      "status": "Terminated",
      "city": "Anasco",
      "state": "PR",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68346",
      "recalling_firm": "AMO Puerto Rico Manufacturing, Inc.",
      "address_1": "Rd # 402, Km 4.2",
      "address_2": "",
      "postal_code": "00610",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - Nationwide (USA and Puerto Rico), and foreign:  Latin America, Asia Pacific, Canada, Europe, Africa and Middle East, and Japan.",
      "recall_number": "Z-2286-2014",
      "product_description": "AMO TECNIS CL Foldable Silicone IOL with OptiEdge Design, Model Z9002. Indicated for primary implantation for the visual correction of aphakia in adult patients. Tecnis CL Mono 13 mm with the following diopter sizes: 26.5, 27, 27.5 & 28.5.    Product Usage:  AMO TECNIS CL monofocal foldable IOLs with OptiEdge design are ultraviolet-absorbing posterior chamber intraocular lenses which compensate for corneal spherical aberrations. TECNIS CL Silicone Intraocular Lenses are indicated for primary implantation for the visual correction of aphakia in adult patients in whom a cataractous lens has been removed by extracapsular cataract extraction or phacoemulsification. These devices are intended to be placed in the capsular bag.",
      "product_quantity": "77,055 IOLs in total",
      "reason_for_recall": "Units may be labeled with the incorrect diopter power.",
      "recall_initiation_date": "20140513",
      "center_classification_date": "20140819",
      "termination_date": "20170518",
      "report_date": "20140827",
      "code_info": "Serial Num  1007351309  1007361309  1007371309  1007381309  1007391309  1007411309  1007431309  1007441309  1007451309  1007461309  1007471309  1007481309  1007491309  1007501309  1007511309  1007521309  1007531309  1007541309  1007551309  1007561309  1007571309  1007581309  1007591309  1007601309  1007611309  1007621309  1007631309  1007641309  1007651309  1007661309  1007671309  1007681309  1007691309  1007701309  1007711309  1007721309  1007731309  1007741309  1007751309  1007761309  1007771309  1007791309  1007801309  1007811309  1007821309  1007831309  1007841309  1007851309  1007861309  1007871309  1007881309  1007891309  1007901309  1007911309  1007921309  1007931309  1007941309  1007951309  1007961309  1007971309  1007981309  1007991309  1008001309  1008011309  1008021309  1008031309  1008041309  1008051309  1008061309  1008101309  1008121309  1008141309  1008161309  1008171309  1008181309  1008191309  1008201309  1008211309  1008221309  1008231309  1008241309  1008251309  1008261309  1008271309  1008281309  1008291309  1008301309  1008311309  1008321309  1008331309  1008341309  1008361309  1008381309  1008401309  1104951312  1119291309  1119311309  1119321309  1119331309  1119341309  1119351309  1119361309  1119371309  1119381309  1119391309  1119401309  1119411309  1119421309  1119431309  1119451309  1119461309  1119471309  1119481309  1119491309  1119501309  1119511309  1119521309  1119531309  1119541309  1119551309  1119561309  1119571309  1119581309  1119591309  1119601309  1119611309  1119621309  1119631309  1119641309  1119651309  1119661309  1119671309  1119681309  1119691309  1119701309  1119711309  1119721309  1119731309  1119741309  1119751309  1119771309  1119811309  1119831309  1119851309  1119871309  1119901309  1119921309  1119931309  1119941309  1119951309  1119971309  1119991309  1120001309  1120021309  1120041309  1120061309  1120081309  1120101309  1120121309  1120141309  1120161309  1120181309  1120201309  1120221309  1120251309  1120261309  1120271309  1120281309  1120331309  1120361309  1120371309  1120381309  1120391309  1120411309  1120421309  1217951310  1217971310  1217991310  1218011310  1218031310  1218051310  1218071310  1218091310  1218111310  1218131310  1218151310  1218171310  1218191310  1218211310  1218231310  1218251310  1218271310  1218291310  1218311310  1218331310  1218351310  1218361310  1218371310  1218381310  1218391310  1218411310  1218431310  1218451310  1218471310  1218491310  1218511310  1218531310  1218551310  1218571310  1218591310  1218611310  1218631310  1218651310  1218671310  1218691310  1218731310  1218751310  1218761310  1218771310  1218781310  1218791310  1218801310  1218811310  