      "status": "Terminated",
      "city": "Anasco",
      "state": "PR",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68346",
      "recalling_firm": "AMO Puerto Rico Manufacturing, Inc.",
      "address_1": "Rd # 402, Km 4.2",
      "address_2": "",
      "postal_code": "00610",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - Nationwide (USA and Puerto Rico), and foreign:  Latin America, Asia Pacific, Canada, Europe, Africa and Middle East, and Japan.",
      "recall_number": "Z-2285-2014",
      "product_description": "AMO TECNIS CL Foldable Silicone IOL with OptiEdge Design, Model Z9002. Indicated for primary implantation for the visual correction of aphakia in adult patients. Tecnis CL Mono 13 mm with the following diopter sizes: 24.5, 25, 25.5 & 26.    Product Usage:  AMO TECNIS CL monofocal foldable IOLs with OptiEdge design are ultraviolet-absorbing posterior chamber intraocular lenses which compensate for corneal spherical aberrations. TECNIS CL Silicone Intraocular Lenses are indicated for primary implantation for the visual correction of aphakia in adult patients in whom a cataractous lens has been removed by extracapsular cataract extraction or phacoemulsification. These devices are intended to be placed in the capsular bag.",
      "product_quantity": "77,055 IOLs in total",
      "reason_for_recall": "Units may be labeled with the incorrect diopter power.",
      "recall_initiation_date": "20140513",
      "center_classification_date": "20140819",
      "termination_date": "20170518",
      "report_date": "20140827",
      "code_info": "Serial Num  1004711311  1004721311  1004731311  1004741311  1004751311  1004761311  1004771311  1004781311  1004791311  1004801311  1004821311  1004841311  1004861311  1004881311  1004901311  1004921311  1004941311  1004961311  1004971311  1004981311  1004991311  1005001311  1005021311  1005041311  1005061311  1005071311  1005091311  1005111311  1005121311  1005151311  1005161311  1005171311  1005181311  1005191311  1005201311  1005211311  1005221311  1005231311  1005241311  1005251311  1005291311  1005331311  1005351311  1005521311  1005531311  1005541311  1005551311  1005561311  1005571311  1005591311  1005611311  1005651311  1005711311  1005731311  1005751311  1005771311  1005791311  1005811311  1005831311  1005841311  1005851311  1014291310  1014301310  1014311310  1014411310  1014421310  1014431310  1014451310  1014511310  1014521310  1014531310  1014541310  1014561310  1014581310  1014591310  1014601310  1014611310  1014641310  1014661310  1014671310  1014701310  1014721310  1014781310  1014801310  1014811310  1014831310  1014861310  1014881310  1014891310  1014911310  1014921310  1014931310  1014951310  1014971310  1014981310  1014991310  1015001310  1015011310  1015021310  1015031310  1015051310  1015061310  1015071310  1015081310  1015091310  1015101310  1015111310  1015131310  1015141310  1015151310  1015171310  1015181310  1015191310  1015211310  1015231310  1015251310  1015271310  1015291310  1015311310  1015331310  1015341310  1020841310  1020851310  1020861310  1020871310  1020881310  1020891310  1020901310  1020911310  1020921310  1020931310  1020941310  1020951310  1020961310  1020971310  1020981310  1020991310  1021001310  1021021310  1021031310  1021041310  1021051310  1021061310  1021071310  1021081310  1021091310  1021101310  1021111310  1021121310  1021131310  1021141310  1021151310  1021161310  1021171310  1021181310  1021191310  1021201310  1021211310  1021221310  1021231310  1021241310  1021251310  1021261310  1021271310  1021281310  1021291310  1021301310  1021311310  1021321310  1021331310  1021341310  1021351310  1021361310  1021371310  1021381310  1021391310  1021401310  1021411310  1021421310  1021431310  1021441310  1021451310  1021461310  1021471310  1021481310  1021491310  1021501310  1021511310  1021521310  1021531310  1021541310  1021551310  1021561310  1021571310  1021581310  1021591310  1021601310  1021611310  1021621310  1021631310  1021641310  1021651310  1021671310  1021681310  1021691310  1021701310  1021711310  1021721310  1021731310  1021741310  1021751310  1021761310  1021771310  1021781310  1021791310  1021801310  1021811310  1021821310  1021831310  1021841310  1021851310  1021861310  1021871310  1021881310  1021891310  1021901310  1021941310  1021951310  1034081309  1034251309  1034311309  1034331309  1034351309  1034731309  1034751309  1034771309  1034791309  1034851309  1034871309  1034891309  1034911309  1034921309  1034931309  1034951309  1034971309  