  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Palm Beach Gardens",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71212",
      "recalling_firm": "The Anspach Effort, Inc.",
      "address_1": "4500 Riverside Dr",
      "address_2": "N/A",
      "postal_code": "33410-4235",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US including AZ, CA, CO, FL, GA, IL, KY, LA, MN, MO, NC, NJ, NY, OH, OR, PA, TX, and Internationally to Canada and China.",
      "recall_number": "Z-2286-2015",
      "product_description": "ANSPACH  - Fluted Barrel;  8MM x 12.5MM   REF L-8BA-G1  Cutting shaping bone including spine and cranium.",
      "product_quantity": "110 tools.",
      "reason_for_recall": "G1 Dissection Tools are not compatible with all attachments listed on the labeling;  G1 cutting tools are labeled with a Specified Shaft Exposure which may not be consistent when used with each attachment on the product label. Shaft exposure may be greater.",
      "recall_initiation_date": "20150413",
      "center_classification_date": "20150729",
      "termination_date": "20160825",
      "report_date": "20150805",
      "code_info": "REF L-8BA-G1 ; ALL lot numbers"
    }
  ]
}