1218821310  1218831310  1218841310  1218851310  1218861310  1218871310  1218881310  1218891310  1218911310  1218921310  1218931310  1218941310  1236041309  1236051309  1236081309  1236091309  1236101309  1236121309  1236141309  1236161309  1236181309  1236191309  1236201309  1236211309  1236221309  1236231309  1236241309  1236251309  1236271309  1236281309  1236291309  1236301309  1236311309  1236321309  1236331309  1236341309  1236351309  1236361309  1236371309  1236381309  1236391309  1236401309  1236411309  1236421309  1236431309  1236451309  1236471309  1236491309  1236511309  1236531309  1236541309  1236551309  1236561309  1236571309  1236581309  1236591309  1236601309  1236611309  1236621309  1236631309  1236641309  1236651309  1236661309  1236671309  1236681309  1236691309  1236701309  1236711309  1236721309  1236731309  1236741309  1236751309  1236761309  1236771309  1236781309  1236791309  1236801309  1236811309  1236831309  1236851309  1236871309  1236901309  1236921309  1236941309  1236961309  1236991309  1237011309  1237021309  1237031309  1237051309  1237461310  1237471310  1237481310  1237491310  1237511310  1237521310  1237531310  1237551310  1237561310  1237581310  1237601310  1237611310  1237621310  1237631310  1237641310  1237651310  1237721310  1237741310  1237761310  1237781310  1237841310  1237921310  1237941310  1237961310  1238041310  1238051310  1238061310  1238081310  1238101310  1238111310  1238121310  1238141310  1238161310  1238181310  1238201310  1238221310  1238241310  1238261310  1238281310  1238291310  1238301310  1238311310  1238321310  1238341310  1238361310  1238381310  1238401310  1238421310  1238461310  1238481310  1238501310  1238531310  1017621310  1017641310  1017651310  1017671310  1017691310  1017711310  1017741310  1017751310  1017771310  1017781310  1017791310  1017801310  1017811310  1017821310  1017831310  1017841310  1017851310  1017861310  1017871310  1017881310  1017891310  1017901310  1017911310  1017921310  1017931310  1017951310  1017961310  1017971310  1018001310  1018051310  1018071310  1018091310  1018101310  1018111310  1018121310  1018131310  1018141310  1018191310  1018201310  1018211310  1018231310  1018241310  1018251310  1018261310  1018271310  1018281310  1018291310  1018301310  1018311310  1018321310  1018331310  1018351310  1018361310  1018371310  1018381310  1018391310  1018401310  1018411310  1018421310  1018431310  1018441310  1018451310  1018461310  1018471310  1018481310  1018491310  1018511310  1018521310  1018531310  1018541310  1018551310  1018561310  1018571310  1018581310  1018591310  1018601310  1018611310  1018621310  1018631310  1018641310  1111741309  1111761309  1111781309  1111801309  1111821309  1111841309  1111861309  1111891309  1111901309  1111931309  1111951309  1112021309  1112031309  1112041309  1112051309  1112061309  1112071309  1112081309  1112091309  1112101309  1112111309  1112121309  1112131309  1112141309  1112151309  1112161309  1112171309  1112181309  1112191309  1112201309  1112211309  1112221309  1112231309  1112241309  1112251309  1112271309  1112291309  1112311309  1112331309  1112351309  1112371309  1112391309  1112411309  1112431309  1112451309  1112471309  1112491309  1112511309  1112531309  1112551309  1112561309  1112581309  1112601309  1112611309  1112621309  1112641309  1112661309  1112681309  1112701309  1112721309  1112741309  1112761309  1112781309  1112811309  1112821309  1112831309  1274741310  1274751310  1274761310  1274771310  1274781310  1274791310  1274801310  1274811310  1274821310  1274831310  1274841310  1274851310  1274861310  1274871310  1274881310  1274891310  1274901310  1274911310  1274921310  1274931310  1274941310  1274951310  1274961310  1274971310  1274981310  1274991310  1275001310  1275011310  1275021310  1275031310  1275041310  1275051310  1275061310  1275091310  1275111310  1275131310  1275151310  1275171310  1275191310  1275291310  1275311310  1275421310  1275441310  1275461310  1275541310  1275571310  1275611310  1275631310  1275651310  1275691310  1275821308  1275831308  1275841308  1275851308  1275861308  1275871308  1275881308  1275891308  1275911308  1275931308  1275951308  1275961308  1275971308  1275981308  1275991308  1276001308  1276011308  1276021308  1276031308  