1034991309  1035011309  1035021309  1035031309  1035051309  1035061309  1035961311  1035971311  1036001311  1036011311  1036041311  1036051311  1036071311  1036081311  1036111311  1036121311  1036151311  1036161311  1036171311  1036191311  1036201311  1036231311  1036241311  1036281311  1036291311  1036311311  1036321311  1036331311  1036351311  1036361311  1036391311  1036401311  1036411311  1036431311  1036441311  1036451311  1036471311  1036481311  1036491311  1036511311  1036521311  1036551311  1036561311  1036591311  1036601311  1036631311  1036641311  1036681311  1036711311  1036721311  1036731311  1036741311  1036751311  1036781311  1036791311  1036821311  1036831311  1036871311  1036901311  1036911311  1036931311  1036951311  1036971311  1036991311  1037011311  1037021311  1037031311  1038921312  1044031309  1044041309  1044051309  1044061309  1044081309  1044091309  1044101309  1044111309  1044121309  1044141309  1044151309  1044161309  1044171309  1044181309  1044191309  1044201309  1044211309  1044221309  1044231309  1044241309  1044261309  1044271309  1044281309  1044291309  1044301309  1044321309  1044341309  1044361309  1044371309  1044381309  1044391309  1044401309  1044411309  1044421309  1044431309  1044441309  1044451309  1044461309  1044471309  1044481309  1044491309  1044501309  1044511309  1044521309  1044531309  1044541309  1044551309  1044561309  1044571309  1044581309  1044591309  1044601309  1044611309  1044621309  1044631309  1044641309  1044651309  1044661309  1044671309  1044681309  1044701309  1044721309  1044741309  1044751309  1044761309  1044771309  1044781309  1044791309  1044801309  1044811309  1044821309  1044831309  1044841309  1044851309  1044861309  1044871309  1044881309  1044891309  1044901309  1044911309  1044921309  1044931309  1044941309  1044951309  1044961309  1044971309  1044981309  1044991309  1045001309  1045011309  1045021309  1045031309  1045041309  1045051309  1045061309  1045071309  1045081309  1045091309  1045101309  1045111309  1045121309  1045131309  1045141309  1045151309  1045161309  1045171309  1047991310  1048001310  1048011310  1048021310  1048031310  1048051310  1048071310  1048091310  1048111310  1048131310  1048151310  1048171310  1048191310  1048211310  1048241310  1048261310  1048281310  1048301310  1048311310  1048321310  1048331310  1048341310  1048351310  1048381310  1048401310  1048421310  1048441310  1048461310  1048481310  1048501310  1048521310  1048541310  1048561310  1048581310  1048601310  1048611310  1048621310  1048631310  1048641310  1048651310  1048661310  1048671310  1048681310  1048691310  1048701310  1048711310  1048721310  1048731310  1048741310  1048751310  1048771310  1048791310  1048811310  1048831310  1048851310  1048871310  1048891310  1048911310  1048931310  1048951310  1048971310  1048991310  1049001310  1049011310  1049021310  1049031310  1049041310  1049051310  1049061310  1049071310  1049081310  1049091310  1049101310  1049111310  1049121310  1049131310  1093111309  1093121309  1093131309  1093141309  1093151309  1093161309  1093171309  1093181309  1093191309  1093201309  1093211309  1093221309  1093231309  1093241309  1093251309  1093261309  1093281309  1093301309  1093321309  1093331309  1093341309  1093361309  1093381309  1093391309  1093401309  1093411309  1093421309  1093431309  1093441309  1093451309  1093461309  1093471309  1093481309  1093491309  1093501309  1093511309  1093521309  1093531309  1093551309  1093561309  1093571309  1093581309  1093591309  1093601309  1093611309  1093631309  1093651309  1093661309  1093671309  1093681309  1093691309  1093701309  1093721309  1093741309  1093781309  1093801309  1093821309  1093841309  1093861309  1093881309  1093901309  1093921309  1093941309  1093961309  1093981309  1094001309  1094021309  1094041309  1094061309  1094081309  1094091309  1094101309  1094111309  1094121309  1094131309  1094141309  1094151309  1116641311  1116651311  1116661311  1116671311  1116681311  1116701311  1116711311  1116721311  1116731311  1116741311  1116761311  1116771311  1116781311  1116801311  1116811311  1116821311  1116831311  1116841311  1116851311  1116861311  1116871311  1116891311  1116911311  1116931311  1116951311  1117011311  1117031311  1117051311  1117061311  1117071311  1117081311  1117091311  1117101311  1117111311  1117121311  1117131311  1117141311  1117151311  1117161311  1117171311  1117181311  1117191311  1117201311  1117211311  1117221311  1117241311  1117261311  1117481311  1117501311  1117521311  1117541311  