  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74500",
      "recalling_firm": "Stryker Howmedica Osteonics Corp.",
      "address_1": "325 Corporate Dr",
      "address_2": "N/A",
      "postal_code": "07430-2006",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.  US nationwide (AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MS, NJ, NY, OH, PA, TX, WA, WI, WV),  Sweden, South Africa,  Chile, Australia,  New Zealand,  Canada,  United Kingdom, Switzerland, Colombia,  and Poland.",
      "recall_number": "Z-2281-2016",
      "product_description": "Stryker Orthopaedics Trident Constrained Liner Inserter/Impactor Tip, 22 mm, 28 mm, and 32 mm.  Non Sterile    Catalog Numbers 2199-2022, 2199-2028 and 2199-2032.    For use in total hip arthroplasty.",
      "product_quantity": "1,864 units",
      "reason_for_recall": "Impactors do not meet the required sterility assurance level (SAL) of 10-6 when using the recommended steam sterilization method.",
      "recall_initiation_date": "20160603",
      "center_classification_date": "20160725",
      "termination_date": "20180801",
      "report_date": "20160803",
      "code_info": "Catalog Nos. 2199-2022, 2199-2028 and 2199-2032  All lots manufactured from May 30, 2012"
    }
  ]
}