  "results": [
    {
      "status": "Ongoing",
      "city": "Irving",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88324",
      "recalling_firm": "Abbott Laboratories, Inc",
      "address_1": "1921 Hurd Dr",
      "address_2": "N/A",
      "postal_code": "75038-4313",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Visit",
      "distribution_pattern": "Worldwide distribution. US nationwide, Australia, Bahrain, Belgium, Brazil, China, Colombia, Croatia, Egypt, Germany, Greece, Hong Kong, Hungary, Iran, Ireland, Israel, Italy, Jordan, Kuwait, Mexico, Netherlands, Pakistan, Poland, Portugal, Saudi Arabia, Serbia, Slovenia, South Africa, Spain, Tanzania, United Arab Emirates, United Kingdom, Ukraine and Zimbabwe.",
      "recall_number": "Z-2276-2021",
      "product_description": "Alinity s System REF 06P16-01.   An automated immunoassay analyzer for In Vitro diagnostic use in blood donor screening, plasma and plasmapheresis screening, and organ donor centers, hospitals, and reference laboratories.",
      "product_quantity": "260 systems",
      "reason_for_recall": "A design defect (hardware and software) allows liquid waste pressure to build up and potentially spray users/operators.",
      "recall_initiation_date": "20210712",
      "center_classification_date": "20210812",
      "report_date": "20210818",
      "code_info": "Model Number:  LN 06P16-01  Serial numbers AS1001 through AS1330"
    }
  ]
}