1117551311  1117561311  1117571311  1117581311  1117591311  1117601311  1117621311  1117641311  1117661311  1117681311  1117701311  1117721311  1125301311  1125311311  1125321311  1125331311  1125341311  1125351311  1125361311  1125371311  1125381311  1125391311  1125401311  1125411311  1125421311  1125431311  1125441311  1125451311  1125461311  1125471311  1125481311  1125491311  1125511311  1125531311  1125551311  1125571311  1125591311  1125611311  1125631311  1125651311  1125671311  1125691311  1125711311  1125751311  1125771311  1125791311  1125811311  1125831311  1125851311  1125871311  1125891311  1125911311  1125931311  1125971311  1125991311  1126011311  1126031311  1126051311  1126071311  1126091311  1126101311  1126111311  1126121311  1126131311  1126141311  1126151311  1126161311  1126171311  1126181311  1126191311  1126201311  1126211311  1126221311  1126231311  1126241311  1126251311  1126261311  1126271311  1126281311  1126291311  1126301311  1126311311  1126321311  1126331311  1126341311  1126351311  1126361311  1126371311  1126381311  1126391311  1126401311  1128691311  1128701311  1128721311  1128731311  1128761311  1128771311  1128791311  1128811311  1128821311  1128831311  1128851311  1128871311  1128881311  1128891311  1128901311  1128911311  1128931311  1128951311  1128961311  1129021311  1129041311  1129121311  1129141311  1129201311  1129221311  1129241311  1129261311  1129271311  1129281311  1129301311  1129321311  1129341311  1129361311  1129381311  1129521311  1129541311  1129641311  1129661311  1129681311  1129711311  1129721311  1129741311  1129751311  1129761311  1129771311  1129781311  1129801311  1129811311  1129821311  1140231309  1140241309  1140251309  1140261309  1140271309  1140281309  1140291309  1140301309  1140311309  1140321309  1140331309  1140341309  1140351309  1140361309  1140371309  1140381309  1140391309  1140401309  1140411309  1140421309  1140431309  1140441309  1140451309  1140461309  1140471309  1140481309  1140491309  1140501309  1140511309  1140521309  1140531309  1140541309  1140551309  1140561309  1140571309  1140581309  1140591309  1140601309  1140611309  1140621309  1140631309  1140641309  1140651309  1140661309  1140671309  1140681309  1140691309  1140701309  1140711309  1140721309  1140731309  1140741309  1140751309  1140761309  1140771309  1140781309  1140791309  1140801309  1140811309  1140821309  1140831309  1140841309  1140861309  1140871309  1140881309  1140891309  1140901309  1140911309  1140921309  1140931309  1140941309  1140951309  1140961309  1140971309  1140981309  1140991309  1141001309  1141011309  1141021309  1141041309  1141051309  1141061309  1141081309  1141101309  1141121309  1141141309  1141151309  1141161309  1141171309  1141181309  1141191309  1141201309  1141211309  1141221309  1141231309  1141241309  1141251309  1141261309  1141271309  1141281309  1141291309  1141301309  1141311309  1141321309  1141331309  1141341309  1141351309  1141361309  1141371309  1162591310  1162601310  1162611310  1162621310  1162631310  1162641310  1162651310  1162661310  1162681310  1162701310  1162721310  1162741310  1162751310  1162761310  1162771310  1162781310  1162791310  1162801310  1162811310  1162821310  1162831310  1162841310  1162851310  1162861310  1162871310  1162881310  1162891310  1162901310  1162911310  1162921310  1162931310  1162951310  1162971310  1162991310  1163011310  1163021310  1163031310  1163051310  1163071310  1163091310  1163111310  1163131310  1163151310  1163171310  1163191310  1163201310  1163211310  1163221310  1163231310  1163241310  1163251310  1163261310  1163271310  1163281310  1163291310  1163301310  1163311310  1163321310  1163331310  1163341310  1163351310  1163371310  1163391310  1163411310  1163431310  1163441310  1163451310  1163461310  1163471310  1163481310  1163491310  1163501310  1163511310  1163521310  1163531310  1163541310  1163551310  1163561310  1163571310  1163581310  1163591310  1163601310  1163611310  1163621310  1163641310  1163661310  1163681310  1175531310  1175541310  1175551310  1175561310  1175571310  1175581310  1175591310  1175601310  1175611310  1175621310  1175631310  1175651310  1175661310  1175671310  1175691310  1175711310  1175731310  1175751310  1175771310  1175801310  1175821310  1175851310  1175871310  1175891310  1175901310  1175911310  1175921310  1175941310  1175961310  1175981310  1176001310  1176021310  1176041310  1176061310  1176171310  1176181310  1176191310  1176201310  1176211310  1176221310  1176231310  1176241310  