      "status": "Terminated",
      "city": "Portage",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65922",
      "recalling_firm": "Stryker Medical Division of Stryker Corporation",
      "address_1": "3800 E Centre Ave",
      "address_2": "N/A",
      "postal_code": "49002-5826",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including MI, WI, VA and internationally to Canada.",
      "recall_number": "Z-2283-2013",
      "product_description": "The FL23SE Bed is an AC-powered adjustable hospital bed with four built-in electric DC motors and remote controls that is intended for medical purposes. It can be operated by the patient or the caregiver to adjust the height and surface contour of the bed. The bed includes movable and latchable siderails.  The FL23SE Bed is intended for use with patients for procedures, therapy, recovery in healthcare environments, and to transport patients between bays and procedural rooms.    The FL23SE Bed is intended for use with patients for procedures, therapy, recovery in healthcare environments, and to transport patients between bays and procedural rooms.",
      "product_quantity": "61 units",
      "reason_for_recall": "The nurse call cable (Part Number: QDF23-0571) on the FL23SE Bed is located right under the opening of the bed head box. If an accessory (i.e. IV pole) is inserted into this slot, the circuit directing the nurse call system can be damaged.",
      "recall_initiation_date": "20130819",
      "center_classification_date": "20130926",
      "termination_date": "20140129",
      "report_date": "20131002",
      "code_info": "Model Numbers: FL23SE with nurse call option: OL235017 & OL235018  List of Affected Devices:  J60267 J60425 J60009 J60001 J60094  J60095 J60096 J60348 J60349 J60350  J60351 J60352 J60353 J60354 J60360  J60361 J60362 J60363 J60364 J60365  J60366 J60367 J60368 J60369 J60370  J60371 J60372 J60373 J60374 J60375  J60376 J60377 J60378 J60379 J60380  J60381 J60382 J60383 J60384 J60385  J60386 J60387 J60388 J60389 J60390  J60391 J60392 J60393 J60394 J60395  J60396 J60397 J60398 J60399 J60400  J60401 J60402 J60403 J60404 J60405  J60512"
    }
  ]
}