  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Danvers",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83429",
      "recalling_firm": "NeuroLogica Corporation",
      "address_1": "14 Electronics Ave",
      "address_2": "N/A",
      "postal_code": "01923-1011",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide",
      "recall_number": "Z-2276-2019",
      "product_description": "WS80A Diagnostic Ultrasound System Version 1.00.16, 3.00.17, 3.00.18, 3.00.19, 3.01.00, 3.01.01, 3.01.02, 3.01.03, 3.01.04, 3.01.05, 3.01.06, 3.01.07,  3.01.08, 3.01.09, 4.00.08, 4.00.09, 4.00.10, 4.00.11, 4.01.00, 4.01.01, 4.01.02,  4.01.03, 4.01.04, 4.01.05, 4.01.06, 4.01.07, 4.01.08, 4.01.09.",
      "product_quantity": "5088 worldwide",
      "reason_for_recall": "There is a potential for probes overheating when decreasing the Doppler SV (Sample Volume) size value set by the user in the Doppler Only mode.",
      "recall_initiation_date": "20190725",
      "center_classification_date": "20190814",
      "termination_date": "20200911",
      "report_date": "20190821",
      "code_info": "UDI: 08806167780081    Serial Numbers:  S116M3HJ600006V  S116M3HJ600002H  S116M3HJ400012W  S116M3HJ400013V  S18EM3HJA00002B  S18EM3HK900002Y  S18EM3HK800001P  S18EM3HJB00003L  S18EM3HJB00004T  S18EM3HJB00005B  S116M3HJ600008T  S116M3HJ500005T  S18EM3HJB00001W  S18EM3HJB00002V  S18EM3HJB00006K  S18EM3HJA00003K  S18EM3HJ900017P  S18EM3HJ900018D  S18EM3HK800005E  S18EM3HK700012D  S1B1M3HK100001X  S18EM3HK700011P  S18EM3HK700013X  S18EM3HK700007M  S18EM3HK700009V  S18EM3HK700010A  S18EM3HK700005H  S18EM3HK700006J  S18EM3HK700001N  S18EM3HK700003F  S18EM3HK700004R  S18EM3HK900001Z  S18EM3HK900003A  S18EM3HK900005D  S18EM3HK700008W  S18EM3HK800002D  S18EM3HK800003X  S18EM3HK800004N  S116M3HJ500004L  S18EM3HJA00001T  S18EM3HK900006X  S18EM3HK900007N  S18EM3HK900008E  S18EM3HK900009F  S18EM3HK900010L  S18EM3HKC00001H  S18EM3HKC00002J  S18EM3HK900004P  S18EM3HK800006F  S18EM3HK700002E  S18EM3HM500009Y  S18EM3HM500010F  S18EM3HM500011R  S18EM3HM500012H  S18EM3HM500007K  S18EM3HM500008Z  S18EM3HM500001M  S18EM3HM500002W  S18EM3HM500003V  S18EM3HM500004L  S18EM3HM500005T  S18EM3HM500006B  S18EM3HM400001L"
    }
  ]
}