1176251310  1176261310  1176271310  1176281310  1176291310  1176301310  1176311310  1176321310  1176331310  1176341310  1176351310  1176361310  1176371310  1176381310  1176391310  1176401310  1176411310  1176421310  1176431310  1176441310  1176451310  1176461310  1176471310  1176481310  1176491310  1176501310  1176511310  1176521310  1176531310  1176541310  1176551310  1176571310  1176581310  1176591310  1176601310  1176611310  1176621310  1176631310  1215401309  1215421309  1215441309  1215451309  1215461309  1215471309  1215481309  1215501309  1215521309  1215541309  1215551309  1215601309  1215621309  1215641309  1215661309  1215681309  1215701309  1215721309  1215741309  1215761309  1215801309  1215821309  1215841309  1215861309  1215881309  1215891309  1215901309  1215911309  1215921309  1215931309  1215941309  1215951309  1215961309  1215991309  1216011309  1216031309  1216051309  1216071309  1216091309  1216111309  1216131309  1216151309  1216171309  1216181309  1216191309  1216201309  1216211309  1216231309  1216251309  1216261309  1216271309  1216291309  1216331309  1216351309  1216371309  1216411309  1216431309  1216441309  1216451309  1216461309  1216471309  1216481309  1216491309  1216501309  1216511309  1216521309  1224341310  1224351310  1224361310  1224371310  1224381310  1224391310  1224401310  1224411310  1224421310  1224431310  1224441310  1224451310  1224461310  1224471310  1224481310  1224491310  1224501310  1224511310  1224521310  1224531310  1224541310  1224551310  1224561310  1224581310  1224601310  1224611310  1224621310  1224631310  1224641310  1224651310  1224661310  1224681310  1224691310  1224701310  1224721310  1224731310  1224771310  1224791310  1224811310  1224851310  1224871310  1224891310  1224931310  1225001310  1225141310  1225161310  1225181310  1225201310  1225221310  1225241310  1225261310  1225281310  1225301310  1225321310  1225341310  1225361310  1257931308  1257941308  1257951308  1257961308  1257971308  1257981308  1257991308  1258001308  1258011308  1258021308  1258031308  1258041308  1258061308  1258071308  1258081308  1258091308  1258101308  1258111308  1258121308  1258131308  1258141308  1258151308  1258161308  1258211308  1258231308  1258251308  1258271308  1258281308  1258291308  1258301308  1258311308  1258321308  1258331308  1258341308  1258351308  1258361308  1258371308  1258381308  1258391308  1258401308  1258411308  1258421308  1258431308  1258441308  1258451308  1258461308  1258471308  1258481308  1258491308  1258501308  1258511308  1258521308  1258531308  1258541308  1258551308  1258561308  1258571308  1258581308  1258591308  1258601308  1258611308  1258621308  1258641308  1258661308  1258681308  1258701308  1258711308  1258721308  1258731308  1258741308  1258761308  1258781308  1258801308  1258821308  1258841308  1258861308  1258881308  1258901308  1258921308  1258941308  1258961308  1258981308  1261201308  1261211308  1261221308  1261231308  1261241308  1261251308  1261261308  1261271308  1261281308  1261291308  1261301308  1261311308  1261321308  1261331308  1261341308  1261351308  1261361308  1261371308  1261391308  1261411308  1261421308  1261431308  1261441308  1261451308  1261461308  1261471308  1261481308  1261491308  1261501308  1261511308  1261521308  1261531308  1261551308  1261571308  1261591308  1261611308  1261701308  1261721308  1261731308  1261741308  1261751308  1261761308  1261771308  1261781308  1261791308  1261801308  1261811308  1261821308  1261831308  1261841308  1261851308  1261861308  1261871308  1261881308  1261891308  1261901308  1261911308  1261921308  1261931308  1261941308  1261951308  1261961308  1261971308  1261981308  1261991308  1262001308  1262011308  1262021308  1262031308  1262041308  1262051308  1262061308  1262071308  1262081308  1262091308  1262101308  1262111308  1262121308  1262141308  1262151308  1262161308  1262171308  1262181308  1262191308  1262201308  1013191310  1013211310  1013231310  1013251310  1013271310  1013291310  1013311310  1013331310  1013341310  1013351310  1013361310  1013371310  1013381310  1013401310  1013411310  1013421310  1013431310  1013441310  1013451310  1013461310  1013471310  1013491310  1013501310  1013521310  1013541310  1013561310  1013581310  1013601310  1013621310  1013641310  1013661310  1013681310  1013701310  1013721310  1013741310  1013761310  1013781310  1013801310  1013821310  1013841310  1013851310  1013861310  1013871310  1013881310  1013891310  1013901310  1013911310  1013921310  1013931310  1013941310  1013951310  