{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Newark",
      "state": "DE",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85602",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "500 Gbc Dr",
      "address_2": "N/A",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution including in the states of North Carolina, California, Indiana, Florida, Ohio, Arizona, Massachusetts, New Jersey, New Hampshire, Illinois, Minnesota, Arkansas, New York, Delaware, Nevada, Kentucky, Washington DC, Montana, Missouri, Colorado, Maryland, Tennessee, Oregon, Virginia, Iowa, Wisconsin, Louisiana, Georgia, Pennsylvania, North Dakota, Mississippi, South Dakota, Washington, Nebraska, Texas, Alaska, Puerto Rico, Idaho, Oregon, Maine, Vermont, West Virginia, Kansas.",
      "recall_number": "Z-2285-2020",
      "product_description": "Dimension Vista Enzymatic Creatinine, Siemens Material Number 10700444, UDI Number 00842768036262 - Product Usage: use in the quantitative determination of creatinine in human serum, plasma, and urine for the diagnosis and treatment of renal disease.",
      "product_quantity": "23,824",
      "reason_for_recall": "The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.",
      "recall_initiation_date": "20200504",
      "center_classification_date": "20200605",
      "termination_date": "20210920",
      "report_date": "20200617",
      "code_info": "Lot #s  20027AC,  19336AA,  19273AA,  19218AB,  19192AB,  19189AA,  19175AA,  ***Updated 8/4/20***  20083AA                 ***Updated 11/2/20***  20128AB  20202AA"
    }
  ]
}