1276041308  1276051308  1276061308  1276071308  1276081308  1276091308  1276101308  1276111308  1276121308  1276131308  1276141308  1276151308  1276161308  1276171308  1276181308  1276191308  1276201308  1276211308  1276231308  1276251308  1276291308  1276301308  1276311308  1276321308  1276331308  1276341308  1276351308  1276361308  1276371308  1276381308  1276391308  1276431308  1276441308  1276451308  1276461308  1276471308  1276481308  1276491308  1276501308  1276511308  1276531308  1276551308  1276571308  1276581308  1276651308  1276661308  1276671308  1276681308  1276691308  1276701308  1276711308  1276721308  1276731308  1276741308  1276751308  1276761308  1276771308  1276781308  1276791308  1276801308  1276811308  1276821308  1276831308  1276841308  1276851308  1276861308  1276871308  1276881308  1276891308  1276901308  1276911308  1008441309  1008461309  1008471309  1008501309  1008521309  1008531309  1008541309  1008561309  1008581309  1008671309  1008681309  1008701309  1008711309  1008721309  1008731309  1008761309  1008771309  1008781309  1008841309  1008861309  1008881309  1008901309  1008921309  1008941309  1008951309  1008961309  1008981309  1008991309  1009001309  1009021309  1009041309  1009061309  1009081309  1009101309  1009121309  1009141309  1009161309  1009181309  1009201309  1009221309  1009231309  1009241309  1009251309  1009271309  1106281309  1106291309  1106301309  1106311309  1106321309  1106331309  1106341309  1106361309  1106381309  1106391309  1106401309  1106411309  1106421309  1106431309  1106441309  1106451309  1106461309  1106471309  1106491309  1106551309  1106571309  1106581309  1106591309  1106601309  1106611309  1106621309  1106641309  1106661309  1106681309  1106701309  1106721309  1106741309  1106761309  1106781309  1106801309  1106821309  1106841309  1106881309  1106901309  1106971309  1107011309  1107031309  1107051309  1107071309  1107081309  1107101309  1107121309  1107141309  1107181309  1107191309  1107201309  1107211309  1107241309  1107301309  1107321309  1107341309  1107361309  1107381309  1107391309  1107401309  1107411309  1238241309  1238281309  1238541309  1238561309  1238581309  1238601309  1238651309  1238671309  1238691309  1238711309  1238731309  1238751309  1238771309  1238791309  1238811309  1238831309  1238851309  1238951309  1238971309  1238991309  1239011309  1239051309  1239071309  1239101309  1239111309  1239131309  1239151309  1239171309  1239191309  1239211309  1239231309  1260041310  1260061310  1260081310  1260101310  1260161310  1260191310  1260231310  1260241310  1260271310  1260291310  1260311310  1260331310  1260401310  1260421310  1260441310  1260461310  1260541310  1260561310  1260571310  1260591310  1260601310  1260641310  1260681310  1260701310  1260721310  1260741310  1260761310  1260781310  1260801310  1260811310  1260821310  1260831310  1260841310  1260901310  1261011310  1261091310  1261111310  1261131310  1000001311  1000011311  1000021311  1000031311  1000041311  1000051311  1000061311  1000081311  1000101311  1000121311  1000131311  1000141311  1000161311  1000181311  1000241311  1000321311  1000341311  1000361311  1000441311  1000461311  1000521311  1117231309  1117241309  1117251309  1117261309  1117271309  1117281309  1117291309  1117301309  1117311309  1117321309  1117331309  1117341309  1117361309  1117371309  1117381309  1117411309  1117461309  1117471309  1117481309  1117491309  1117501309  1117551309  1117571309  1117591309  1117611309  1117621309  1117631309  1117701309  1117731309  1117741309  1117751309  1117761309  1117771309  1117781309  1117811309  1117831309  1117841309  1117851309  1117861309  1117871309  1117881309  1117901309  1117921309  1117941309  1117961309  1117971309  1117991309  1118021309  1118131309  1118151309  1118161309  1118171309  1118181309  1118191309  1118201309  1118221309  1118241309  1118261309  1118281309  1118301309  1118321309  1118341309  1118361309  1118371309  1118381309  1289291308  1289301308  1289311308  1289321308  1289331308  1289341308  1289361308  1289381308  1289421308  1289441308  1289451308  1289461308  1289531308  1289591308  1289611308  1289631308  1289651308  1289711308  1289731308  1289751308  1289771308  1289811308  1289821308",
      "more_code_info": ""
    }
  ]
}