1013961310  1013971310  1013981310  1013991310  1014001310  1014011310  1014021310  1014031310  1014041310  1014051310  1014061310  1014071310  1014081310  1014091310  1014101310  1014111310  1014121310  1014131310  1014141310  1014151310  1014161310  1014171310  1014181310  1014191310  1014201310  1014211310  1014221310  1014231310  1014241310  1014251310  1014261310  1014271310  1019631312  1019801312  1019841312  1019901312  1019921312  1019941312  1019961312  1019981312  1020001312  1020081312  1020151312  1020171312  1020191312  1020431312  1020451312  1044871310  1044881310  1044891310  1044901310  1044921310  1044941310  1044961310  1044981310  1045001310  1045011310  1045021310  1045041310  1045061310  1045081310  1045101310  1045121310  1045141310  1045151310  1045161310  1045171310  1045181310  1045191310  1045201310  1045211310  1045221310  1045231310  1045241310  1045251310  1045261310  1045271310  1045281310  1045291310  1045301310  1045311310  1045321310  1045341310  1045351310  1045361310  1045371310  1045381310  1045401310  1045411310  1045421310  1045431310  1045441310  1045451310  1045461310  1045471310  1045481310  1045491310  1045501310  1045511310  1045521310  1045531310  1045541310  1045551310  1045561310  1045571310  1045581310  1045591310  1045601310  1045611310  1045621310  1045631310  1045641310  1045651310  1045661310  1045671310  1045691310  1045731310  1045751310  1045771310  1045881310  1059471309  1059531309  1059551309  1059571309  1059591309  1059611309  1059631309  1059651309  1059671309  1059691309  1059711309  1059731309  1059741309  1059751309  1059761309  1059771309  1059781309  1059791309  1059801309  1059811309  1059821309  1059831309  1059841309  1059851309  1059861309  1059871309  1059881309  1059891309  1059901309  1059911309  1059921309  1059941309  1059961309  1059981309  1060001309  1060021309  1060041309  1060061309  1060081309  1060101309  1060121309  1060141309  1060161309  1060181309  1060201309  1060221309  1060241309  1060261309  1060281309  1060301309  1060321309  1060341309  1060361309  1060371309  1060381309  1060391309  1060401309  1060411309  1060421309  1060431309  1060441309  1060451309  1060461309  1060471309  1060481309  1060491309  1060501309  1060521309  1060531309  1060541309  1060551309  1060561309  1060571309  1060581309  1060591309  1060601309  1069361309  1069371309  1069401309  1069411309  1069431309  1069441309  1069451309  1069471309  1069481309  1069491309  1069511309  1069521309  1069531309  1069551309  1069561309  1069571309  1069591309  1069611309  1069631309  1069651309  1069681309  1069691309  1069721309  1069731309  1069761309  1069771309  1069781309  1069811309  1069821309  1069831309  1069851309  1069861309  1069871309  1069891309  1069901309  1069911309  1069931309  1069951309  1069961309  1069971309  1069991309  1070001309  1070011309  1070031309  1070041309  1070051309  1070071309  1070081309  1070121309  1070141309  1070151309  1070161309  1070191309  1070201309  1070231309  1070241309  1070261309  1070271309  1070281309  1070301309  1070311309  1070321309  1070341309  1070351309  1070361309  1070381309  1104011312  1104031312  1104051312  1104071312  1104091312  1104111312  1104131312  1104151312  1104391312  1104411312  1104431312  1104491312  1104571312  1104591312  1104611312  1104631312  1104651312  1104671312  1104691312  1104711312  1104731312  1104751312  1104771312  1104791312  1104811312  1104831312  1104851312  1104871312  1129841310  1129851310  1129861310  1129871310  1129881310  1129891310  1129901310  1129911310  1129921310  1129931310  1129941310  1129951310  1129961310  1129971310  1129981310  1129991310  1130001310  1130011310  1130021310  1130031310  1130041310  1130051310  1130061310  1130081310  1130101310  1130111310  1130121310  1130131310  1130141310  1130151310  1130181310  1130191310  1130231310  1130251310  1130271310  1130281310  1130291310  1130301310  1130311310  1130321310  1130331310  1130341310  1130351310  1130361310  1130371310  1130381310  1130391310  1130401310  1130411310  1130421310  1130431310  1130441310  1130451310  1130461310  1130471310  1130481310  1130491310  1130501310  1130511310  1130521310  1130531310  1130541310  1130561310  1130581310  1130741310  1130761310  1139991310  1140001310  1140021310  1140041310  1140061310  1140081310  1140101310  1140121310  1140141310  1140151310  1140161310  1140171310  1140181310  1140211310  1140221310  1140231310  1140251310  1140261310  1140271310  1140281310  1140311310  1140331310  