  "results": [
    {
      "status": "Terminated",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66150",
      "recalling_firm": "CareFusion 303, Inc.",
      "address_1": "10020 Pacific Mesa Blvd",
      "address_2": "N/A",
      "postal_code": "92121-4386",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "CA, FL, AZ, and TN",
      "recall_number": "Z-2279-2013",
      "product_description": "SmartSite Low Sorbing Infusion Set, Model#72313E    The SmartSite Low Sorbing Infusion Sets are used to administer fluid and medications through a needle or catheter inserted into the patient's artery or vein. The SmartSite Low Sorbing set is comprised of components commonly found on intravascular administration sets and extension sets. It includes a 1.2 micron filter, needleless SmartSite connector, low sorbing tubing segment, pinch clamps, and luer connectors.",
      "product_quantity": "600 units",
      "reason_for_recall": "CareFusion is recalling the SmartSite Low Sorbing Infusion Set, Model 72313E, Lot Number  13016314, because of an incorrect filter was assembled on this lot.  The correct filter is a 1.2 micron  filter (blue color) but a 0.2 micron filter (clear color) may have been added instead.",
      "recall_initiation_date": "20130830",
      "center_classification_date": "20130925",
      "termination_date": "20131028",
      "report_date": "20131002",
      "code_info": "Lot Number 13016314"
    }
  ]
}