  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Allentown",
      "address_1": "901 Marcon Blvd",
      "reason_for_recall": "There is a potential for a different set to be mixed within the lot, where the incorrect set does not contain an air eliminating filter.",
      "address_2": "",
      "product_quantity": "1,872",
      "code_info": "Lot Number: 0061583973",
      "center_classification_date": "20180625",
      "distribution_pattern": "US Distribution to the states of :  CA, IL, and TX., and Internationally to Canada.",
      "state": "PA",
      "product_description": "Space Pump IV Set, 2 Caresites, ASV, 0.2 Micron Air Eliminating Filter, Product Code 363420    The Infusomat Space Pump IV Set, 2 Caresites, ASV with 0.2 Micron Air Eliminating Filter and Anti-Syphon Valve (ASV) is a latex-free product intended to be used with the Infusomat Space Pump, to deliver intravenous solutions. The 0.2 Micron filter is a clear acrylic in-line filter designed to eliminate air within the IV infusion set. While it is not listed in the labeling, it is known that the filter is used for removal of particulate matter for some medications and at risk patient populations.",
      "report_date": "20180704",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "B. Braun Medical, Inc.",
      "recall_number": "Z-2285-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80142",
      "termination_date": "20191212",
      "more_code_info": "",
      "recall_initiation_date": "20180509",
      "postal_code": "18109-9512",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}