1140351310  1140391310  1140411310  1140431310  1140571310  1140611310  1140631310  1140651310  1140661310  1140671310  1140681310  1140691310  1140701310  1140711310  1140721310  1140731310  1140741310  1140751310  1140771310  1140781310  1140791310  1140811310  1140831310  1140841310  1140851310  1140861310  1140871310  1140881310  1140891310  1140901310  1140911310  1140921310  1140931310  1145951309  1146221309  1146241309  1146261309  1146281309  1146301309  1146321309  1146341309  1146361309  1146381309  1146401309  1146411309  1146421309  1146431309  1146441309  1146451309  1146461309  1146471309  1146481309  1146491309  1146501309  1146511309  1146521309  1146531309  1146541309  1146551309  1146561309  1146571309  1146581309  1146591309  1146601309  1146611309  1146621309  1146631309  1146671309  1146791309  1146811309  1146851309  1165681311  1165701311  1165711311  1165721311  1165731311  1165741311  1165751311  1165761311  1165781311  1165791311  1165801311  1165821311  1165841311  1165881311  1165901311  1165911311  1165921311  1165931311  1165941311  1165971311  1165991311  1166001311  1166021311  1166081311  1166101311  1166351311  1166451311  1166471311  1166561311  1166581311  1166591311  1166601311  1166621311  1166641311  1166661311  1166681311  1166771311  1166801311  1205411310  1205421310  1205441310  1205461310  1205471310  1205481310  1205501310  1205521310  1205531310  1205541310  1205551310  1205561310  1205581310  1205601310  1205651310  1205661310  1205671310  1205681310  1205691310  1205701310  1205711310  1205731310  1205751310  1205771310  1205791310  1205801310  1205821310  1205831310  1205841310  1205851310  1205861310  1205871310  1205881310  1205891310  1205901310  1205911310  1205921310  1205941310  1205951310  1205961310  1205971310  1205981310  1205991310  1206001310  1206011310  1206021310  1206031310  1206041310  1206051310  1206061310  1206071310  1206081310  1206091310  1206101310  1206111310  1206121310  1206131310  1206141310  1206151310  1206161310  1206171310  1206181310  1206191310  1206201310  1206221310  1206241310  1206261310  1206281310  1206291310  1206301310  1206321310  1206331310  1206341310  1206361310  1206381310  1206401310  1214511309  1214531309  1214551309  1214561309  1214571309  1214581309  1214591309  1214601309  1214611309  1214631309  1214641309  1214651309  1214671309  1214681309  1214691309  1214711309  1214731309  1214751309  1214761309  1214771309  1214781309  1214811309  1214831309  1214851309  1214871309  1214881309  1214891309  1214901309  1214911309  1214921309  1214931309  1214941309  1214951309  1214961309  1214971309  1214981309  1214991309  1215001309  1215011309  1215021309  1215031309  1215041309  1215051309  1215061309  1215071309  1215081309  1215091309  1215101309  1215111309  1215121309  1215131309  1215141309  1215151309  1215161309  1215171309  1215181309  1215191309  1215201309  1215211309  1215221309  1215231309  1215241309  1215251309  1215291309  1215331309  1215351309  1215361309  1215371309  1215381309  1215391309  1230441311  1230451311  1230461311  1230471311  1230481311  1230491311  1230501311  1230511311  1230521311  1230531311  1230541311  1230551311  1230561311  1230571311  1230581311  1230591311  1230601311  1230611311  1230621311  1230641311  1230661311  1230671311  1230681311  1230691311  1230701311  1230711311  1230721311  1230731311  1230741311  1230751311  1230761311  1230771311  1230781311  1230791311  1230801311  1230811311  1230821311  1230831311  1230851311  1230871311  1230891311  1230911311  1230931311  1230951311  1230971311  1230991311  1231011311  1231031311  1231051311  1231071311  1231091311  1231111311  1231131311  1231151311  1231171311  1231191311  1231211311  1231241311  1231261311  1231281311  1231301311  1231321311  1231341311  1231361311  1231381311  1231401311  1231421311  1231441311  1276971310  1276991310  1277011310  1277021310  1277051310  1277071310  1277091310  1277111310  1277141310  1277161310  1277201310  1277261310  1277271310  1277281310  1277301310  1277321310  1277361310  1277381310  1277401310  1277421310  1277441310  1277451310  1277471310  1277501310  1277521310  1277541310  1277561310  1277571310  1277581310  1277591310  1277601310  1277611310  1277621310  1277641310  1277661310  1277671310  1277681310  1277701310  1277721310  1277741310  1277751310  1277761310  1277781310  1277801310  1277821310  1277841310  1277861310  1277901310  1037991309  1038011309  1038031309  1038041309  1038051309  1038071309  1038081309  