflex Medical Europe Ltd",
      "address_1": "Ida Business Park",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and the countries of  Canada, India",
      "recall_number": "Z-2287-2018",
      "product_description": "Teleflex MEDICAL Pilling(R) UNIVERSAL SHEARS 7 1/2\", REF 142542    Product Usage:  general surgical scissors",
      "product_quantity": "573 units",
      "reason_for_recall": "The incorrect plastic was used to manufacture the handles of these shears, which causes the handles to melt if sterilized.",
      "recall_initiation_date": "20180612",
      "center_classification_date": "20180626",
      "termination_date": "20210402",
      "report_date": "20180704",
      "code_info": "Lot Numbers:  F7, G7",
      "more_code_info": ""
    }
  ]
}