{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Burlington",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74563",
      "recalling_firm": "LeMaitre Vascular, Inc.",
      "address_1": "63 2nd Ave",
      "address_2": "N/A",
      "postal_code": "01803-4413",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-2286-2016",
      "product_description": "AnastoCiip Universal Clip Remover (6 Pack}   REF# 4001-00     The Anastoclip Clip applier consists of a rotating shaft and an integral cartridge containing titanium clips.  The Anastoclip Universal Clip Remover is intended to remove Anastoclip AC Clips as necessary. It is an accessory to the Anastoclip AC Clip Applier.",
      "product_quantity": "40 units (6-Pack)",
      "reason_for_recall": "US marketing brochure for the AnastoClip AC Closure  System states that the Anastoclip catalog numbers  can be used for dura closure, but the catalog numbers are only intended for cardiovascular use.",
      "recall_initiation_date": "20160622",
      "center_classification_date": "20160725",
      "termination_date": "20210408",
      "report_date": "20160803",
      "code_info": "Lot Numbers:  ANR1064¿  ANR1066¿  ANR1068¿  ANR1069¿"
    }
  ]
}