1038091309  1038101309  1038111309  1038121309  1038131309  1038151309  1038171309  1038191309  1038211309  1038231309  1038251309  1038271309  1038291309  1038311309  1038331309  1038351309  1038371309  1038381309  1038391309  1038411309  1038421309  1038431309  1038441309  1038451309  1038471309  1038491309  1038501309  1038521309  1038531309  1038541309  1038551309  1038571309  1038591309  1038611309  1038621309  1038631309  1038641309  1038651309  1038671309  1038681309  1038691309  1038711309  1038731309  1038751309  1038761309  1038771309  1038781309  1038791309  1038801309  1038811309  1038821309  1038831309  1038921309  1038961309  1038981309  1039001309  1111371312  1111391312  1111411312  1111431312  1111481312  1111701312  1111741312  1111761312  1111781312  1111801312  1111821312  1111861312  1111881312  1111901312  1111921312  1111941312  1111961312  1111981312  1112001312  1112021312  1112041312  1112061312  1112071312  1112081312  1112111312  1112251312  1113461312  1113481312  1113521312  1113581312  1113601312  1113741312  1114231312  1114251312  1114271312  1115951310  1115961310  1115971310  1115981310  1115991310  1116001310  1116011310  1116021310  1116031310  1116041310  1116051310  1116061310  1116071310  1116081310  1116091310  1116101310  1116111310  1116131310  1116151310  1116161310  1116171310  1116181310  1116201310  1116211310  1116221310  1116231310  1116241310  1116251310  1116261310  1116271310  1116281310  1116291310  1116301310  1116311310  1116321310  1116331310  1116341310  1116351310  1116361310  1116371310  1116381310  1116391310  1116401310  1116411310  1116421310  1116431310  1116441310  1116451310  1116461310  1116471310  1116481310  1116491310  1116501310  1116511310  1116531310  1116551310  1116571310  1116581310  1116611310  1116631310  1116821310  1116831310  1116841310  1116851310  1116861310  1116871310  1116881310  1116891310  1116901310  1116911310  1116931310  1116941310  1116951310  1116961310  1116971310  1116991310  1117011310  1117031310  1120081310  1120091310  1120101310  1120111310  1120121310  1120131310  1120141310  1120161310  1120171310  1120181310  1120191310  1120201310  1120211310  1120221310  1120231310  1120241310  1120251310  1120261310  1120271310  1120281310  1120291310  1120301310  1120311310  1120321310  1120331310  1120351310  1120371310  1120391310  1120411310  1120431310  1120451310  1120471310  1120481310  1120501310  1120521310  1120541310  1120561310  1120571310  1120581310  1120591310  1120601310  1120611310  1120621310  1120631310  1120641310  1120661310  1120701310  1120721310  1120741310  1120761310  1120781310  1120851310  1120861310  1120871310  1120881310  1120891310  1120901310  1120911310  1120921310  1120931310  1120951310  1120961310  1120971310  1120981310  1120991310  1121001310  1121011310  1121021310  1121031310  1121041310  1121051310  1121071310  1121091310  1121111310  1121131310  1121151310  1121161310  1121181310  1131771310  1131791310  1131811310  1131831310  1131851310  1131871310  1131891310  1131911310  1131931310  1131941310  1131951310  1131961310  1131971310  1131981310  1131991310  1132001310  1132011310  1132021310  1132031310  1132041310  1132051310  1132071310  1132091310  1132101310  1132111310  1132121310  1132131310  1132141310  1132151310  1132161310  1132171310  1132181310  1132191310  1132211310  1132221310  1132231310  1132251310  1132271310  1132291310  1132311310  1132331310  1132341310  1132351310  1132361310  1132371310  1132381310  1132391310  1132401310  1132411310  1132421310  1132431310  1132441310  1132451310  1132461310  1132471310  1132481310  1132491310  1132501310  1132511310  1132521310  1132531310  1132541310  1132551310  1132561310  1132571310  1132581310  1132591310  1144681309  1144691309  1144701309  1144721309  1144741309  1144761309  1144771309  1144781309  1144801309  1144821309  1144951309  1144971309  1144991309  1145011309  1145031309  1145051309  1145071309  1145091309  1145111309  1145121309  1145131309  1145141309  1145151309  1145161309  1145171309  1145181309  1145191309  1145201309  1145211309  1145221309  1145231309  1145241309  1145251309  1145261309  1145271309  1145281309  1145291309  1145301309  1145311309  1145321309  1145331309  1145341309  1145361309  1145381309  1145401309  1145421309  1145441309  1145461309  1145481309  1145501309  1145521309  1145541309  1145601309  1145621309  1145641309  1145661309  1145681309  1145701309  1145721309  1220091310  1220101310  1220111310  1220121310  1220131310  