  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mentor",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62776",
      "recalling_firm": "Steris Corporation",
      "address_1": "5960 Heisley Rd",
      "address_2": "N/A",
      "postal_code": "44060-1834",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Visit",
      "distribution_pattern": "Nationwide distribution:  USA including states of: AR, AK, CA, CO, DE, FL, IL, MA, MD, NC, NE, NH, NY, OH, PA, SD, TN, TX, WI, and WV.",
      "recall_number": "Z-2284-2012",
      "product_description": "QKC1692E     The QKC1692E Quick Connect device connects endoscopes to the sterilization tray for C1220E liquid chemical sterilization processor.",
      "product_quantity": "64 units",
      "reason_for_recall": "STERIS has been actively seeking devices for quick connect revalidation projects.  STERIS was unable to obtain two (2) of the needed devices; these two devices represent the design of seven (7) models in one quick connect.  Due to the unavailability of these two validation devices, STERIS will remove 7 models from the labeling of QKC1692E and will re-label all QKC1692Es in the field..",
      "recall_initiation_date": "20120810",
      "center_classification_date": "20120828",
      "termination_date": "20130705",
      "report_date": "20120905",
      "code_info": "Lot numbers: 1879717, 1943786, 2054914, 2143717, 2177723, 2342319, 2730752, 2922565, 3094190"
    }
  ]
}