1220141310  1220151310  1220161310  1220171310  1220181310  1220191310  1220201310  1220211310  1220221310  1220231310  1220241310  1220251310  1220261310  1220271310  1220281310  1220291310  1220301310  1220311310  1220321310  1220331310  1220341310  1220351310  1220361310  1220371310  1220381310  1220391310  1220401310  1220421310  1220431310  1220441310  1220451310  1220471310  1220491310  1220511310  1220531310  1220551310  1220571310  1220591310  1220611310  1220651310  1220671310  1220861310  1220881310  1220901310  1220951310  1220971310  1220991310  1221001310  1221011310  1221021310  1221031310  1221041310  1221051310  1221061310  1221071310  1221081310  1221091310  1221101310  1221121310  1221131310  1221141310  1221151310  1221161310  1221181310  1221201310  1221211310  1226451310  1226461310  1226471310  1226481310  1226491310  1226501310  1226511310  1226521310  1226531310  1226541310  1226551310  1226561310  1226571310  1226581310  1226591310  1226601310  1226611310  1226621310  1226631310  1226641310  1226651310  1226661310  1226671310  1226681310  1226691310  1226701310  1226711310  1226721310  1226731310  1226741310  1226751310  1226761310  1226771310  1226781310  1226791310  1226801310  1226811310  1226821310  1226841310  1226861310  1226891310  1226911310  1226921310  1226931310  1226951310  1226971310  1226991310  1227011310  1227031310  1227051310  1227061310  1227071310  1227081310  1227091310  1227101310  1227111310  1227121310  1227131310  1227141310  1227151310  1227161310  1227171310  1227181310  1227191310  1227201310  1227211310  1227221310  1227231310  1227251310  1227271310  1227281310  1227291310  1227301310  1227311310  1227321310  1227331310  1227341310  1227351310  1227371310  1227391310  1227411310  1227431310  1227451310  1227471310  1227491310  1227511310  1227531310  1227551310  1227571310  1231851309  1231861309  1231871309  1231881309  1231891309  1231901309  1231911309  1231921309  1231931309  1231941309  1231951309  1231961309  1231971309  1231981309  1231991309  1232001309  1232011309  1232021309  1232031309  1232041309  1232051309  1232061309  1232071309  1232081309  1232091309  1232101309  1232111309  1232121309  1232131309  1232151309  1232171309  1232181309  1232191309  1232201309  1232211309  1232221309  1232231309  1232241309  1232251309  1232261309  1232271309  1232281309  1232291309  1232301309  1232311309  1232321309  1232331309  12323413",
      "more_code_info": "09  1232351309  1232361309  1232371309  1232381309  1232401309  1232421309  1232521309  1232541309  1232561309  1232581309  1232601309  1232621309  1232641309  1232651309  1232661309  1232681309  1232691309  1232701309  1232711309  1232731309  1232751309  1232771309  1232791309  1232801309  1232811309  1232831309  1232841309  1232861309  1232891309  1232911309  1232931309  1232951309  1232961309  1232971309  1232981309  1233001309  1233021309  1233041309  1233061309  1233081309  1233091309  1233101309  1233121309  1233141309  1233161309  1233181309  1233201309  1233221309  1233241309  1233251309  1233261309  1233271309  1233281309  1233291309  1233301309  1233311309  1233321309  1233331309  1233341309  1233351309  1233361309  1233371309  1233381309  1233391309  1233401309  1233411309  1233421309  1233431309  1233441309  1233451309  1233461309  1233471309  1233481309  1233491309  1233501309  1233511309  1233521309  1233531309  1233551309  1233561309  1233571309  1233581309  1233591309  1233601309  1233611309  1233621309  1233631309  1233641309  1233651309  1233661309  1233671309  1233681309  1233691309  1233711309  1233731309  1233741309  1233751309  1233771309  1233791309  1233811309  1233831309  1233851309  1233871309  1233891309  1233901309  1233911309  1243871310  1243891310  1243901310  1243921310  1243931310  1243941310  1243951310  1243961310  1243971310  1243981310  1243991310  1244001310  1244021310  1244041310  1244051310  1244061310  1244081310  1244101310  1244121310  1244141310  1244161310  1244181310  1244191310  1244201310  1244221310  1244241310  1244261310  1244281310  1244321310  1244331310  1244341310  1244351310  1244361310  1244371310  1244381310  1244391310  1244401310  1244411310  1244421310  1244441310  1244461310  1244481310  1244501310  1244511310  1244521310  1244531310  1244541310  1244561310  1244571310  1244581310  1244591310  1244601310  1244611310  1244621310  1244631310  1244641310  1244651310  1244661310  1244671310  1244681310  1244691310  1244701310  1244711310  1244731310  1244771310  1244791310"
    }
  ]
}